A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table. Product Usage: Product provides a flexible, easy-to-use, articulated posturing of a patient for either general, I.A., or specialized surgical procedures. There are movable components and accessories intended for use during diagnostic examinations or surgical procedures to support and position a patient.
Enforcement
- Recall Number
- Z-2402-2012
- Event ID
- 54705
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Steris Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 26, 2012
- Initiation Date
- February 17, 2010
- Classification Date
- September 17, 2012
- Termination Date
- March 13, 2013
- Address
- 5960 Heisley Rd, Mentor, OH, 44060, United States
Description
A mobile, electro-hydraulically operated, fluoroscopic/C-Arm compatible general surgical table. Product Usage: Product provides a flexible, easy-to-use, articulated posturing of a patient for either general, I.A., or specialized surgical procedures. There are movable components and accessories intended for use during diagnostic examinations or surgical procedures to support and position a patient.
Customers may be storing objects on the base and/or around the column of the surgical table. This prohibited practice and misuse known by the customers can damage the override switch assembly located at the top of the column cover. Depending upon the nature and severity of the damage to the electronics of the table, it is possible that the table could move without prompting by an operator .
B420702032 to 0402110097
Worldwide Distribution - US Nationwide including District of Columbia and Puerto Rico and the following countries: American Samoa, Argentina, Australia, Bahrain, Bangladesh, Bermuda, Brazil, Canada, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Egypt, Germany, Greece, Guam, Hong Kong, India, Indonesia, Israel, Italy, Japa, Jordan, Kazakhstan, Korea, Kuwait, Lebanon , Libya , Malaysia , Mexico, Morocco, New Zealand, Panama, Peru, Philippines , Poland, Portugal, Qatar, Russia, Saudi Arabia , Singapore, South Africa, Spain, Sri Lanka, Taiwan, Thailand, Trinidad and Tobago, Turkey & United Kingdom.
11,044 Units