AXIOM ARTIS MODULAR ANGIOGRAPHY SYSTEM, system, x-ray, angiographic AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen systems can also support the acquisition of position triggered imaging for spatial data synthesis.
Enforcement
- Recall Number
- Z-2393-2016
- Event ID
- 74716
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 17, 2016
- Initiation Date
- July 13, 2016
- Classification Date
- August 11, 2016
- Termination Date
- November 8, 2017
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418, United States
Description
AXIOM ARTIS MODULAR ANGIOGRAPHY SYSTEM, system, x-ray, angiographic AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen systems can also support the acquisition of position triggered imaging for spatial data synthesis.
The possibility exists for contaminants, in the form of biomass, to develop in the cooling system of Artis systems resulting in damage to the pump system.
Model Numbers:10094135, 10094137, 10094139, 10094141, 10094142, 10094143, 10280959, 10848280, 10848281, 10848282, 10848283, 10848353, 10848354, 10848355, 5904433, 5904441, 5904466, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555365, 7555373, 7727717, 7728350, 7728392
Nationwide Distribution to: AK,AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,WA,WI,WV,WY, Puerto Rico and Guam
11,111 worldwide, 3,187 in the USA