Ingenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106
Enforcement
- Recall Number
- Z-2399-2023
- Event ID
- 92727
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Philips North America
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 16, 2023
- Initiation Date
- June 12, 2023
- Classification Date
- August 10, 2023
- Address
- 222 Jacobs St, N/A, Cambridge, MA, 02141-2296, United States
Description
Ingenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106
During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.
Product Number: 781357: UDI-DI: 00884838088108 Serial Numbers: 46119 46111 46109 46107 46104 46101 46097 46095 4609446093 46092 46089 46088 46086 46085 46084 46083 46082 46081 46078 46077 46076 46075 46071 46069 46068 46067 4606646065 46061 46060 46059 46058 46057 46055 46054 46053 46052 46050 46049 46048 46046 46045 46044 46042 46041 4603946037 46036 46034 46033 46032 46031 46030 46029 46027 46026 46025 46024 46023 46021 46020 46019 46018 46017 4601646015 46014 46013 46012 46011 46008 46007 46005 46004 46003 46002 46001 46121; Product Number 782106: UDI-DI:00884838098329 Serial Numbers: 61017 61010 46324 46315 46311 46310 46304 46289 46286 46285 46282 46270 46269 4626846265 46258 46252 46249 46247 46243 46241 46240 46239 46237 46235 46234 46233 46231 46228 46226 46225 46224 4622246220 46219 46218 46217 46215 46214 46213 46212 46211 46209 46208 46207 46206 46205 46204 46203 46202 46201 4620046242 46245 46263 46305 46264 46281
US Nationwide. Global Distribution.
601 Total