MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, IC Tray REF: 8888101004HP
Enforcement
- Recall Number
- Z-2370-2023
- Event ID
- 92660
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Covidien LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 23, 2023
- Initiation Date
- June 28, 2023
- Classification Date
- August 11, 2023
- Address
- 15 Hampshire St, Mansfield, MA, 02048-1113, United States
Description
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, IC Tray REF: 8888101004HP
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
GTIN: 20884521127989 LOT Numbers: 2130900107 2133700059
US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
1325 units