FDA Enforcement Class II Terminated

da Vinci Xi EndoWrist Stapler 45 Reload Green, Model 48445G-03, 12 reloads/box.

Recall: Z-2381-2018 · Reported July 18, 2018

Enforcement

Recall Number
Z-2381-2018
Event ID
80318
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Intuitive Surgical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 18, 2018
Initiation Date
April 23, 2018
Classification Date
July 6, 2018
Termination Date
July 15, 2020
Address
1266 Kifer Rd Bldg 101, Sunnyvale, CA, 94086-5304, United States

Description

da Vinci Xi EndoWrist Stapler 45 Reload Green, Model 48445G-03, 12 reloads/box.

Reason

Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.

Code Info

UDI: 00886874112823. Device expiration: 10/14/2017 02/21/2020

Distribution

U.S., foreign and government accounts. Foreign: Australia, Austria, Belgium, Brazil, Canada, China (Hong Kong), Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Qatar, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.

Quantity

6816