MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit REF: 8888345629HP
Enforcement
- Recall Number
- Z-2389-2023
- Event ID
- 92660
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Covidien LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 23, 2023
- Initiation Date
- June 28, 2023
- Classification Date
- August 11, 2023
- Address
- 15 Hampshire St, Mansfield, MA, 02048-1113, United States
Description
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit REF: 8888345629HP
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
GTIN: 0884521128163/ 20884521128160 LOT Numbers: Lot Number 1828200086 1925300170 1925300171 2018800013 2018800024 2018800043 2019500199 2019500200 2019500222 2021000106 2021000107 2029600102 2029600103 2116600083 2131200176 2133700166 2227800108 2230400250 2230400251 2234800071
US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
15209 units