FDA Enforcement
Class II
Terminated
Femoral Nail, A/R, R1500 T2 Femur ¿12x300 mm, Product Number 18281230S
Recall: Z-2399-2018
·
Reported July 18, 2018
Enforcement
- Recall Number
- Z-2399-2018
- Event ID
- 80131
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 18, 2018
- Initiation Date
- April 27, 2018
- Classification Date
- July 10, 2018
- Termination Date
- May 21, 2020
- Address
- Bohnackerweg 1, Selzach, N/A, Switzerland
Description
Femoral Nail, A/R, R1500 T2 Femur ¿12x300 mm, Product Number 18281230S
Reason
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
Code Info
GTIN: 04546540200655; Lot code: K07AFF5
Distribution
The products were distributed to the following US states: AL, Co, FL, KY, NC, TN, UT, and WA.