FDA Enforcement
Class II
Terminated
Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087
Recall: Z-2392-2016
·
Reported August 17, 2016
Enforcement
- Recall Number
- Z-2392-2016
- Event ID
- 74714
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Hitachi Medical Systems America Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 17, 2016
- Initiation Date
- June 30, 2015
- Classification Date
- August 11, 2016
- Termination Date
- November 1, 2016
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087-2371, United States
Description
Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087
Reason
The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
Code Info
Serial #'s: S5002-S5014; S5017-S5038
Distribution
US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.
Quantity
33 units