FDA Enforcement Class I Ongoing

Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 CM/ASK-15123-UOC1, AHDC KIT: 3-L 12 FR X 20 CM/ASK-15123-UPM, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-P1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XP1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPCN1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-P1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPCN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPN1A, AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XPCN1A, HEMODIALYSIS KIT 2L: 12 FR X 20CM AGB/AK-25122-CDC, HEMODIALYSIS KIT: 2-LUMEN 14 FR X 20 CM/AK-25142-F, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM1, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM2, AHDC KIT: 2L 14 FR X 15 CM/ASK-22142-UOC1, AHDC AGB KIT: 2-L 14 FR X 15 CM/ASK-22142-UPM, HEMODIALYSIS KIT: 2L 14 FR X 20 CM/ASK-25142-UNC1, AHDC KIT: 2L 14 FR X 20 CM/ASK-25142-UOC1, AHDC AGB KIT: 2L 14 FR X 20 CM/ASK-25142-UPM, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/ASK-26142-UNC1, HEMODIALYSIS KIT: 2LUMEN 12 FR X 16 CM/AU-22122-F, HEMODIALYSIS KIT: 2L 12 FR X 16 CM AGB/CDC-22122-1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A-U, HEMODIALYSIS KIT 2L: 14 FR X 15 CM AGB/CDC-22142-1A, HEMODIALYSIS KIT: 2L 14 FR X 15 CM AGB/CDC-22142-XC1A, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A-C, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XCN1A-U, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XN1A-U, HEMODIALYSIS KIT: 2L 12 FR X 13 CM AGB/CDC-23122-XU1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-X1A, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XN1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-XU1A, HEMODIALYSIS KIT 2L: 14 FR X 20 CM AGB/CDC-25142-1A, HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB/CDC-25142-XC1A, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A-C, HEMODIALYSIS KIT: 2L 12 FR X 25 CM AGB/CDC-26122-1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XCN1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XN1A, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/CDC-26142-1A, HEMODIALYSIS KIT: 2L 14 FR X 25 CM AGB/CDC-26142-X1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XCN1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XN1A, HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-25122-F, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 16 CM/CDC-22123-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 20 CM/CDC-25123-XCN1A, LBCVC AGB KIT:3-L 12 FR X 20 CM/CDC-25123-XN1A

Recall: Z-2388-2026 · Reported June 24, 2026

Enforcement

Recall Number
Z-2388-2026
Event ID
98793
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
ARROW INTERNATIONAL, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 24, 2026
Initiation Date
April 21, 2026
Classification Date
June 12, 2026
Address
3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437, United States

Description

Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 CM/ASK-15123-UOC1, AHDC KIT: 3-L 12 FR X 20 CM/ASK-15123-UPM, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-P1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XP1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPCN1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-P1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPCN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPN1A, AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XPCN1A, HEMODIALYSIS KIT 2L: 12 FR X 20CM AGB/AK-25122-CDC, HEMODIALYSIS KIT: 2-LUMEN 14 FR X 20 CM/AK-25142-F, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM1, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM2, AHDC KIT: 2L 14 FR X 15 CM/ASK-22142-UOC1, AHDC AGB KIT: 2-L 14 FR X 15 CM/ASK-22142-UPM, HEMODIALYSIS KIT: 2L 14 FR X 20 CM/ASK-25142-UNC1, AHDC KIT: 2L 14 FR X 20 CM/ASK-25142-UOC1, AHDC AGB KIT: 2L 14 FR X 20 CM/ASK-25142-UPM, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/ASK-26142-UNC1, HEMODIALYSIS KIT: 2LUMEN 12 FR X 16 CM/AU-22122-F, HEMODIALYSIS KIT: 2L 12 FR X 16 CM AGB/CDC-22122-1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A-U, HEMODIALYSIS KIT 2L: 14 FR X 15 CM AGB/CDC-22142-1A, HEMODIALYSIS KIT: 2L 14 FR X 15 CM AGB/CDC-22142-XC1A, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A-C, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XCN1A-U, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XN1A-U, HEMODIALYSIS KIT: 2L 12 FR X 13 CM AGB/CDC-23122-XU1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-X1A, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XN1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-XU1A, HEMODIALYSIS KIT 2L: 14 FR X 20 CM AGB/CDC-25142-1A, HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB/CDC-25142-XC1A, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A-C, HEMODIALYSIS KIT: 2L 12 FR X 25 CM AGB/CDC-26122-1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XCN1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XN1A, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/CDC-26142-1A, HEMODIALYSIS KIT: 2L 14 FR X 25 CM AGB/CDC-26142-X1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XCN1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XN1A, HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-25122-F, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 16 CM/CDC-22123-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 20 CM/CDC-25123-XCN1A, LBCVC AGB KIT:3-L 12 FR X 20 CM/CDC-25123-XN1A

Reason

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Code Info

REF(Material)/UDI-DI/Lot(Batch): AK-15122-F/10801902197136, 10801902193275, 10801902168525/33F24E0348, 33F24J0280, 33F24K0119, 33F24L0601, 33F25A0477, 33F25C0908, 33F24A0850; ASK-12123-UOC1/10801902193275/33F25B0430, 33F25D0119; ASK-12123-UPM/10801902193275, 10801902222463/33F25E0535, 33F25G0188, 33F25G0189, 33F26C0333, 33F25D0262; ASK-15123-UOC1/10801902193275/33F25C0378, 33F25D0121; ASK-15123-UPM/10801902193275, 10801902159950/33F25E0718, 33F25H0447, 33F26A1389, 33F24E0649; CDC-12123-P1A/40801902111768, 10801902193275/33F24D0632, 33F24F0758, 33F24L0465, 33F25A0281, 33F25E0794; CDC-12123-XP1A/10801902193275, 30801902111785, 10801902213416/33F25J0723, 33F26A1382, 33F26B0433, 36F25A0007, 36F25B0009, 36F25E0018, 33F24D0299; CDC-12123-XPCN1A/10801902213416, 10801902193275/33F24D0950, 33F24K0417, 33F24L0466, 33F25A0216, 33F25B0667, 33F25D0436, 33F25D0666, 33F25E0693, 33F25G0239, 33F25J0247, 33F26A1556, 33F26B0801, 33F26C0767, 33F25B0668; CDC-12123-XPN1A/10801902193275, 10801902213409/33F25D0554, 33F25G0240, 33F25J0312, 33F26B0214,33F26C0199, 33F25A0181; CDC-15123-P1A/10801902193275/33F25D0437, 33F25H0547,33F25B0613; CDC-15123-XPCN1A/10801902213430, 10801902193275/33F24F0107, 33F24F0998, 33F24K0403, 33F24L0834, 33F25A0283, 33F25B0527, 33F25E0650, 33F25E0745, 33F25F0085, 33F25H0596, 33F25J0314, 33F26A1464, 33F26B0883, 33F25B0669; CDC-15123-XPN1A/0801902193275, 10801902213423, 10801902116793/33F25D0439, 33F25G0238, 33F25H0597, 33F25J0315, 33F26A1468, 33F26B0826, 33F24C0461; CDC-16123-XPCN1A/10801902193275/33F25B0671, 33F25D0440, 33F25E0651, 33F25G0236, 33F25H0529, 33F25J0316, 33F25G0237, AK-25122-CDC/10801902193275/33F24G0370, 33F24K0135, 33F25C0120, 33F25C0969, 33F25E0942, 33F24G0371; AK-25142-F/10801902193275, 70801902123205, 10801902167795/33F25C0863, 33F25E0087, 33F25H0038, 33F25J0846, 33F26B0531, 33F24D0089; ASK-22122-MM1/10801902193275/33F25E0546; ASK-22122-MM2/10801902193275, 10801902221923, 10801902144956/33F25G0172, 33F25L0031, 33F25M0310, 33F26A1652, 33F26B0151, 33F26B0450, 33F24F0854; ASK-22142-UOC1/10801902193275/33F25C0545, 33F25D0122; ASK-22142-UPM/10801902193275, 10801902222487, 10801902202281/33F25D0594, 33F25F0116, 33F25G0187, 33F26A1390, 33F26B0205, 33F24D0396; ASK-25142-UNC1/10801902193275/33F25C0655, 33F24K0561; ASK-25142-UOC1/10801902193275/33F25B0362, 33F25D0123; ASK-25142-UPM/10801902193275/33F25E0595, 33F25H0446, 33F25J0381, 33F26A1446, 33F25J0092; ASK-26142-UNC1/10801902193275/33F25B0427, 33F25K0058; AU-22122-F/10801902193275/33F24K0269, 33F25A0584, 33F25C0862, 33F25E0245, 33F25J0567; CDC-22122-1A/20801902123484, 10801902193275/, CDC-22122-XCN1A-U/33F24F0996, 33F24L0374, 33F25C0461; CDC-22122-XN1A/10801902193275/33F25C0239, 33F25E0875, 33F25J0419; CDC-22122-XN1A-U/10801902193275, 10801902213621/33F25A0309, 33F25E0876, 33F26B0217; CDC-22142-1A/50801902123515, 10801902193275/33F24D0928, 33F24F0103, 33F24F0896, 33F24L0111, 33F24L0955, 33F25B0500, 33F25D0447, 33F25E0654, 33F25G0278, 33F25H0577, 33F25J0319; CDC-22142-XC1A/10801902193275/33F25A0284, 33F25J0180; CDC-22142-XCN1A/10801902193275, 10801902213539/33F25A0285, 33F25B0529, 33F25D0449, 33F25E0655, 33F25E0818, 33F25F0214, 33F25G0233, 33F26A1477, 33F26B0898, 33R25J0455; CDC-22142-XCN1A-C/10801902193275/33F24L0473, 33F25B0615, 33F25F0213; CDC-22142-XN1A/10801902193275, 10801902213522/33F25C0240, 33F25E0578, 33F25F0215, 33F25H0528, 33F25J0321, 33F26A1383, 33F26B0215, 33F26C0016; CDC-22142-XN1A-C/10801902193275/33F25B0485, 33F25E0879; CDC-23122-XCN1A-U/10801902193275/33F24L0279, 33F25D0492; CDC-23122-XN1A-U/10801902193275/33F25E0579; CDC-23122-XU1A/10801902193275/33F25D0493, 33F25H0432, 33F25J0248; CDC-25122-1A/10801902193275/33F25B0619, 33F25E0881; CDC-25122-X1A/10801902193275/33F25E0882, 33F25J0422; CDC-25122-XCN1A/10801902213492, 10801902193275/33F24F0929, 33F25C0242, 33F25D0498, 33F25E0883, 33F25G0234, 33F26B0439, 33F26C0214; CDC-25122-XCN1A-U/10801902193275/33F24L1041, 33F25B0620, 33F25F0121, 33F25G0226; CDC-25122-XN1A/10801902193275, 10801902213485/33F25C0243, 33F25E0884, 33F25K0183, 33R25E0884; CDC-25122-XU1A/10801902193275/33F25B0621, 33F25E0886, 33F25H0568; CDC-25142-1A/50801902123584, 10801902193275/33F24F0837, 33F24K0457, 33F24L0113, 33F24L1031, 33F25B0622, 33F25D0503, 33F25E0889, 33F25F0106, 33F25J0127; CDC-25142-XC1A/10801902193275/33F25B0467; CDC-25142-XCN1A/10801902193275, 10801902213553/33F25A0312, 33F25E0657, 33F25E0821, 33F25G0231, 33F25H0527, 33F25J0428, 33F26B0452, 33F26C0133; CDC-25142-XCN1A-C/10801902193275/33F24L0377, 33F25D0238; CDC-25142-XN1A/10801902193275, 10801902213546/33F25B0624, 33F25D0504, 33F25E0823, 33F25G0232, 33F26B0437; CDC-25142-XN1A-C/10801902193275/33F24L0485, 33F25B0676; CDC-26122-1A/60801902123611, 10801902193275/33F24F0084, 33F24K0475, 33F25G0389; CDC-26122-XCN1A/10801902213515, 10801902193275/33F24F1028, 33F24L0117, 33F25D0308, 33F25D0557, 33F25F0219; CDC-26122-XN1A/10801902193275/33F25E0451; CDC-26142-1A/10801902193275/33F24L0488, 33F24L1025, 33F25C0368, 33F25E0891, 33F25H0406, 33F25J0116; CDC-26142-X1A/10801902193275/33F25D0558, 33F25E0892; CDC-26142-XCN1A/10801902193275, 10801902213577/33F24L0941, 33F25B0530, 33F25D0507, 33F25E0699, 33F25H0525, 33F26B0396, 33F26B0895, 33F26C0780; CDC-26142-XN1A/10801902193275, 10801902213560/33F24L0927, 33F25E0580, 33F25H0526, 33F26B0393; AK-25122-F/ CDC-22122-XCN1A/10801902193275/33F24G0371, 33F24L0742, 33F25C0822, 33F25E0084, 33F25J0483, 33F25L0400; CDC-22123-XCN1A/10801902213591, 10801902193275/33F24D0842,33F24F1018, 33F24L0110, 33F25C0205, 33F25D0613, 33F25E0878, 33F26B0455; CDC-25123-XCN1A/10801902213614, 10801902193275/33F24D0818, 33F24L0483, 33F25C0244, 33F25D0502, 33F25E0888, 33F25G0228, 33F25J0425, 33F26B0395, 33F26B0894; CDC-25123-XN1A/10801902193275/33F25C0332, 33F25G0229

Distribution

US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.

Quantity

62,719