Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

Recall: Z-2392-2016 · Reported August 17, 2016

Enforcement

Recall Number
Z-2392-2016
Event ID
74714
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hitachi Medical Systems America Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 17, 2016
Initiation Date
June 30, 2015
Classification Date
August 11, 2016
Termination Date
November 1, 2016
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087-2371, United States

Description

Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

Reason

The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.

Code Info

Serial #'s: S5002-S5014; S5017-S5038

Distribution

US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.

Quantity

33 units