FDA Enforcement Class II Terminated

Norton Arm Kit, Product Code 711400, (previous product code 031001), Sample Kit S711400

Recall: Z-2390-2021 · Reported September 8, 2021

Enforcement

Recall Number
Z-2390-2021
Event ID
88476
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Aspen Surgical Products, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 8, 2021
Initiation Date
July 23, 2021
Classification Date
August 30, 2021
Termination Date
July 26, 2022
Address
6945 Southbelt Dr SE, N/A, Caledonia, MI, 49316-7664, United States

Description

Norton Arm Kit, Product Code 711400, (previous product code 031001), Sample Kit S711400

Reason

Affected products were manufactured with materials found to have a bioburden level exceeding the limit at which the product's sterilization process was validated. As a result, the firm cannot provide full assurance that the affected products were effectively sterilized, and the product may compromise a sterile field and increase the risk of a surgical site infection.

Code Info

All lots manufactured between July 9, 2019 to July 9, 2021

Distribution

US nationwide distribution. No foreign distribution.

Quantity

124