Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. 1) The Varian Low Energy Linear Accelerator is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment i is indicated. (2) The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Recall: Z-2383-2012 · Reported September 26, 2012

Enforcement

Recall Number
Z-2383-2012
Event ID
63033
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Varian Medical Systems, Inc. Oncology Systems
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 26, 2012
Initiation Date
August 14, 2012
Classification Date
September 14, 2012
Termination Date
December 4, 2013
Address
911 Hansen Way, Palo Alto, CA, 94304-1028, United States

Description

Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. 1) The Varian Low Energy Linear Accelerator is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment i is indicated. (2) The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Reason

Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.

Code Info

LIST OF SERIAL NUMBERS: H140863 ,H294608 ,H294669 ,H294732 ,H140865 ,H294610 ,H294670 ,H294733 ,H140866 ,H294611 ,H294671 ,H294734 ,H140867 ,H294612 ,H294672 ,H294735 ,H140868 ,H294613 ,H294673 ,H294736 ,H140869 ,H294614 ,H294674 ,H294737 ,H140870 ,H294615 ,H294675 ,H294738 ,H140871 ,H294616 ,H294676 ,H294739 ,H140872 ,H294617 ,H294677 ,H294740 ,H140873 ,H294618 ,H294678 ,H294741 ,H140874 ,H294619 ,H294679 ,H294742 ,H140875 ,H294620 ,H294680 ,H294743 ,H140876 ,H294621 ,H294681 ,H294744 ,H140877 ,H294622 ,H294682 ,H294745 ,H140878 ,H294623 ,H294683 ,H294746 ,H140879 ,H294624 ,H294684 ,H294747 ,H140880 ,H294625 ,H294685 ,H294748 ,H140881 ,H294626 ,H294686 ,H294749 ,H140882 ,H294627 ,H294687 ,H294750 ,H140883 ,H294628 ,H294688 ,H294751 ,H140884 ,H294629 ,H294689 ,H294752 ,H140885 ,H294630 ,H294690 ,H294753 ,H140886 ,H294631 ,H294691 ,H294754 ,H140887 ,H294632 ,H294692 ,H294755 ,H140888 ,H294633 ,H294693 ,H294756 ,H140889 ,H294634 ,H294694 ,H294757 ,H140890 ,H294635 ,H294695 ,H294758 ,H140891 ,H294636 ,H294696 ,H294759 ,H140892 ,H294637 ,H294697 ,H294760 ,H140893 ,H294638 ,H294698 ,H294761 ,H140895 ,H294639 ,H294699 ,H294762 ,H140896 ,H294640 ,H294700 ,H294763 ,H140897 ,H294641 ,H294701 ,H294764 ,H140898 ,H294642 ,H294702 ,H294765 ,H140899 ,H294643 ,H294704 ,H294766 ,H140900 ,H294644 ,H294705 ,H294767 ,H181175 ,H294645 ,H294706 ,H294768 ,H181176 ,H294646 ,H294707 ,H294769 ,H181177 ,H294647 ,H294708 ,H294770 ,H181178 ,H294648 ,H294709 ,H294771 ,H181179 ,H294649 ,H294710 ,H294772 ,H181180 ,H294650 ,H294711 ,H181181 ,H294651 ,H294712 ,H181182 ,H294652 ,H294713 ,H181183 ,H294653 ,H294714 ,H181184 ,H294654 ,H294715 ,H181185 ,H294655 ,H294716 ,H181186 ,H294656 ,H294717 ,H181187 ,H294657 ,H294718 ,H181188 ,H294658 ,H294719 ,H181189 ,H294659 ,H294720 ,H181190 ,H294660 ,H294722 ,H181191 ,H294661 ,H294723 ,H181192 ,H294662 ,H294724 ,H182000 ,H294663 ,H294726 ,H182001 ,H294664 ,H294727 ,H182002 ,H294665 ,H294728 ,H182003 ,H294666 ,H294729 ,H182005 ,H294667 ,H294730 ,H294607 ,H294668 ,H294731

Distribution

Worldwide Distribution.

Quantity

221 Units