FDA Enforcement Class I Ongoing

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637HP

Recall: Z-2391-2023 · Reported August 23, 2023

Enforcement

Recall Number
Z-2391-2023
Event ID
92660
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Covidien LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 23, 2023
Initiation Date
June 28, 2023
Classification Date
August 11, 2023
Address
15 Hampshire St, Mansfield, MA, 02048-1113, United States

Description

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637HP

Reason

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Code Info

GTIN: 10884521128187 / 20884521128184 LOT Numbers: 1916500156 2017400085 2028300088 2102600094 2104600149 2107700145 2113100057 2119400230 2123000100 2133700164

Distribution

US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom

Quantity

3965 units