Description/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION KIT/AK-05502D; EPIDURAL CATHETERIZATION KIT/AK-05503; EPIDURAL CATHETERIZATION KIT/AK-05503-L; SPINAL ANES/EPIDURAL CATH KIT/AK-05560; SPINAL ANES/EPIDURAL CATH KIT/ASK-02520-SBMC1; EPIDURAL CATHETERIZATION KIT/ASK-05400-BW1; EPIDURAL CATHETERIZATION KIT/ASK-05400-GH; SPINAL ANES/EPIDURAL CATH KIT/ASK-05401-LMC; EPIDURAL CATHETERIZATION KIT/ASK-05401-NM; SPINAL ANES/EPIDURAL CATH KIT/ASK-05500-BID1; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-JHH1; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-MCV1; EPIDURAL CATHETER KIT: 19GA X 90CM ASK-05500-MVH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05500-UMC; 19 GA X 35.5 IN EPIDURAL CATHETER KIT/ASK-05502-MLH; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-JMH1; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-MKD2; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-OM1; EPIDURAL CATHETERIZATION KIT/ASK-17019-SM; EPIDURAL CATHETERIZATION KIT/BP-05501; EPIDURAL CATHETERIZATION KIT/CK-05401; EPIDURAL CATHETERIZATION KIT/HS-05501; EPIDURAL CATHETERIZATION KIT/MM-05501; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TK; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TKD; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TKL; EPIDURAL CATHETERIZATION KIT/MSO-05401-JMH; EPIDURAL CATHETERIZATION KIT/NM-05401; EPIDURAL CATHETERIZATION KIT/SJ-05501; EPIDURAL CATHETERIZATION KIT/SM-05401; EPIDURAL CATHETER KIT: 19 GA X 90 CM/TM-05401; EPIDURAL CATHETERIZATION KIT/TM-05502; SPINAL ANES/EPIDURAL CATH KIT/UI-05502; SPINAL ANES/EPIDURAL CATH KIT: 19 GA/ASK-05400-SHM; EPIDURAL CATH KIT: 19 GA/ASK-05400-UOM; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05500-MR; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-TM; EPIDURAL CATHETERIZATION KIT/ASK-05501-BHL1; EPIDURAL CATHETERIZATION KIT/ASK-05501-GH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-SM; EPIDURAL CATHETERIZATION KIT/ASK-05502-BSM; EPIDURAL CATHETERIZATION KIT/ASK-05502-MGH1; EPIDURAL CATHETERIZATION KIT/ASK-05503-BWB1; EPIDURAL CATHETERIZATION KIT/ASK-05503-HMH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05503-SPH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05503-UPM1; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-UCI2; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-17019-UMP; EPIDURAL CATHETERIZATION KIT/MH-05503-1; EPIDURAL CATHETERIZATION KIT/ASK-05401-VB2; Epidural Catheterization Kit: 19GA/ASK-05500-SC1; Epidural Catheterization Kit: 19GA/ASK-05500-SC3; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-AFH1; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-BMC; EPIDURAL CATHETERIZATION KIT/ASK-05502-NY; EPIDURAL CATHETERIZATION KIT/ASK-17019-RMH
Enforcement
- Recall Number
- Z-2371-2026
- Event ID
- 98793
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- ARROW INTERNATIONAL, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 24, 2026
- Initiation Date
- April 21, 2026
- Classification Date
- June 12, 2026
- Address
- 3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437, United States
Description
Description/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION KIT/AK-05502D; EPIDURAL CATHETERIZATION KIT/AK-05503; EPIDURAL CATHETERIZATION KIT/AK-05503-L; SPINAL ANES/EPIDURAL CATH KIT/AK-05560; SPINAL ANES/EPIDURAL CATH KIT/ASK-02520-SBMC1; EPIDURAL CATHETERIZATION KIT/ASK-05400-BW1; EPIDURAL CATHETERIZATION KIT/ASK-05400-GH; SPINAL ANES/EPIDURAL CATH KIT/ASK-05401-LMC; EPIDURAL CATHETERIZATION KIT/ASK-05401-NM; SPINAL ANES/EPIDURAL CATH KIT/ASK-05500-BID1; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-JHH1; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-MCV1; EPIDURAL CATHETER KIT: 19GA X 90CM ASK-05500-MVH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05500-UMC; 19 GA X 35.5 IN EPIDURAL CATHETER KIT/ASK-05502-MLH; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-JMH1; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-MKD2; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-OM1; EPIDURAL CATHETERIZATION KIT/ASK-17019-SM; EPIDURAL CATHETERIZATION KIT/BP-05501; EPIDURAL CATHETERIZATION KIT/CK-05401; EPIDURAL CATHETERIZATION KIT/HS-05501; EPIDURAL CATHETERIZATION KIT/MM-05501; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TK; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TKD; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TKL; EPIDURAL CATHETERIZATION KIT/MSO-05401-JMH; EPIDURAL CATHETERIZATION KIT/NM-05401; EPIDURAL CATHETERIZATION KIT/SJ-05501; EPIDURAL CATHETERIZATION KIT/SM-05401; EPIDURAL CATHETER KIT: 19 GA X 90 CM/TM-05401; EPIDURAL CATHETERIZATION KIT/TM-05502; SPINAL ANES/EPIDURAL CATH KIT/UI-05502; SPINAL ANES/EPIDURAL CATH KIT: 19 GA/ASK-05400-SHM; EPIDURAL CATH KIT: 19 GA/ASK-05400-UOM; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05500-MR; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-TM; EPIDURAL CATHETERIZATION KIT/ASK-05501-BHL1; EPIDURAL CATHETERIZATION KIT/ASK-05501-GH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-SM; EPIDURAL CATHETERIZATION KIT/ASK-05502-BSM; EPIDURAL CATHETERIZATION KIT/ASK-05502-MGH1; EPIDURAL CATHETERIZATION KIT/ASK-05503-BWB1; EPIDURAL CATHETERIZATION KIT/ASK-05503-HMH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05503-SPH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05503-UPM1; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-UCI2; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-17019-UMP; EPIDURAL CATHETERIZATION KIT/MH-05503-1; EPIDURAL CATHETERIZATION KIT/ASK-05401-VB2; Epidural Catheterization Kit: 19GA/ASK-05500-SC1; Epidural Catheterization Kit: 19GA/ASK-05500-SC3; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-AFH1; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-BMC; EPIDURAL CATHETERIZATION KIT/ASK-05502-NY; EPIDURAL CATHETERIZATION KIT/ASK-17019-RMH
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
REF(Material)/UDI-DI/Lot(Batch): ASK-05502-VM1/10801902210217, 10801902209938/33F25C0168, 33F25C0880, 33F25E0202, 33F25G0134, 33F25H0126, 33F26C0297, 33F24C0216; AK-05501/)10801902210217, 10801902209501, 10801902209518/33F24E0057, 33F24E0058, 33F24J0496, 33F24K0094, 33F24L0564, 33F25A0460, 33F25A0617, 33F25C0728,33F25E0043, 33F25E0913, 33F25G0081, 33F25H0015, 33F25J0465, 33F25J0822, 33F25K0254, 33F25L0376, 33F26C0300, 33F24B0119; AK-05502/10801902209518, 10801902210217, 10801902209525/33F24C0015, 33F24G0581, 33F24G0582, 33F24G0583, 33F24J0497, 33F24J0678, 33F24J0679, 33F24J0680, 33F24J0681, 33F24K0095, 33F24K0310, 33F24L0580, 33F24L0624, 33F24L0628, 33F24L0630, 33F24L0757, 33F24L0785, 33F25A0461, 33F25A0618, 33F25A0619, 33F25A0620, 33F25C0103, 33F25C0724, 33F25C0725, 33F25C0741, 33F25C0791, 33F25E0044, 33F25E0291, 33F25E0292, 33F25E0293, 33F25E0914, 33F25E1067, 33F25G0082, 33F25G0159, 33F25H0016, 33F25H0197, 33F25H0198, 33F25H0199, 33F25J0466, 33F25J0643, 33F25J0790, 33F25J0805, 33F25J0806, 33F25J0807, 33F25J0808, 33F25K0024, 33F25K0252, 33F25K0358, 33F25K0359, 33F25L0099, 33F25L0266, 33F25L0267, 33F25M0970, 33F25M0978, 33F26A1100, 33F24C0096; AK-05502D/10801902210217, 10801902209532/33F24G0331, 33F24J0498, 33F24K0096, 33F24L0713, 33F25A0462, 33F25C0745, 33F25E0045, 33F25H0017, 33F25J0467, 33F25L0366, 33F24C0017; AK-05503/10801902210217, 10801902209549/33F24J0277, 33F24J0499, 33F24K0097, 33F24L0674, 33F25A0463, 33F25C0949, 33F25E0046, 33F25E0915, 33F25H0018, 33F25J0823, 33F25L0406, 33F25M0912, 33F24C0018; AK-05503-L/10801902209549, 10801902210217, 10801902209549/33F24C0019, 33F24E0066, 33F24J0682, 33F24K0098, 33F24L0613, 33F24L0758, 33F25A0464, 33F25C0104, 33F25C0731, 33F25C0793, 33F25E0047, 33F25E0916, 33F25G0083, 33F25H0019, 33F25H0200, 33F25J0791, 33F25J0809, 33F25K0026, 33F25K0253, 33F25L0057, 33F25L0352, 33F25M0861, 33F26A0989, 33F24J0278; AK-05560/10801902210217, 10801902209556, 30801902121722/33F24J0501, 33F24K0099, 33F24L0552, 33F25A0145, 33F25A0465, 33F25C0743, 33F25E0048, 33F25E0917, 33F25G0084, 33F25H0020, 33F25J0737, 33F25J0824, 33F25M0895, 33F26B0538, 33F26C0592, 33F24A0812; ASK-02520-SBMC1/10801902209570, 10801902210217, 10801902193275/33F24D0103, 33F24F0275, 33F24G0396, 33F25A0137; ASK-05400-BW1/10801902209617, 10801902209631/33F24D0153, 33F24E0474, 33F24G0461, 33F24J0307, 33F24L0556, 33F25B0246, 33F25C0729, 33F25E0178, 33F25G0121, 33F25J0899, 33F25L0042, 33F25L0450, 33F24C0196; ASK-05400-GH/10801902209631, 10801902210217, 10801902202199/33F24D0154, 33F24E0480, 33F24G0292, 33F24J0310, 33F24J0600, 33F24L0597, 33F25A0153, 33F25C0161, 33F25C0754, 33F25E0181, 33F25G0123, 33F25J0532, 33F25L0391, 33F25M0897, 33F24D0155; ASK-05401-LMC/10801902210217, 10801902209662/33F24J0603, 33F25F0240, 33F25G0754, 33F25J0534, 33F24D0031; ASK-05401-NM/10801902210217/33F24E0104, 33F24G0295, 33F24J0311, 33F24J0604, 33F24K0223, 33F24L0559, 33F25A0548, 33F25C0840, 33F25E0184, 33F25G0125, 33F25H0113, 33F25J0535, 33F25L0024, 33F25L0361, 33F24E0027; ASK-05500-BID1/10801902210217/33F24E0490, 33F24G0297, 33F24H0488, 33F24J0607, 33F24L0594, 33F25A0551, 33F25C0823, 33F25E0187, 33F25G0127, 33F25J0536, 33F25L0438, 33F24K0228; ASK-05500-JHH1/10801902210217, 10801902209716, 10801902209723/33F24L0572, 33F25B0379, 33F25D0215, 33F25E0190, 33F25G0129, 33F25H0744, 33F25J0538, 33F25K0407, 33F25M0955, 33S24L0572, 33F24C0203; ASK-05500-MCV1/10801902210217/33F24E0494, 33F24E0495, 33F24G0470, 33F24J0611, 33F24K0229, 33F24L0707, 33F25A0554, 33F25C0784, 33F25E0191, 33F25E1005, 33F25G0761, 33F25H0115, 33F24E0496; ASK-05500-MVH/10801902209730, 10801902210217/33F24D0380, 33F24E0498, 33F24G0472, 33F24L0681; ASK-05500-UMC/10801902209754, 10801902210217/33F24D0382, 33F24J0316, 33F24K0233, 33F24L0085, 33F24L0772, 33F25C0166, 33F25C0737, 33F25E0196, 33F25E0317, 33F25E1008, 33F25H0120, 33F25J0540, 33F25L0397, 33F25M0904, 33F24E0502; ASK-05502-MLH/10801902210217/33F24J0618, 33F24K0239, 33F24L0555, 33F25A0562, 33F25C0996, 33F25E0201, 33F25H0125, 33F25J0544, 33F25L0440, 33F26B0977, 33F24L0583; ASK-05560-JMH1/10801902210217, 10801902209983/33F24G0483, 33F24L0773, 33F25A0566, 33F25C0894, 33F25E1014, 33F25H0130, 33F25J0548, 33F24E0519; ASK-05560-MKD2/10801902193275, 10801902210217/33F24K0246, 33F25G0468, 33F24J0476; ASK-05560-OM1/10801902210217, 10801902210002/33F24K0550, 33F25A0155, 33F25C0171, 33F25F0383, 33F25G0560, 33F25H0131, 33F25H0746, 33F25J0784, 33F25L0006, 33F25M0903, 33F24D0391; ASK-17019-SM/10801902210217, 10801902211221/33F25G0137, 33F25H0136, 33F25J0553, 33F25L0465, 33F24B0348; BP-05501/10801902210217, 10801902193275/33F24E0156, 33F24J0642, 33F24K0270, 33F25B0689, 33F25C0868, 33F25E1034, 33F25G0145, 33F25H0158, 33F25J0568, 33F25L0369, 33F26B0518, 33F25C0444; CK-05401/10801902202144, 10801902210217, 10801902193275/33F24C0256, 33F25A0591, 33F25C0972, 33F25E0256, 33F25E1040, 33F25H0756, 33F25J0935, 33F25L0244, 33F26B0620, 33F25D0674; HS-05501/10801902210217, 10801902202090/33F24E0580, 33F24J0329, 33F24B0389; MM-05501/10801902210217/33F24J0655, 33F24K0287, 33F24L0788, 33F25J0628, 33F25L0445, 33F24C0263; MP-17019-TK/10801902210217/33F24D0409, 33F24J0390, 33F24K0288, 33F24L0546, 33F25A0595, 33F25C0727, 33F25E0262, 33F25E1045, 33F25H0170, 33F25J0629, 33F25L0441, 33F25M0939, 33F24D0410; MP-17019-TKD/10801902210217, 10801902210224/33F24J0657, 33F24L0698, 33F25C0760, 33F25E0263, 33F25H0171, 33F25J0938, 33F24B0152; MP-17019-TKL/10801902210224, 10801902210217/33F24D0038, 33F24K0289, 33F24L0614, 33F25C0190, 33F25C0839, 33F25E0264, 33F25G0149, 33F25J0630, 33F25K0380, 33F24K0290; MSO-05401-JMH/10801902210217/33F25C0193, 33F25H0172, 33F25J0631, 33F25L0375, 33F25M0880, 33F24D0413; NM-05401/10801902210217, 10801902210262, 10801902140644/33F25J0944, 33F25M0968, 33F26B0600, 33F24C0780; SJ-05501/10801902210330, 10801902210217, 10801902202496/33F24C0049, 33F24C0050, 33F24J0382, 33F24J0383, 33F24J0384, 33F24J0385, 33F24J0386, 33F24K0552, 33F24K0553, 33F24K0554, 33F24K0555, 33F24K0556, 33F24K0557, 33F24L0625, 33F24L0721, 33F24L0760, 33F24L0786, 33F25A0605, 33F25A0623, 33F25A0624, 33F25C0202, 33F25C0789, 33F25C0790, 33F25C0945, 33F25C0997, 33F25E0278, 33F25E0295, 33F25E0296, 33F25E1057, 33F25E1068, 33F25G0154, 33F25G0160, 33F25H0183, 33F25H0202, 33F25H0203, 33F25J0639, 33F25J0644, 33F25J0798, 33F25J0811, 33F25J0812, 33F25J0813, 33F26B0601, 33F26C0299, 33F26C0307, 33F24A1336; SM-05401/10801902210217, 10801902210354, 10801902193275/33F24G0561, 33F24K0302, 33F24L0607, 33F25C0204, 33F25C0805, 33F25E0279, 33F25E1058, 33F25H0184, 33F25J0948, 33F26A1030, 33F24G0270; TM-05401/10801902210217, 30801902202155, 10801902210361/33F24D0418, 33F24G0563, 33F24J0669, 33F24L0089, 33F24L0741, 33F25C0207, 33F25C0918, 33F25E0282, 33F25E1059, 33F25J0950, 33F25L0447, 33F26C0308, 33F24B0410; TM-05502/10801902210378/33F24B0411; UI-05502/10801902210217/33F24G0565, 33F24J0670, 33F24L0562, 33F25A0607, 33F25C0881, 33F25E0283, 33F25E1060, 33F25G0752, 33F25H0186, 33F25J0951, 33F24A0359; ASK-05400-SHM/10801902210217/33F24E0482, 33F24G0465, 33F24J0601, 33F24K0221, 33F24L0800, 33F25C0162, 33F25C0888, 33F25E0182, 33F25H0741, 33F25L0457, 33F24E0483; ASK-05400-UOM/10801902210217/33F25G0124, 33F25H0742, 33F25K0263, 33F24G0294; ASK-05500-MR/10801902210217, 10801902209730/33F24E0497, 33F24G0471, 33F24K0230, 33F25A0555, 33F25C0796, 33F25E0192, 33F25H0116, 33F25K0408, 33F25M0854, 33F24C0204; ASK-05500-TM/10801902210217, 10801902209747, 10801902209754/33F25C0165, 33F25E1007, 33F25M0871, 33F26C0256, 33F24C0206; ASK-05501-BHL1/10801902210217/33F24L0566, 33F25A0558, 33F25C0810, 33F25E0197, 33F25E1009, 33F25H0121, 33F25J0541, 33F25L0439, 33F24E0503; ASK-05501-GH/10801902210217, 10801902210392/33F24K0235, 33F25A0559, 33F25E0198, 33F25H0122, 33F25J0542, 33F24D0384; ASK-05501-SM/10801902210217, 10801902210392, 10801902214857/33F24E0505, 33F24K0236, 33F25A0560, 33F25C0808, 33F25E0199, 33F25E1010, 33F25H0123, 33F25J0907, 33F26C0304, 33F24C0210; ASK-05502-BSM/10801902209853, 10801902210217/33F24D0385, 33F24E0508, 33F24G0301, 33F24K0237, 33F25A0154, 33F25B0687, 33F25H0124, 33F25J0543, 33F25K0409, 33F25L0466, 33F26B0592, 33F24E0509; ASK-05502-MGH1/10801902210217, 10801902209860/33F24J0617, 33F24L0569, 33F25B0247, 33F25C0938, 33F25E0304, 33F25G0133, 33F25J0909, 33F25L0098, 33F26B0545, 33F24J0318; ASK-05503-BWB1/10801902210217, 10801902209938, 10801902209952/33F24E0514, 33F24G0481, 33F24J0619, 33F24K0241, 33F24L0690, 33F25C0169, 33F25C0756, 33F25E1011, 33F25J0545, 33F25L0453, 33F26B0584, 33F24C0217; ASK-05503-HMH/10801902210217, 10801902209952, 10801902209969/33F24L0643, 33F25A0563, 33F25C0170, 33F25C0765, 33F25E1012, 33F25G0135, 33F25H0127, 33F25J0546, 33F25L0451, 33F26A1079, 33F24C0218; ASK-05503-SPH/10801902210217, 10801902209969, 10801902211405/33F24E0516, 33F24J0620, 33F24K0243, 33F24L0598, 33F25A0564, 33F25E0204, 33F25L0458, 33F25M0951, 33F24C0219; ASK-05503-UPM1/10801902211405, 10801902210217, 10801902209976/33F24D0389, 33F24E0160, 33F24J0621, 33F24K0244, 33F24L0549, 33F25A0565, 33F25C0992, 33F25E0205, 33F25E1013, 33F25H0129, 33F25J0547, 33F25L0454, 33F24D0390; ASK-05560-UCI2/10801902210217, 10801902193275/33F24G0485, 33F24J0623, 33F24L0593, 33F25A0567, 33F25C0893, 33F25E0206, 33F25E1015, 33F25H0132, 33F25J0911, 33F25L0061; ASK-17019-UMP/10801902210217, 10801902211221, 10801902210057/33F24J0628, 33F24K0250, 33F25B0694, 33F25E0216, 33F25H0137, 33F25J0916, 33F26B1000, 33F24B0350; MH-05503-1/10801902210217/33F25H0169, 33F24G0542 ASK-05401-VB2/10801902214840/33F24C0199, 33F24D0374, 33F24E0486, 33F24G0296, 33F24J0312, 33F24J0605, 33F24K0224, 33F24L0733, 33F25A0549; ASK-05500-SC1 /10801902202250/33F24B0319, 33F24C0205, 33F24D0381, 33F24E0499, 33F24G0299, 33F24J0315, 33F24J0613, 33F24K0232, 33F24L0567, 33F25A0557, 33F25C0884, 33F25E0194; ASK-05500-SC3/10801902223651/33F25G0751, 33F25H0118, 33F26B0644; ASK-05501-AFH1/10801902209761/33F24G0475; ASK-05501-BMC/10801902209785/33F24J0479; ASK-05502-NY/10801902209884/33F24E0512; ASK-17019-RMH/10801902214918/33F24B0346, 33F24E0531
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
1,525,831