Mueller Knife Handle, Hexagonal - Product Usage: Intended to hold surgical blades.
Recall: Z-2400-2020
·
Reported June 24, 2020
Enforcement
- Recall Number
- Z-2400-2020
- Event ID
- 85747
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Carefusion 2200 Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 24, 2020
- Initiation Date
- May 18, 2020
- Classification Date
- June 18, 2020
- Termination Date
- April 25, 2021
- Address
- 75 N Fairway Dr, Vernon Hills, IL, 60061-1845, United States
Description
Mueller Knife Handle, Hexagonal - Product Usage: Intended to hold surgical blades.
Reason
Becton Dickinson and Company, is conducting a voluntary recall on the V. MuellerTM 3H Knife Handles catalog number listed in the table above. BD has been informed by our supplier that the external threads on some of the collets (Figure 1) contain a machining burr, potentially causing the collet to not fully seat in the knife handle (Figure 2), inhibiting complete grip of the inserted knife blade possibly resulting in a loose blade.
Code Info
Model Numbers: OP0702 and OP0703. Model OP0702, Lot C20VAA ( 4 inch length) Model OP0703, Lot D17VAA (6 inch length)
Distribution
US Nationwide distribution including in the states of CA, FL, VA and KY.
Quantity
358 units