FDA Enforcement Class II Terminated

Sterile Distraction Screws of 3 different sizes: 12mm, 14mm and 16mm sold under TSI, 3COR, DB Surgical, and American Surgical Products brands and Sterile XL Screws of 7 different sizes: 90mm, 100mm, 110mm, 120mm, 130mm, 140mm, 150mm sold under TSI brand Sterile Distraction Screws are intended for the distraction in the anterior approach to the cervical spine. Sterile XL Screws are intended for the fixation of the blade to the vertebral body to prohibit drifting in the extreme lateral approach to the lumbar spine.

Recall: Z-2385-2012 · Reported September 26, 2012

Enforcement

Recall Number
Z-2385-2012
Event ID
62899
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Tedan Surgical Innovations Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 26, 2012
Initiation Date
June 11, 2012
Classification Date
September 14, 2012
Termination Date
July 31, 2013
Address
12615 W Airport Blvd, Ste 200, Sugar Land, TX, 77478-6203, United States

Description

Sterile Distraction Screws of 3 different sizes: 12mm, 14mm and 16mm sold under TSI, 3COR, DB Surgical, and American Surgical Products brands and Sterile XL Screws of 7 different sizes: 90mm, 100mm, 110mm, 120mm, 130mm, 140mm, 150mm sold under TSI brand Sterile Distraction Screws are intended for the distraction in the anterior approach to the cervical spine. Sterile XL Screws are intended for the fixation of the blade to the vertebral body to prohibit drifting in the extreme lateral approach to the lumbar spine.

Reason

This letter is to notify you that several of our sterile screw products are being voluntarily recalled. Product sterility may be compromised due to a mechanical failure of the protective packaging tube. Tube end plugs may dislodge creating the potential for the screw to come out of the protective tube and then the sharp end of the screw pierce the outer sterile barrier pouch.

Code Info

Lot numbers: 019715, 019716, 019942, 019943, 019977, 019978, 020102, 020103, 020104, 020105, 020106, 020107, 020108, 020609, 020782, 021539, 021663, 022097, 022098, 022099, 022122, 022123, 022124, 022371, 022372, 022790, 022791, 022792, 022794, 022795, 022796, 022827, 022828, and 023266.

Distribution

Worldwide distribution: USA (nationwide) and countries of : Australia and Taiwan.

Quantity

12,065 Sterile Distraction Screws (2413 boxes); 384 Sterile XL Screws (128 boxes) in distribution