FDA Enforcement Class I Ongoing

Description/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-PMAH; 3L PI CVC KIT: 7 FR X 16 CM/ASK-12703-PNM; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-PUMH; MULTILUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-15703-PRH; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/CDC-15703-P1A; PI CVC KIT: 3-L 7 FR X 6IN (16 CM)/NA-12703-PCDC; PI CVC KIT: 3-L 7 FR X 8IN (20 CM)/NA-15703-PCDC; CVC KIT: 3-LUMEN 5.5 FR X 8 CM/AK-25553; 2-L CVC KIT: 4 FR X 5 CM/CDC-22402-X1A; 2-L CVC KIT: 5 FR X 5 CM/CDC-22502-X1A; 2-L CVC KIT: 4 FR X 8 CM/CDC-25402-X1A; 2-L CVC KIT: 5 FR X 8 CM/CDC-25502-X1A; 3-L CVC KIT: 5.5 FR X 8 CM/CDC-25553-X1A; 3-L CVC KIT: 5.5 FR X 13 CM/CDC-26553-X1A; CVC KIT 3L: 12 FR X 16 CM AGB/AK-22123-CDC; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-22123-F; CVC KIT: 2-LUMEN 4 FR X 5 CM/AK-22402; CVC KIT: 2-LUMEN 5 FR X 5 CM/AK-22502; CVC KIT: 16 GA X 8IN/AK-24301-SK; CVC KIT: 2-LUMEN 4 FR X 13 CM/AK-24402; CVC KIT: 3-LUMEN 7 FR X 30 CM/AK-24703-ASK; CVC KIT 3L: 12FR X 20 CM AGB/AK-25123-CDC; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-25123-F; CVC KIT: 2-LUMEN 4 FR X 8 CM/AK-25402; CVC KIT: 2-LUMEN 5 FR X 8 CM/AK-25502; CVC KIT: 3-LUMEN 5.5 FR X 13 CM/AK-26553; LBCVC KIT: 3L 12 FR X 16 CM AGB/CDC-22123-1A; LBCVC KIT: 3L 12 FR X 16 CM AGB/CDC-22123-X1A; CVC KIT: 16 GA X 20 CM/CDC-24301-1A; PI CVC KIT: 2L 7 FR X 16 CM AGB/AK-22702-P1A; PI CVC KIT: 2L 7 FR X 16 CM AGB/CDC-22702-XP1A; CVC KIT: 3L 7 FR X 20 CM/ASK-45703-CPR; MULTILUMEN PI CVC KIT: 3L 7 FR X 16 CM/ASK-42703-AKP1A; MULTI-LUMEN PI CVC KIT: 3-L 7 FR X 6IN/ASK-42703-KH; PI CVC KIT: 3L 7FR X 16CM AGB+/ASK-42703-MGH1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6IN/ASK-42703-PCCH2; PI CVC KIT: 3LUMEN 7 FR X 6IN (16 CM)/ASK-42703-PFS1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNO2; MULTI LUMEN PI CVC KIT: 3L 7 FR/ASK-42703-PSCH1; MULTILUMEN PI CVC KIT: 2L 8 FR X 16 CM/ASK-42802-AKP1A; CVC KIT: 2L 8 FR X 16CM/ASK-42802-LMDTG1; PI CVC KIT: 2L 8 FR X 6 IN (16 CM) AGB+/ASK-42802-PUCI1; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PUMH; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-UMM; PI CVC KIT: 4-L 8.5 FR X 6 IN/ASK-42854-AKP1A; PI CVC KIT: 4L 8.5FR X 16CM AGB+/ASK-42854-MGH; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-AKP1A; PI CVC KIT: 3L 7 FR X 20CM AGB+/ASK-45703-MGH; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PCCH2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PECL1; PI CVC KIT: 3 LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-PFS1; PI CVC KIT: 3L 7 FR X 8 IN (20 CM) AGB+/ASK-45703-PHM1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PHU; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PMSG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PNCB; MULTILUMEN CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PPCM; PI CVC KIT: 3-LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-PRH; MULTILUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PSJ2; PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PSVM; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PUHS; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PUKL; PI CVC KIT: 3L 7FR X 20CM AGB+/ASK-45703-PUMF1; PI CVC KIT: 3L 7 FR X 8 IN (20 CM)/ASK-45703-PUR1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-RHR1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-WA; MULTI LUMEN PI CVC KIT: 2L 8 FR X 8 IN/ASK-45802-AKP1A; PI CVC KIT: 4L 8.5 FR X 20 CM/ASK-45854-AKP1A; PI CVC KIT: 4L 8.5FR X 20CM AGB+/ASK-45854-MGH; PI CVC KIT: 4L 8.5 FR X 8IN/ASK-45854-PCCH2; CVC KIT: 4L 8.5 FR X 20 CM AGB/ASK-45854-PNM; MULTI LUMEN CVC KIT: 4L 8.5 FR X 8 IN/ASK-45854-PPCM; PI CVC KIT: 4L 8.5 FR X 20 CM AGB+/ASK-45854-PUCH2; MULTILUMEN PI CVC KIT: 2L 7 FR X 20 CM/ASK-46702-PWBH1; MULTILUMEN CVC KIT: 2L 8 FR X 16 CM/CDC-42802-P1A; PI CVC KIT: 3-L 7 FR X 20 CM AGB/CDC-45703-PA1A; CVC KIT: 2L 8 FR X 6IN (16 CM)/ASK-42802-ECL4; PI CVC KIT: 3-L 7 FR X 16 CM AGB+/AK-42703-P1A; PI CVC KIT: 2-L 8 FR X 16 CM AGB+/AK-42802-P1A; PI CVC KIT: 4-L 8.5 FR X 16 CM AGB+/AK-42854-P1A; PI CVC KIT: 3-L 7 FR X 20 CM AGB+/AK-45703-P1A; PI CVC KIT: 2-L 8 FR X 20 CM AGB

Recall: Z-2376-2026 · Reported June 24, 2026

Enforcement

Recall Number
Z-2376-2026
Event ID
98793
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
ARROW INTERNATIONAL, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 24, 2026
Initiation Date
April 21, 2026
Classification Date
June 12, 2026
Address
3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437, United States

Description

Description/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-PMAH; 3L PI CVC KIT: 7 FR X 16 CM/ASK-12703-PNM; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-PUMH; MULTILUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-15703-PRH; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/CDC-15703-P1A; PI CVC KIT: 3-L 7 FR X 6IN (16 CM)/NA-12703-PCDC; PI CVC KIT: 3-L 7 FR X 8IN (20 CM)/NA-15703-PCDC; CVC KIT: 3-LUMEN 5.5 FR X 8 CM/AK-25553; 2-L CVC KIT: 4 FR X 5 CM/CDC-22402-X1A; 2-L CVC KIT: 5 FR X 5 CM/CDC-22502-X1A; 2-L CVC KIT: 4 FR X 8 CM/CDC-25402-X1A; 2-L CVC KIT: 5 FR X 8 CM/CDC-25502-X1A; 3-L CVC KIT: 5.5 FR X 8 CM/CDC-25553-X1A; 3-L CVC KIT: 5.5 FR X 13 CM/CDC-26553-X1A; CVC KIT 3L: 12 FR X 16 CM AGB/AK-22123-CDC; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-22123-F; CVC KIT: 2-LUMEN 4 FR X 5 CM/AK-22402; CVC KIT: 2-LUMEN 5 FR X 5 CM/AK-22502; CVC KIT: 16 GA X 8IN/AK-24301-SK; CVC KIT: 2-LUMEN 4 FR X 13 CM/AK-24402; CVC KIT: 3-LUMEN 7 FR X 30 CM/AK-24703-ASK; CVC KIT 3L: 12FR X 20 CM AGB/AK-25123-CDC; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-25123-F; CVC KIT: 2-LUMEN 4 FR X 8 CM/AK-25402; CVC KIT: 2-LUMEN 5 FR X 8 CM/AK-25502; CVC KIT: 3-LUMEN 5.5 FR X 13 CM/AK-26553; LBCVC KIT: 3L 12 FR X 16 CM AGB/CDC-22123-1A; LBCVC KIT: 3L 12 FR X 16 CM AGB/CDC-22123-X1A; CVC KIT: 16 GA X 20 CM/CDC-24301-1A; PI CVC KIT: 2L 7 FR X 16 CM AGB/AK-22702-P1A; PI CVC KIT: 2L 7 FR X 16 CM AGB/CDC-22702-XP1A; CVC KIT: 3L 7 FR X 20 CM/ASK-45703-CPR; MULTILUMEN PI CVC KIT: 3L 7 FR X 16 CM/ASK-42703-AKP1A; MULTI-LUMEN PI CVC KIT: 3-L 7 FR X 6IN/ASK-42703-KH; PI CVC KIT: 3L 7FR X 16CM AGB+/ASK-42703-MGH1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6IN/ASK-42703-PCCH2; PI CVC KIT: 3LUMEN 7 FR X 6IN (16 CM)/ASK-42703-PFS1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNO2; MULTI LUMEN PI CVC KIT: 3L 7 FR/ASK-42703-PSCH1; MULTILUMEN PI CVC KIT: 2L 8 FR X 16 CM/ASK-42802-AKP1A; CVC KIT: 2L 8 FR X 16CM/ASK-42802-LMDTG1; PI CVC KIT: 2L 8 FR X 6 IN (16 CM) AGB+/ASK-42802-PUCI1; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PUMH; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-UMM; PI CVC KIT: 4-L 8.5 FR X 6 IN/ASK-42854-AKP1A; PI CVC KIT: 4L 8.5FR X 16CM AGB+/ASK-42854-MGH; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-AKP1A; PI CVC KIT: 3L 7 FR X 20CM AGB+/ASK-45703-MGH; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PCCH2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PECL1; PI CVC KIT: 3 LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-PFS1; PI CVC KIT: 3L 7 FR X 8 IN (20 CM) AGB+/ASK-45703-PHM1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PHU; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PMSG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PNCB; MULTILUMEN CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PPCM; PI CVC KIT: 3-LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-PRH; MULTILUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PSJ2; PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PSVM; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PUHS; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PUKL; PI CVC KIT: 3L 7FR X 20CM AGB+/ASK-45703-PUMF1; PI CVC KIT: 3L 7 FR X 8 IN (20 CM)/ASK-45703-PUR1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-RHR1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-WA; MULTI LUMEN PI CVC KIT: 2L 8 FR X 8 IN/ASK-45802-AKP1A; PI CVC KIT: 4L 8.5 FR X 20 CM/ASK-45854-AKP1A; PI CVC KIT: 4L 8.5FR X 20CM AGB+/ASK-45854-MGH; PI CVC KIT: 4L 8.5 FR X 8IN/ASK-45854-PCCH2; CVC KIT: 4L 8.5 FR X 20 CM AGB/ASK-45854-PNM; MULTI LUMEN CVC KIT: 4L 8.5 FR X 8 IN/ASK-45854-PPCM; PI CVC KIT: 4L 8.5 FR X 20 CM AGB+/ASK-45854-PUCH2; MULTILUMEN PI CVC KIT: 2L 7 FR X 20 CM/ASK-46702-PWBH1; MULTILUMEN CVC KIT: 2L 8 FR X 16 CM/CDC-42802-P1A; PI CVC KIT: 3-L 7 FR X 20 CM AGB/CDC-45703-PA1A; CVC KIT: 2L 8 FR X 6IN (16 CM)/ASK-42802-ECL4; PI CVC KIT: 3-L 7 FR X 16 CM AGB+/AK-42703-P1A; PI CVC KIT: 2-L 8 FR X 16 CM AGB+/AK-42802-P1A; PI CVC KIT: 4-L 8.5 FR X 16 CM AGB+/AK-42854-P1A; PI CVC KIT: 3-L 7 FR X 20 CM AGB+/AK-45703-P1A; PI CVC KIT: 2-L 8 FR X 20 CM AGB

Reason

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Code Info

REF(Material)/UDI-DI/Lot(Batch): ASK-42802-PLH/10801902193275, 10801902133110, 10801902212013/33F25L0140 33F26B0222, 33F24F0799; AK-12123-F/10801902168464, 10801902193275, 10801902186338/33F24B0185, 33F24C0102, 33F24D0077, 33F24F0323, 33F24G0346, 33F24K0112, 33F25C0929, 33F25E0062, 33F25H0028, 33F25J0833, 33F24G0129; AK-15123-F/10801902168525, 10801902193275, 10801902186499/33F24C0108, 33F24E0349, 33F24G0353, 33F24L0650, 33F25A0478, 33F25C0779, 33F25E0068, 33F25M0946, 33F24E0713; ASK-12703-PMAH/10801902193275/33F24J0116; ASK-12703-PNM/10801902193275, 10801902214338/33F24L0197, 33F24D0725; ASK-12703-PUMH/10801902193275, 10801902164602/33F25L0030, 33F24F0914; ASK-15703-PRH/10801902193275, 10801902133226/33F25C0271, 33F25F0168, 33F25J0101, 33F25J0368, 33F26A1500, 33F24C0858; CDC-15703-P1A/10801902116793, 10801902193275/33F24D0980, 33F24E0236, 33F24E0832, 33F24G0252, 33F24H0074, 33F24H0401, 33F24H0457, 33F24J0195, 33F24K0024, 33F24M0016, 33F24M0017, 33F24M0064, 33F25B0116, 33F25C0577, 33F25D0719, 33F25G0427, 33F25G0441, 33F25G0663, 33F25H0361, 33F25J1039, 33F25M0495, 33F25M0512, 33F25B0670; NA-12703-PCDC/10801902193275, 10801902119978, 10801902119985/33F24G0555, 33F24K0298, 33F25A0602, 33F25E1055, 33F25H0180, 33F25J0942, 33F25L0474, 33F26B0598, 33F24D0414; NA-15703-PCDC/10801902119985, 10801902193275, 10801902119985, 10801902210217/33F24F0191, 33F24G0556, 33F24J0336, 33F24K0572, 33F25A0157, 33F25C0200, 33F25C0833, 33F25E0275, 33F25H0181, 33F25H0762, 33F25J0943, 33F25L0456, 33F26B0587, 33F25A0603; AK-25553/10801902193275, 10801902167801/33F24H0250, 33F24M0062, 33F25A0382, 33F25B0310, 33F25C0069, 33F25C0506, 33F26C0405, 33F24G0377; CDC-22402-X1A/)10801902193275/33F25K0177, 33F26B0893, 33F25E1192; CDC-22502-X1A/10801902193275/33F25F0216, 33F25C0331; CDC-25402-X1A/10801902193275, 20801902116974/33F25A0287, 33F25D0505, 33F25E0824, 33F25H0569, 33F26A1470, 33F26B0216, 33F24L0486; CDC-25502-X1A/10801902193275/33F24L0980, 33F25D0506, 33F25E0698, 33F25G0315, 33F25J0429, 33F26A1472, 33F24L0832; CDC-25553-X1A/10801902193275, 10801902168129, 60801902123611/33F25E0890, 33F25E1191, 33F26A1384, 33F26B0708, 33F24F0084; CDC-26553-X1A/10801902193275, 10801902168136, 50801902117019/33F25E0749, 33F25E1190, 33F25J0432, 33F26B0807, 33F26C0210, 33F26D0063, 33F24D0807; AK-22123-CDC/10801902193275, 10801902167740/33F24K0127, 33F25A0483, 33F25E0076, 33F25E0938, 33F25H0032, 33F25J0480, 33F25K0398, 33F25M0960, 33F24E0358; AK-22123-F/10801902193275/33F24G0362, 33F24J0526, 33F24K0128, 33F24L0762, 33F25C0963, 33F25E0939, 33F25H0033, 33F24G0363; AK-22402/10801902193275, 10801902167764/33F24L0704, 33F25A0484, 33F25C0817, 33F25E0079, 33F25H0035, 33F25L0373, 33F25M0879, 33F24B0203; AK-22502/10801902193275, 10801902167764/33F24E0363, 33F24G0366, 33F24K0131, 33F24L0708, 33F25C0885, 33F25J0842, 33F24K0487; AK-24301-SK/10801902193275/33F24G0367, 33F24J0528, 33F24K0132, 33F24L0680, 33F25C0118, 33F25C0775, 33F25E0080, 33F25G0092, 33F25H0036, 33F25H0724, 33F25J0482, 33F25L0411, 33F25M0915, 33F24K0133; AK-24402/10801902193275/33F24L0059, 33F24L0727, 33F25C0119, 33F25C0976, 33F25E0081, 33F24K0134; AK-24703-ASK/10801902193275, 10801902118971, 60801902123178/33F24L0778, 33F25C0928, 33F25E0082, 33F25H0037, 33F25J0843, 33F26B0495, 33F24F0281; AK-25123-CDC/10801902193275/33F24K0136, 33F24L0730; AK-25123-F/10801902193275/33F25C0913, 33F25E0085, 33F25E0943, 33F25J0484, 33F26B0496, 33F24G0374; AK-25402/10801902193275, 10801902167795/33F24G0376, 33F24J0282, 33F24K0139, 33F24L0752, 33F25C0121, 33F25C0968, 33F25G0093, 33F25J0486, 33F25L0381, 33F26B0532, 33F24G0688; AK-25502/10801902193275, 10801902180381/33F24H0051, 33F24M0001, 33F25A0125, 33F25B0333, 33F25F0256, 33F25F0500, 33F25H0807, 33F25K0228, 33F25L0163, 33F26A0051, 33F26B0266, 33F24G0689; AK-26553/10801902193275, 20801902117506/33F24K0140, 33F24L0776, 33F25E0089, 33F25G0094, 33F25J0847, 33F25L0410, 33F25M0914, 33F24D0024; CDC-22123-1A/10801902168044, 10801902193275/33F24F0943, 33F24L0472, 33F25A0325, 33F25B0672, 33F25E0653, 33F25E0877, 33F25M0114, 33F25M0424, 33F26A1508, 33F25A0326; CDC-22123-X1A/10801902193275, 20801902116851, 20801902116851,10801902213591/33F25B0673, 33F25D0446, 33F25D0667, 33F25F0212, 33F25G0413, 33F25J0317, 33F25M0396, 33F26B0442, 33F24D0842; CDC-24301-1A/30801902116902, 10801902193275, 30801902116902, 30801902116919/33F24D0874, 33F24F0076, 33F24F0863, 33F24K0501, 33F24L0979, 33F25B0617, 33F25D0494, 33F25D0668, 33F25E0697, 33F25E0820, 33F25H0571, 33F25J0249, 33F25K0181, 33F25L0084, 33F25L0104, 33F26A1413, 33F26B0784, 33S25E0697, 33F24C0702; AK-22702-P1A/10801902193275, 30801902122378/33F25D0322, 33F25H0424, 33F25J0097, 33F25K0085, 33F24E0364; CDC-22702-XP1A/10801902193275, 30801902116896/33F25D0451, 33F25F0217, 33F25H0573, 33F25J0322, 33F26A1386, 33F26B0706, 33F25K0180; ASK-45703-CPR/10801902193275/33F24J0477, 33F24L1068; ASK-42703-AKP1A/10801902193275, 10801902202519/33F24J0324, 33F25A0577, 33F25C0841, 33F25E0233, 33F25E1027, 33F25G0140, 33F25H0148, 33F25H0754, 33F25J0924, 33F25M0869, 33F24F0754; ASK-42703-KH/10801902193275, 10801902205657/33F24G0269, 33F24L0087, 33F25E0235, 33F25G0141, 33F25L0213, 33F24D0891; ASK-42703-MGH1/10801902193275, 10801902223682, 10801902193275/33F25J0735, 33F26C0344, 33F25E0509; ASK-42703-PCCH2/10801902193275, 10801902141030/33F24L0231, 33F24L0416, 33F24L0945, 33F25B0443, 33F25B0561, 33F25D0352, 33F25E0726, 33F25G0316, 33F25J0155, 33F25M0013, 33F26B0792, 33F26C0228, 33F25D0540; ASK-42703-PFS1/10801902193275/33F25E0598, 33F25B0275; ASK-42703-PNO2/10801902193275/33F26A1661, 33F24L0335; ASK-42703-PSCH1/10801902137330, 10801902193275/33F24F1038, 33F24H0204, 33F25B0566, 33F25H0491, 33F25J0274, 33F25M0092, 33F26B0124, 33F24H0100; ASK-42802-AKP1A/10801902193275, 10801902217155/33F24G0511, 33F24K0264, 33F24L0745, 33F25C0931, 33F25H0150, 33F25J0561, 33F25L0481, 33F25M0971, 33F24F0676; ASK-42802-LMDTG1/10801902193275, 10801902205701/33F26A1254, 33F26B0814, 33F24H0215; ASK-42802-PUCI1/10801902193275/33F24L0195; ASK-42802-PUMH/10801902193275, 10801902164251/33F25H0493, 33F25K0059, 33F25L0141, 33F25M0674, 33F24D0712; ASK-42802-UMM/10801902159653, 10801902193275/33F24F1027, 33F24G0513; ASK-42854-AKP1A/10801902193275, 10801902202540/33F24J0635, 33F24K0265, 33F24L0610, 33F25A0580, 33F25E0239, 33F25H0152, 33F25J0562, 33F24E0632; ASK-42854-MGH/10801902223668, 10801902193275/33F26C0876, 33F25E1198; ASK-45703-AKP1A/10801902193275, 10801902158120/33F24G0515, 33F24J0636, 33F24K0266, 33F24L0577, 33F25A0581, 33F25C0843, 33F25E0241, 33F25G0142, 33F25H0154, 33F25J0564, 33F25L0238, 33F25M0902, 33F24F0106; ASK-45703-MGH/10801902223675/33F26C0875, 33F25B0580; ASK-45703-PCCH2/10801902193275, 10801902141047, 10801902193275/33F24L0356, 33F24L1092, 33F25B0582, 33F25D0550, 33F25E0615, 33F25G0352, 33F25J0163, 33F25K0154, 33F26B0772, 33F24L0088; ASK-45703-PECL1/10801902193275, 10801902130386/33F25E0618, 33F25F0204, 33F25G0289, 33F25J0167, 33F25M0192, 33F26B0191, 33F26B0765, 33F25C0313; ASK-45703-PFS1/10801902193275/33F25E0510, 33F25B0274; ASK-45703-PHM1/10801902193275, 10801902116229/33F25E0620, 33F25G0290, 33F25H0572, 33F25J0405, 33F26B0715, 33F26B0854, 33F25K0313; ASK-45703-PHU/10801902193275/33F25L0135, 33F26A0243, 33F25G0685; ASK-45703-PMSG1/10801902193275, 30801902116315/33F25E1199, 33F25F0357, 33F26A0411, 33F24D0674; ASK-45703-PNCB/30801902116315, 10801902193275/33F24F0818, 33F24L0256, 33F24L1075, 33F25B0592, 33F25D0486, 33F25E0731, 33F25H0433, 33F25M0159, 33F25M0426, 33F26B0760, 33F25B0456; ASK-45703-PPCM/10801902193275, 10801902116335/33F25F0358, 33F25G0688, 33F25K0048, 33F25L0028, 33F26B0903, 33F24K0067; ASK-45703-PRH/10801902193275, 10801902116359, 10801902116373/33F25A0275, 33F25C0318, 33F25E0807, 33F25G0358, 33F25J0301, 33F26A1420, 33F26B0799, 33F24E0560; ASK-45703-PSJ2/10801902193275, 10801902119732/33F25C0319, 33F25E0733, 33F25G0357, 33F25H0641, 33F25J0302, 33F26A1507, 33F26B0878, 33F24L0259; ASK-45703-PSVM/10801902119749, 10801902193275, 10801902119749/33F24F0755, 33F24L0446, 33F25B0459, 33F25C0320, 33F25D0551, 33F25F0120, 33F25G0356, 33F25J0303, 33F26A1523, 33F26B0762, 33F25D0237; ASK-45703-PUHS/10801902193275, 10801902114058/33F24L0942, 33F25B0596, 33F25E0687, 33F25G0303, 33F25H0642, 33F26A1424, 33F24F0717; ASK-45703-PUKL/10801902193275, 10801902114058, 10801902193275/33F24H0185, 33F24K0433, 33F24L0262, 33F25B0597, 33F25D0372, 33F25E0628, 33F25H0566, 33F25J0173, 33F26B0233, 33S25E0628, 33F25B0461; ASK-45703-PUMF1/10801902193275/33F25K0060, 33F25L0134, 33F26A0235, 33F24K0567; ASK-45703-PUR1/10801902116465, 10801902193275/33F24F0790, 33F24H0206, 33F24L0266, 33F24L0922, 33F25B0601, 33F25D0375, 33F25E0630, 33F25E0870, 33F25J0408, 33F25M0128,33F26A1432, 33F26B0836, 33F24G0616; ASK-45703-RHR1/10801902116465, 10801902193275/33F24F0790, 33F24H0206, 33F24L0266,33F24L0922, 33F25B0601,33F25D0375, 33F25E0630, 33F25E0870, 33F25J0408, 33F25M0128, 33F26A1432, 33F26B0836, 33F24G0616; ASK-45703-WA/10801902193275, 10801902116564, 30801902119774/33F25A0443, 33F25E0011, 33F25K0309, 33F26B0909, 33F24C0746; ASK-45802-AKP1A/10801902193275, 30801902119774/33F24L0023, 33F26A1657, 33F26B0277, 33F25D0034; ASK-45854-AKP1A/10801902193275, 30801902119781/33F24G0671, 33F24L0039, 33F25A0366, 33F25B0094, 33F25D0715, 33F25G0495, 33F25J0033, 33F25K0349, 33F25L0226, 33F26A1654, 33F26B0025, 33F26B0698, 33F24J0130; ASK-45854-MGH/10801902223699, 10801902193275/33F26C0688, 33F25C0363; ASK-45854-PCCH2/10801902193275/33F25J0234, 33F25E1175; ASK-45854-PNM/10801902193275/33F24K0570, 33F25B0431, 33F25E0006, 33F25F0377, 33F25H0780, 33F25L0261; ASK-45854-PPCM/10801902143867/33F24F0736; ASK-45854-PUCH2/10801902193275/33F24M0177; ASK-46702-PWBH1/10801902193275/33F24K0321, 33F24K0269; CDC-42802-P1A/10801902170870, 10801902193275, 10801902170870/33F24F0244, 33F24H0070, 33F24H0426, 33F24J0223, 33F24J0468,33F24L0009,33F24M0079, 33F24M0139, 33F25B0319, 33F25C0533, 33F25E1094, 33F25E1156,33F25F0384, 33F25H0338, 33F25J0766, 33F25J0990, 33F25L0086, 33F25M0016, 33F26A1619, 33F26B1051, 33F25H0880; CDC-45703-PA1A/10801902193275, 30801902117190, 10801902117202/33F24M0081, 33F25F0241, 33F26A1656, 33F26B0334, 33F24C1072; ASK-42802-ECL4/10801902224283, 10801902205664/33F26B0489, 33F24D0651; AK-42703-P1A/10801902193275, 10801902171372/33F25D0323, 33F25D0585, 33F25F0122, 33F25F0153, 33F25G0376, 33F25J0135, 33F26B0459, 33F24L1104; AK-42802-P1A/10801902193275/33F25B0539, 33F25D0386, 33F25F0137, 33F25J0136, 33F25B0512; AK-42854-P1A/10801902193275, 10801902168167, 10801902119152/33F25E0664, 33F25H0442, 33F25J0341, 33F26B0752, 33F23H0557; AK-45703-P1A/10801902193275, 10801902171389/33F25C0341, 33F25D0387, 33F25D0643, 33F25F0154, 33F25G0257, 33F25G0375, 33F25J0137, 33F25J0342, 33F26A1416, 33F26B0150, 33F26B0766, 33F25C0191; AK-45802-P1A/10801902193275, 10801902168181/33F25E0665, 33F25J0191, 33F26B0754, 33F25A0223; AK-45854-P1A/10801902193275, 10801902168198/33F25D0459, 33F25F0138, 33F26B0145, 33F25B0540; AK-47702-P1A/10801902193275/33F25E0502, 33F25G0162, 33F25H0407, 33F25J0343, 33F24G0383; ASK-42703-PECL3/10801902193275/33F26A1397, 33F25B0562; ASK-42703-PUPM3/10801902225150, 10801902156966/33F26C0238, 33F24F0977; ASK-42802-PUPM3/10801902225143, 10801902193275/33F26C0239, 33F25B0498; ASK-45703-PECL3/10801902193275, 10801902224313/33F26A1577, 33F26B0773, 33F25C0234; ASK-45703-PUPM3/10801902225181, 10801902116465/33F26C0219, 33F24D0840; ASK-45854-PECL3/10801902193275/33F24L0191; ASK-45854-PUPM3/10801902225167, 10801902193275/33F26C0240, 33F24H0111; ASK-45854-SMW/10801902224436, 10801902214888/33F26B0830, 33F24F0872; CDC-42802-XP1A/10801902193275, 10801902168365, 10801902162103/33F25J0295, 33F25J0728, 33F25K0304, 33F25K0390, 33F25L0228, 33F25M0037, 33F25M0485, 33F25M0717, 33F26A1630, 36F25C0008, 36F25D0001, 36F25D0034, 36F25E0001, 36F25G0022, 36F25H0015, 33F24D0258; CDC-42854-XP1A/10801902193275, 20801902117148, 10801902162134/33F25J0681, 33F26A0390, 33F26C0209, 33F26D0062, 36F25D0003, 36F25H0014, 33F24D0827; CDC-45703-XP1A/10801902193275, 20801902117216/33F25G0286, 33F25H0876,33F25J0089,33F25J0682, 33F25J0702, 33F25J0705, 33F25J0711,33F25J0718, 33F25J0729, 33F25J0733, 33F25K0305, 33F25L0063,33F25L0093, 33F25L0094, 33F25L0231, 33F25L0252, 33F25L0253, 33F25L0254, 33F25L0286, 33F25L0302, 33F25L0322, 33F25M0487, 33F26A0062, 33F26B0637, 33F26B0638, 33F26D0185, 36F25A0008, 36F25A0014, 36F25D0010, 36F25D0011, 36F25D0014, 36F25D0021, 36F25D0022, 36F25D0023, 36F25D0028, 36F25D0041, 33F25C0372; CDC-45854-XP1A/10801902193275, 10801902169720, 10801902162158/33F25J0695, 33F25J0730, 33F25K0394, 33F26A1620, 33F26B0572, 33F26B0735, 36F25G0003, 36F25G0021, 36F25G0024, 36F25H0011, 36F25H0017, 33F24D0424 ASK-42703-PGM3/10801902213751/33F24F0951, 33F24K0743; ASK-42802-UH1/10801902156386/33F24H0122; ASK-45703-PGM3/10801902213768/33F24D0844, 33F24F0840;

Distribution

US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.

Quantity

308,263