FDA Enforcement Class I Ongoing

Description/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHILIC/ASK-00520-JH; CATH S-L 20 GA X 5 CM HYDROPHILIC/ASK-00520-UHC; ARTERIAL LINE KIT: 20 GA x 5IN/ASK-00820-CL; RA CATH KIT: 20 GA X 8 CM X 12 CM/ASK-00820-WC; ARTERIAL CATHETERIZATION KIT/ASK-01218-HM; SAC KIT: 20 GA X 12CM/ASK-01220-HMH; ARTERIAL CATH KIT: 20GA X 12CM/ASK-01220-NWM; ARTERIAL CATH KIT: 20 GA X 12 CM/ASK-01220-UH; ARTERIAL CATH: 18 GA X 6 1/2IN/ASK-01618-AU; RA CATH KIT: 18 GA X 6 IN/ASK-01618-CL; ARTERIAL CATH KIT: 18 GA X 6 1/2 IN/ASK-01618-JH; ARTERIAL CATHETERIZATION KIT/ASK-01618-PH; ARTERIAL CATH KIT: 18 GA X 6 1/2IN/ASK-01618-RU; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018-HF; ARTERIAL CATH KIT: 18 GA X 6 IN (16 CM)/ASK-04018-NS; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-ST; RA CATH KIT: 20 GA X 1 1/2 IN (3.81 CM)/ASK-04120-BGH; ART/VESS CATH KIT:20GAX4.45CM/20GAX12CM/ASK-04500-SM; ARTERIAL LINE KIT/ASK-04500-UCH; ARTERIAL LINE KIT/ASK-04510-EUH2; ARTERIAL LINE KIT: 20GA X 5IN (12CM)/ASK-04510-MS4; ARTERIAL LINE KIT/ASK-04510-UPM; ARTERIAL CATH KIT: 20GA X 5CM/SAC-00520-1A; ARTERIAL CATH KIT: 20GA X 5CM/SAC-00520-X1A; ARTERIAL CATH KIT: 22GA X 5CM/SAC-00522-1A ARTERIAL CATH KIT: 22 GA X 5CM/SAC-00522-X1A; ARTERIAL CATH KIT: 20GA X 8CM/SAC-00820-1A; ARTERIAL CATH KIT: 20GA X 8CM/SAC-00820-X1A; SAC KIT: 18 GA X 12CM/SAC-01218-1A; SAC KIT: 18 GA X 12CM/SAC-01218-X1A; SAC KIT: 20 GA X 12CM/SAC-01220-1A; SAC KIT: 20 GA X 12CM/SAC-01220-X1A; SAC KIT: 18 GA X 16CM/SAC-01618-1A; SAC KIT: 18 GA X 16CM/SAC-01618-X1A; ACCESS TRAY/UM-04018; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-ABM; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-BGH; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-LOL; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-MHS; RA CATH KIT: 20 GA X 1 3/4IN (4.45 CM)/ASK-04020-MIB3; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-PMC1; RA CATH SET: 20 GA X 1 3/4 IN/ASK-04020-URMC; RA CATH KIT: 20 GA X 1 1/2 IN (3.81 CM)/ASK-04220-FMH1; RA CATH KIT: 20 GA X 4.45 CM/NA-04020-1A; RA CATH KIT: 20 GA X 4.45 CM W BIO/NA-04020-B; RA CATH KIT: 20 GA X 4.45 CM W CHG/NA-04020-C; VESSEL CATH KIT: 20 GA X 5IN (12 CM)/ASK-04550-NYP; CVC KIT: 16 GA X 8 IN (20 CM)/ASK-04300-PHP1; CVC KIT: 16 GA X 8 IN (20 CM)/ASK-04301-TJ; ARTERIAL ACCESS KIT/ASK-04001-VAMC1; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-WBH

Recall: Z-2391-2026 · Reported June 24, 2026

Enforcement

Recall Number
Z-2391-2026
Event ID
98793
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
ARROW INTERNATIONAL, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 24, 2026
Initiation Date
April 21, 2026
Classification Date
June 12, 2026
Address
3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437, United States

Description

Description/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHILIC/ASK-00520-JH; CATH S-L 20 GA X 5 CM HYDROPHILIC/ASK-00520-UHC; ARTERIAL LINE KIT: 20 GA x 5IN/ASK-00820-CL; RA CATH KIT: 20 GA X 8 CM X 12 CM/ASK-00820-WC; ARTERIAL CATHETERIZATION KIT/ASK-01218-HM; SAC KIT: 20 GA X 12CM/ASK-01220-HMH; ARTERIAL CATH KIT: 20GA X 12CM/ASK-01220-NWM; ARTERIAL CATH KIT: 20 GA X 12 CM/ASK-01220-UH; ARTERIAL CATH: 18 GA X 6 1/2IN/ASK-01618-AU; RA CATH KIT: 18 GA X 6 IN/ASK-01618-CL; ARTERIAL CATH KIT: 18 GA X 6 1/2 IN/ASK-01618-JH; ARTERIAL CATHETERIZATION KIT/ASK-01618-PH; ARTERIAL CATH KIT: 18 GA X 6 1/2IN/ASK-01618-RU; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018-HF; ARTERIAL CATH KIT: 18 GA X 6 IN (16 CM)/ASK-04018-NS; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-ST; RA CATH KIT: 20 GA X 1 1/2 IN (3.81 CM)/ASK-04120-BGH; ART/VESS CATH KIT:20GAX4.45CM/20GAX12CM/ASK-04500-SM; ARTERIAL LINE KIT/ASK-04500-UCH; ARTERIAL LINE KIT/ASK-04510-EUH2; ARTERIAL LINE KIT: 20GA X 5IN (12CM)/ASK-04510-MS4; ARTERIAL LINE KIT/ASK-04510-UPM; ARTERIAL CATH KIT: 20GA X 5CM/SAC-00520-1A; ARTERIAL CATH KIT: 20GA X 5CM/SAC-00520-X1A; ARTERIAL CATH KIT: 22GA X 5CM/SAC-00522-1A ARTERIAL CATH KIT: 22 GA X 5CM/SAC-00522-X1A; ARTERIAL CATH KIT: 20GA X 8CM/SAC-00820-1A; ARTERIAL CATH KIT: 20GA X 8CM/SAC-00820-X1A; SAC KIT: 18 GA X 12CM/SAC-01218-1A; SAC KIT: 18 GA X 12CM/SAC-01218-X1A; SAC KIT: 20 GA X 12CM/SAC-01220-1A; SAC KIT: 20 GA X 12CM/SAC-01220-X1A; SAC KIT: 18 GA X 16CM/SAC-01618-1A; SAC KIT: 18 GA X 16CM/SAC-01618-X1A; ACCESS TRAY/UM-04018; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-ABM; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-BGH; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-LOL; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-MHS; RA CATH KIT: 20 GA X 1 3/4IN (4.45 CM)/ASK-04020-MIB3; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-PMC1; RA CATH SET: 20 GA X 1 3/4 IN/ASK-04020-URMC; RA CATH KIT: 20 GA X 1 1/2 IN (3.81 CM)/ASK-04220-FMH1; RA CATH KIT: 20 GA X 4.45 CM/NA-04020-1A; RA CATH KIT: 20 GA X 4.45 CM W BIO/NA-04020-B; RA CATH KIT: 20 GA X 4.45 CM W CHG/NA-04020-C; VESSEL CATH KIT: 20 GA X 5IN (12 CM)/ASK-04550-NYP; CVC KIT: 16 GA X 8 IN (20 CM)/ASK-04300-PHP1; CVC KIT: 16 GA X 8 IN (20 CM)/ASK-04301-TJ; ARTERIAL ACCESS KIT/ASK-04001-VAMC1; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-WBH

Reason

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Code Info

REF(Material)/UDI-DI/Lot(Batch): AK-04020-SPC/10801902121551, 10801902193275/33F23M0171, 33F24C0729, 33F24E0699, 33F24G0596, 33F24H0057, 33F24J0350, 33F24J0458, 33F24K0615, 33F24M0126, 33F25A0338, 33F25B0039, 33F25B0131, 33F25D0192, 33F25D0281, 33F25E1109, 33F25F0496, 33F25J0764, 33F26B0054; ASK-04020-MW/10801902142983, 10801902193275/33F23M0188, 33F24B0447, 33F24C0744, 33F24E0725, 33F24G0231, 33F24G0669, 33F24H0344, 33F24L0005, 33F24L0024, 33F25A0062, 33F25B0064, 33F25D0290, 33F25D0729, 33F25E0428, 33F25H0328, 33F26B0089, 33F26B0263; ASK-04510-MM5/10801902193275/33F24K0360, 33F25A0237, 33F25B0546; ASK-00520-JH/10801902138788, 10801902193275/33F24D0201, 33F24F0686, 33F24J0128, 33F24K0373, 33F25A0204, 33F25B0541, 33F25D0324, 33F25D0645, 33F25H0429, 33F25H0507, 33F25J0195, 33F26A1380, 33F26B0159, 33F26C0192; ASK-00520-UHC/10801902139839, 10801902193275/33F23L0771, 33F24A0775, 33F24B0219, 33F24C0125, 33F24D0093, 33F24E0383, 33F24G0386, 33F24J0284, 33F24K0149, 33F24L0712, 33F25A0495, 33F25C0125, 33F25C0826, 33F25E0099, 33F25H0047, 33F25J0851, 33F25L0392, 33F26B0540; ASK-00820-CL/10801902193275, 10801902193916/33F24G0275, 33F24J0285, 33F24K0152, 33F24L0590, 33F25A0497, 33F25C0797, 33F25E0102, 33F25H0050, 33F25J0853, 33F25L0385, 33F26B0536; ASK-00820-WC/10801902193275/33F24G0150, 33F24G0227, 33F24G0691, 33F24H0253, 33F24K0624, 33F25B0091, 33F25B0334, 33F25J0574, 33F26B0113; ASK-01218-HM/10801902193275/33F24G0276, 33F24K0551, 33F24L0215, 33F25A0296, 33F25B0543, 33F25E0450, 33F25F0364, 33F25G0163, 33F25G0208, 33F25H0499, 33F25J0100, 33F25J0139; ASK-01220-HMH/10801902193275/33F25B0245, 33F25B0544, 33F25C0698, 33F25D0388; ASK-01220-NWM/10801902193275/33F24H0181, 33F24J0154, 33F24K0497, 33F25F0157, 33F25H0658; ASK-01220-UH/10801902193275/33F24G0391, 33F24J0286, 33F24J0541, 33F24K0154, 33F24L0663, 33F25C0126, 33F25E0104, 33F25E0955, 33F25H0052, 33F25J0494, 33F25J0855, 33F25L0459, 33F25M0952; ASK-01618-AU/10801902205909, 10801902193275/33F24F0213, 33F24K0673, 33F25C0653, 33F25E1171, 33F25L0062, 33F26A0233; ASK-01618-CL/10801902195934, 10801902193275/33F24E0648, 33F24K0355, 33F25A0224,33F25B0493, 33F25D0327, 33F25E0588, 33F25F0158, 33F25J0140, 33F25M0112, 33F26C0366; ASK-01618-JH/10801902138801, 10801902193275/33F24D0584, 33F24K0490, 33F24L0844, 33F25B0494, 33F25D0530, 33F25H0834, 33F25K0264, 33F25K0357, 33S25D0530; ASK-01618-PH/10801902193275/33F24K0749, 33F24L0837, 33F25C0251, 33F25E0019, 33F25G0220; ASK-01618-RU/10801902202304, 10801902193275/33F24B0869, 33F24D1027, 33F25A0102, 33F25C0713, 33F25J0792, 33F25M0618; ASK-04018-HF/10801902193275/33F24H0156, 33F24L0397, 33F25D0391, 33F25G0401, 33F25K0069, 33F25M0142; ASK-04018-NS/10801902193275/33F25E1176, 33F25G0103; ASK-04018-ST/10801902117639, 10801902193275/33F24A0112, 33F24A0826, 33F24B0237, 33F24C0140, 33F24E0402, 33F24E0403, 33F24G0281, 33F24K0163, 33F25A0504, 33F25C0780, 33F25E0115, 33F25G0104, 33F25J0498, 33F25M0917, 33F26B1011; ASK-04120-BGH/10801902117936, 10801902193275/33F24D0518, 33F24G0262, 33F24J0474, 33F25A0150, 33F25C0740, 33F25E1207, 33F25H0096, 33F25K0331; ASK-04500-SM/30801902118012, 10801902193275/33F24A0117, 33F24A1163, 33F24D0059, 33F24D0968, 33F24E0782, 33F24G0263, 33F24J0302, 33F25C0500, 33F25E0162, 33F25H0738, 33F25L0209, 33F25M0617, 33F26B0020, 33F26B0746; ASK-04500-UCH/10801902139464, 10801902193275/33F24E0454, 33F24G0446, 33F24J0588, 33F24K0207, 33F24L0715, 33F25C0155, 33F25C0924, 33F25E0163, 33F25H0739, 33F25L0332, 33F26B0655; ASK-04510-EUH2/10801902193275, 10801902213386/33F24J0303, 33F24L0728, 33F25C0156, 33F25C0803, 33F25E0164, 33F25H0103, 33F25J0522, 33F25L0343, 33F25M0850, 33F26B0660; ASK-04510-MS4/10801902193275, 10801902139709/33F24K0212, 33F24L0764, 33F25C0860, 33F25E0169, 33F25E0994, 33F25J0526, 33F25L0423, 33F26B0654; ASK-04510-UPM/10801902118087, 10801902193275/33F24B0144, 33F24E0101, 33F24E0102, 33F24G0456, 33F24J0595, 33F24K0216, 33F24L0675, 33F25A0544, 33F25C0947, 33F25E0174, 33F25H0107, 33F25J0529, 33F25J0738, 33F25L0022, 33F26B0647, 33F26B0665, 33F26C0617; SAC-00520-1A/10801902140644, 10801902193275/33F24C0780, 33F24E0772, 33F24G0025, 33F24G0675, 33F24H0061, 33F24J0190, 33F24J0207, 33F24J0415, 33F24M0010, 33F25A0097, 33F25B0052, 33F25B0126, 33F25B0410, 33F25C0540, 33F25D0185, 33F25E0408, 33F25E0472, 33F25G0530, 33F25H0832, 33F25J0045, 33F25J1013, 33F25M0704, 33F26B0259; SAC-00520-X1A/10801902193275, 10801902140583/33F24J0208, 33F25A0098, 33F25B0023, 33F25B0127, 33F25B0287, 33F25D0086, 33F25E1180, 33F25F0288, 33F25F0425, 33F25H0347, 33F25H0866, 33F25L0177, 33F26B0058, 33F26B0279; SAC-00522-1A/10801902140606, 10801902193275/33F23L0976, 33F24B0461, 33F24C0783, 33F24E0775, 33F24G0243, 33F24J0242, 33F24J0416, 33F24M0122, 33F25A0389, 33F25B0076, 33F25E0382, 33F25H0348, 33F26B0257; SAC-00522-X1A/10801902193275/33F24L0536, 33F26B0059; SAC-00820-1A/10801902140637, 10801902193275/33F24B0463, 33F24C0785, 33F24E0777, 33F24F0081, 33F24G0136, 33F24H0059, 33F24H0352, 33F24H0388, 33F24L0057, 33F24M0011, 33F24M0123, 33F25A0134, 33F25B0109, 33F25B0282, 33F25B0646, 33F25C0543, 33F25C0568, 33F25D0094, 33F25D0186, 33F25D0302, 33F25E0364, 33F25F0045, 33F25J0581, 33F25J0980, 33F25K0298, 33F25K0338; SAC-00820-X1A/10801902193275, 10801902140620/33F25C0491, 33F25C0544, 33F25D0073, 33F25E0409, 33F25F0055, 33F25H0801, 33F25J0660, 33F25K0337, 33F25L0193, 33F25M0697, 33F26B0068, 33F26B0702; SAC-01218-1A/10801902199734, 10801902193275/33F24A1699, 33F24D0559, 33F24E0668, 33F24F1031, 33F25A0203, 33F25C0455, 33F25D0580, 33F25J0189, 33F26C0035; SAC-01218-X1A/10801902193275, 10801902199741/33F24J0138, 33F24L0960, 33F25A0294, 33F25B0537, 33F25C0479, 33F25E0768, 33F25H0618, 33F25J0451, 33F26A1374, 33F26B0775; SAC-01220-1A/10801902199758, 10801902193275/33F24C0971, 33F24D0361, 33F24F0986, 33F24L0509, 33F25A0221, 33F25B0481, 33F25C0374, 33F25D0635, 33F25E0711, 33F25G0297, 33F25J0133, 33F26A1362, 33F26B0458, 33F26C0680; SAC-01220-X1A/10801902193275, 10801902140989/33F24J0163, 33F24L0148, 33F25A0194, 33F25B0470, 33F25B0680, 33F25D0636, 33F25E0841, 33F25G0411, 33F25J0259, 33F26A1449, 33F26B0202, 33F26C0168; SAC-01618-1A/10801902199765, 10801902193275/33F24A0983, 33F24A1565, 33F24C0373, 33F24C1010, 33F24D0323, 33F24D0875, 33F24F0992, 33F24K0364, 33F24L0394, 33F24L1126, 33F25A0295, 33F25B0510, 33F25C0250, 33F25D0258, 33F25D0385, 33F25D0581, 33F25E0712, 33F25E0903, 33F25F0231, 33F25G0402, 33F25J0123, 33F25J0340, 33F26A1356, 33F26A1509, 33F26B0941, 33F26D0311; SAC-01618-X1A/10801902193275, 10801902145489/33F24J0144, 33F24L0395, 33F25A0222, 33F25B0511, 33F25D0259, 33F25D0321, 33F25D0458, 33F25E0770, 33F25F0232, 33F25G0264, 33F25G0383, 33F25J0134, 33F25J0452, 33F26A1435, 33F26B0842, 33F26B0939, 33F26B0940; UM-04018/10801902192063, 10801902193275/33F24A0359, 33F24A0360, 33F24A0361, 33F24A1061, 33F24A1062, 33F24B0162, 33F24B0163, 33F24B0164, 33F24C0053, 33F24D0042, 33F24D0043, 33F24H0463, 33F24H0464, 33F24J0466, 33F24K0306, 33F24K0311, 33F24K0543, 33F24L0723, 33F24L0761, 33F25A0608, 33F25C0209, 33F25E0284, 33F25E0297, 33F25E0742, 33F25E1061, 33F25E1069; ASK-04020-ABM/10801902193275, 10801902218947/33F24J0341, 33F24J0551, 33F24L0700, 33F25C0923, 33F25E0117, 33F25J0500, 33F25L0468, 33F26B0995; ASK-04020-BGH/10801902117660, 10801902193275/33F24A0814, 33F24C0142, 33F24D0109, 33F24E0406, 33F24G0404, 33F24J0554, 33F24L0634, 33F25A0507, 33F25C0809, 33F25E0120, 33F25H0064, 33F25J0867, 33F25L0386, 33F25M0890; ASK-04020-LOL/10801902193275, 10801902144765/33F24G0417, 33F24J0566, 33F24K0180, 33F24L0780, 33F25C0993, 33F25H0077, 33F25H0733, 33F25J0670, 33F25L0425, 33F26C0318, 33S24B0254, 33S24C0155; ASK-04020-MHS/20801902117711, 10801902193275/33F24A0380, 33F24A1033, 33F24B0255, 33F24C0157, 33F24D0121, 33F24E0420, 33F24H0476, 33F24J0568, 33F24K0181, 33F25C0990, 33F25E0137, 33F25G0111, 33F25H0078, 33F25J0505, 33F25L0377, 33F26B0526, 33S24A1033, 33S24B0255, 33S24C0157, 33S24D0121, 33X24K0003; ASK-04020-MIB3/10801902193275/33F24L0687, 33F25C0859, 33F25E0981, 33F25J0878, 33F26A0967; ASK-04020-PMC1/10801902193275/33F25C0489; ASK-04020-URMC/10801902117868, 10801902193275/33F24A0142, 33F24A0870, 33F24B0274, 33F24C0173, 33F24D0136, 33F24E0442, 33F24G0436, 33F24L0644, 33F25A0530, 33F25E0151, 33F25H0091, 33F25J0888; ASK-04220-FMH1/10801902193275/33F24K0203, 33F25A0535, 33F25E0158, 33F25J0890; NA-04020-1A/10801902119916, 10801902193275/33F24A1057, 33F24B0153, 33F24C0045, 33F24D0039, 33F24G0549, 33F24J0333, 33F24K0293, 33F24L0678, 33F25A0597, 33F25C0900, 33F25E0267, 33F25E1047, 33F25G0150, 33F25H0174, 33F25H0666, 33F25J0664, 33F25K0403, 33F25L0023, 33F25L0436, 33F25M0936, 33F26B0577; NA-04020-B/10801902193275, 10801902221312/33F24L0072, 33F25A0598, 33F25C0763, 33F25E0268, 33F25E1048, 33F25H0175, 33F25H0667, 33F25J0632, 33F26B0530; NA-04020-C/10801902193275, 10801902221329/33F24L0073, 33F25A0599, 33F25C0926, 33F25E0269, 33F25E1049, 33F25G0151, 33F25H0176, 33F25H0668, 33F26B0529; ASK-04550-NYP/10801902130836, 10801902193275/33F24F0685, 33F24H0144, 33F24K0424, 33F25A0166, 33F25B0170, 33F25B0495, 33F25D0337, 33F25G0355, 33F25H0431, 33F25H0639, 33F26A1367, 33F26A1542, 33F26C0585; ASK-04300-PHP1/10801902143805, 10801902193275/33F24F0619, 33F24H0147, 33F24K0072, 33F24M0047, 33F25C0649, 33F25K0314, 33F25L0029, 33F25M0066; ASK-04301-TJ/20801902114987, 10801902193275/33F24D0733, 33F24F0798, 33F25A0235, 33F25B0516; ASK-04001-VAMC1/10801902158854, 10801902193275/33F24E0400, 33F24G0400, 33F24K0161, 33F25A0502, 33F25E0114, 33F25J0863, 33F25L0387, 33F26B1016; ASK-04020-WBH/10801902117899/33F24C0174, 33F24D0138, 33F24E0445, 33F24G0438, 33F24J0582, 33F24L0793;

Distribution

US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.

Quantity

491,489