Description/REF: RA CATH KIT: 20 GA X 1 3/4 IN/TJ-04020-1; RA CATH KIT: 20 GA X 1 1/2IN/NA-04220-S1A; ARTERIAL LINE KIT/AK-04510-S; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018-HUMC; ARTERIAL LINE KIT: 18 GA X 6 IN (16 CM)/ASK-04018-OH1; RA CATH KIT: 20 GA X 1 3/4IN (4.45 CM)/ASK-04020-BID; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-MAR; ARTERIAL LINE KIT: 20 GA X 5 IN/ASK-04020-PM4; RA CATH KIT: 20 GA X 1 1/2IN (3.81 CM)/ASK-04120-HF3; RA CATH SET: 20 GA X 2 1/2 IN (6.35 CM)/ASK-04120-UR; ARTERIAL CATH KIT: 20 GA X 3 1/16 IN/ASK-04125-MGH; ARTERIAL CATH KIT: 20 GA X 5 IN/ASK-04500-HFH4; ARTERIAL LINE KIT: 20GA X 5IN/ASK-04510-BID2; ARTERIAL LINE KIT: 20GA X 5IN (12 CM)/ASK-04510-EHP1; ARTERIAL LINE KIT/ASK-04510-FEM1; ARTERIAL LINE KIT: 20 GA X 5 IN (12 CM)/ASK-04510-HFH; ARTERIAL CATH KIT: 20 GA X 5 IN/ASK-04510-HFH4; ARTERIAL LINE KIT/ASK-04510-SE; ARTERIAL LINE KIT: 20 GA X 5 IN (12CM)/ASK-04510-TGH2; RA CATH KIT: 20 GA X 12 CM/NA-04550-1A; RA CATH KIT: 20 GA X 12 CM/NA-04550-X1A; RA CATH KIT: 18 GA X 1-3/4IN/AK-04018; RA CATH KIT: 20 GA X 1-3/4IN/AK-04020; RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-C; RA CATH KIT: 20 GA X 1-1/2IN/AK-04120; RA CATH KIT: 22 GA X 1-3/8IN/AK-04122; RA CATH KIT: 20 GA X 1-1/2IN/AK-04220; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-AH; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-ATH; RA CATH KIT: 20 GA X 1 3/4IN (4.45CM)/ASK-04020-AU2; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-BHS; ARTERIAL CATHETERIZATION KIT/ASK-04020-BIDB2; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-CCD1; ARTERIAL CATHETERIZATION KIT/ASK-04020-CR; ART LINE KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-EUH2; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-FAM1; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-FMC; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-KSF2; RA CATH KIT: 20 GA X 1 3/4 IN (4.5 CM)/ASK-04020-LHC2; RA CATH KIT: 20 GA X 1 3/4 IN (4.5 CM)/ASK-04020-LHC3; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-LM; RA CATH KIT: 20 GA X 1 3/4IN/ASK-04020-MCH1; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-MIH1; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-NWMX; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-PHP; RA CATH KIT: 20 GA X 1-3/4 IN (4.45CM)/ASK-04020-RIHM; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-RMC1; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-RU; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-SE; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-SHM; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-SRMC1; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-UCL1; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-UMI; RA CATH KIT: 20 GA X 1-3/4 IN (4.45CM)/ASK-04020-UN; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-UNC; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-UNC1; RA CATH SET: 20 GA X 1-1/2IN (3.81CM)/ASK-04220-PSU; RA CATH KIT: 20 GA X 1-1/2 IN (3.81 CM)/ASK-04220-UCL1; RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-S1A; CVC KIT: 20 GA X 5 IN (12CM)/ASK-04550-JMC1; ARTERIAL ACCESS TRAY/ASK-04001-BW2; ARTERIAL ACCESS TRAY WITH .025 IN SWG/ASK-04001-WUH1; ARTERIAL LINE KIT/ASK-04510-HUM
Enforcement
- Recall Number
- Z-2368-2026
- Event ID
- 98793
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- ARROW INTERNATIONAL, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 24, 2026
- Initiation Date
- April 21, 2026
- Classification Date
- June 12, 2026
- Address
- 3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437, United States
Description
Description/REF: RA CATH KIT: 20 GA X 1 3/4 IN/TJ-04020-1; RA CATH KIT: 20 GA X 1 1/2IN/NA-04220-S1A; ARTERIAL LINE KIT/AK-04510-S; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018-HUMC; ARTERIAL LINE KIT: 18 GA X 6 IN (16 CM)/ASK-04018-OH1; RA CATH KIT: 20 GA X 1 3/4IN (4.45 CM)/ASK-04020-BID; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-MAR; ARTERIAL LINE KIT: 20 GA X 5 IN/ASK-04020-PM4; RA CATH KIT: 20 GA X 1 1/2IN (3.81 CM)/ASK-04120-HF3; RA CATH SET: 20 GA X 2 1/2 IN (6.35 CM)/ASK-04120-UR; ARTERIAL CATH KIT: 20 GA X 3 1/16 IN/ASK-04125-MGH; ARTERIAL CATH KIT: 20 GA X 5 IN/ASK-04500-HFH4; ARTERIAL LINE KIT: 20GA X 5IN/ASK-04510-BID2; ARTERIAL LINE KIT: 20GA X 5IN (12 CM)/ASK-04510-EHP1; ARTERIAL LINE KIT/ASK-04510-FEM1; ARTERIAL LINE KIT: 20 GA X 5 IN (12 CM)/ASK-04510-HFH; ARTERIAL CATH KIT: 20 GA X 5 IN/ASK-04510-HFH4; ARTERIAL LINE KIT/ASK-04510-SE; ARTERIAL LINE KIT: 20 GA X 5 IN (12CM)/ASK-04510-TGH2; RA CATH KIT: 20 GA X 12 CM/NA-04550-1A; RA CATH KIT: 20 GA X 12 CM/NA-04550-X1A; RA CATH KIT: 18 GA X 1-3/4IN/AK-04018; RA CATH KIT: 20 GA X 1-3/4IN/AK-04020; RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-C; RA CATH KIT: 20 GA X 1-1/2IN/AK-04120; RA CATH KIT: 22 GA X 1-3/8IN/AK-04122; RA CATH KIT: 20 GA X 1-1/2IN/AK-04220; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-AH; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-ATH; RA CATH KIT: 20 GA X 1 3/4IN (4.45CM)/ASK-04020-AU2; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-BHS; ARTERIAL CATHETERIZATION KIT/ASK-04020-BIDB2; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-CCD1; ARTERIAL CATHETERIZATION KIT/ASK-04020-CR; ART LINE KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-EUH2; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-FAM1; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-FMC; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-KSF2; RA CATH KIT: 20 GA X 1 3/4 IN (4.5 CM)/ASK-04020-LHC2; RA CATH KIT: 20 GA X 1 3/4 IN (4.5 CM)/ASK-04020-LHC3; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-LM; RA CATH KIT: 20 GA X 1 3/4IN/ASK-04020-MCH1; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-MIH1; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-NWMX; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-PHP; RA CATH KIT: 20 GA X 1-3/4 IN (4.45CM)/ASK-04020-RIHM; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-RMC1; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-RU; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-SE; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-SHM; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-SRMC1; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-UCL1; RA CATH KIT: 20 GA X 4.45 CM/ASK-04020-UMI; RA CATH KIT: 20 GA X 1-3/4 IN (4.45CM)/ASK-04020-UN; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-UNC; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-UNC1; RA CATH SET: 20 GA X 1-1/2IN (3.81CM)/ASK-04220-PSU; RA CATH KIT: 20 GA X 1-1/2 IN (3.81 CM)/ASK-04220-UCL1; RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-S1A; CVC KIT: 20 GA X 5 IN (12CM)/ASK-04550-JMC1; ARTERIAL ACCESS TRAY/ASK-04001-BW2; ARTERIAL ACCESS TRAY WITH .025 IN SWG/ASK-04001-WUH1; ARTERIAL LINE KIT/ASK-04510-HUM
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
REF(Material)/UDI-DI/Lot(Batch): TJ-04020-1/10801902193275, 30801902202155/33F24J0338, 33F25A0158, 33F25C0753, 33F25E0281, 33F24B0409; NA-04220-S1A/10801902193275/33F25E1052, 33F24G0553; AK-04510-S/10801902192230, 10801902193275, 10801902196719/33F24A0337, 33F24A0338, 33F24A0339, 33F24A0341, 33F24A0342, 33F24A0343, 33F24A1012, 33F24B0107, 33F24B0108, 33F24B0109, 33F24B0110, 33F24B0111, 33F24C0010, 33F24C0011, 33F24D0015, 33F24D0016, 33F24E0046, 33F24E0047, 33F24G0327, 33F24G0570, 33F24G0571, 33F24G0572, 33F24G0573, 33F24G0574, 33F24G0575, 33F24J0493, 33F24J0674, 33F24J0675, 33F24J0676, 33F24J0677, 33F24K0091, 33F24K0309, 33F24K0313, 33F24L0623, 33F24L0627, 33F24L0629, 33F24L0631, 33F24L0754, 33F24L0783, 33F25A0457, 33F25A0613, 33F25A0614, 33F25A0615, 33F25A0616, 33F25C0738, 33F25C0751, 33F25C0794, 33F25C0846, 33F25C0907, 33F25C0955, 33F25C0956, 33F25E0040, 33F25E0288, 33F25E0289, 33F25E0290, 33F25E0910, 33F25E1065, 33F25E1066, 33F25G0079, 33F25G0158, 33F25H0012, 33F25H0193, 33F25H0194, 33F25H0195, 33F25H0196, 33F25J0464, 33F25J0642, 33F25J0789, 33F25J0801, 33F25J0802, 33F25J0803, 33F25J0804, 33F25L0486, 33F25L0488, 33F26B0668, 33F26B0669, 33F26B0671, 33F26B0672, 33F23L0360; ASK-04018-HUMC/10801902193275/33F25B0360, 33F25K0050, 33F25M0680, 33F25E1176; ASK-04018-OH1/10801902117622, 10801902193275, 10801902117622, 10801902117639/33F24F0741, 33F24L0218, 33F24L0865, 33F25B0173, 33F25C0253, 33F25E0667, 33F25F0162, 33F25J0143, 33F26A1440, 33F26B0794, 33F24A0112; ASK-04020-BID/10801902211429, 10801902193275, 10801902211429/33F24D0870, 33F24E0650, 33F24F0646, 33F24L0220, 33F24L0399, 33F25B0175, 33F25D0335, 33F25E0668, 33F25G0248, 33F25L0123, 33F26A0254, 33F26A1441, 33S23J0777, 33F24G0405; ASK-04020-MAR/10801902193275, 10801902159776/33F24J0567, 33F24L0586, 33F25A0519, 33F25E0136, 33F25E0979, 33F25G0675, 33F25J0504, 33F25L0210, 33F26B0742, 33F25K0032; ASK-04020-PM4/10801902138498, 10801902193275, 10801902138498, 30801902158896/33F24E0428, 33F24E0429, 33F24G0424, 33F24K0188, 33F24L0789, 33F25C0142, 33F25C0821, 33F25E0142, 33F25H0085, 33F25J0883, 33F25L0470, 33F26B0659, 33F24F0095; ASK-04120-HF3/10801902200447, 10801902193275, 10801902200447, 10801902117943/33F24D0857, 33F24F0638, 33F24H0168, 33F24K0389, 33F24L0289, 33F25D0336, 33F25F0126, 33F25F0163, 33F25H0616, 33F25J0202, 33F25M0158, 33F25M0362, 33F26A1429, 33F26B0127, 33S24A1516, 33S24C0854, 33F24A0136; ASK-04120-UR/10801902117943, 10801902193275, 10801902117943/33F24A0769, 33F24B0278, 33F24C0175, 33F24D0140, 33F24E0447, 33F24G0287, 33F24K0201, 33F25B0692, 33F25C0962, 33F25E0156, 33F25H0097, 33F25J0889, 33F25L0383, 33F26C0309, 33F24G0441; ASK-04125-MGH/10801902193275, 10801902139976/33F24J0584, 33F25C0985, 33F25D0307, 33F25E0157, 33F25E0990, 33F25H0098, 33F25J0519, 33F25L0421, 33F26A1057, 33F26B0983, 33F24K0203; ASK-04500-HFH4/10801902200461, 10801902193275, 10801902118001/33F24E0651, 33F24F0722, 33F24L0292, 33F25C0437, 33F25D0228, 33F25E0775, 33F25F0109, 33F25F0164, 33F25H0615, 33F25M0345, 33F26A1363, 33F26B0405, 33S24C0436, 33S24L0292, 33F24A0141; ASK-04510-BID2/10801902193275, 10801902211412, 10801902115024/33F24J0171, 33F25A0165, 33F25C0438, 33F25D0229, 33F25E0852, 33F25G0249, 33F25J0204, 33F25M0233, 33F26A1443, 33F26B0410, 33F24F0820; ASK-04510-EHP1/10801902193275, 10801902115024/33F24L0293, 33F25C0192, 33F25D0230, 33F25E0589, 33F25E0853, 33F25H0646, 33F25L0125, 33F26A0255, 33F26A1520, 33F26B0838, 33F24J0303; ASK-04510-FEM1/10801902214796, 10801902193275/33F24A0831, 33F24B0286, 33F24C0185, 33F24E0456, 33F24E0457, 33F24G0448, 33F24J0304, 33F24L0771, 33F25A0539, 33F25E0165, 33F25G0118, 33F25J0523, 33F26A1051, 33F26B0658, 33F24J0591; ASK-04510-HFH/10801902115031, 10801902193275, 10801902200454/33F24G0053, 33F24L0022, 33F24M0108, 33F24M0190, 33F25A0126, 33F25A0356, 33F25A0384, 33F25B0043, 33F25D0200, 33F25H0329, 33F25H0848, 33F25J0613, 33F26A1358, 33F26B0677, 33F24D0865; ASK-04510-HFH4/10801902200454, 10801902193275/33F24F0865, 33F24J0167, 33F24L0965, 33F25A0236, 33F25C0260, 33F25E0776, 33F25G0332, 33F25J0205, 33F25M0276, 33F26A0305, 33F26B0194, 33F24J0306; ASK-04510-SE/10801902193275, 20801902214038, 10801902217018/33F24K0213, 33F24L0592, 33F25B0251, 33F25C0159, 33F25E0171, 33F25J0527, 33F25L0333, 33F26B0664, 33F24A0873; ASK-04510-TGH2/10801902217018, 10801902193275, 10801902118070/33F24B0292, 33F24C0188, 33F24E0463, 33F24G0289, 33F24K0214, 33F24L0633, 33F25A0542, 33F25C0879, 33F25E0172, 33F25G0120, 33F25J0528, 33F25L0354, 33F25M0862, 33F24C0189; NA-04550-1A/30801902117367, 10801902193275, 33F24A1562, 33F24C0709, 33F24D0331, 33F24E0628, 33F24F0946, 33F24K0382, 33F24L0144, 33F24L1015, 33F25C0223, 33F25D0320, 33F25E0512, 33F25E0663, 33F25E0839, 33F25J0187, 33F25K0199, 33F24K0374; NA-04550-X1A/10801902193275, 30801902117374, 10801902119930/33F24K0704, 33F24L0145, 33F24L1138, 33F25C0465, 33F25C0472, 33F25D0633, 33F25G0387, 33F25J0188, 33F25K0200, 33F26C0354, 33F26C0763, 33F24A1326; AK-04018/10801902190915, 10801902193275, 10801902190939/33F24E0696, 33F24L0051, 33F25A0353, 33F25B0078, 33F25D0714, 33F25G0732, 33F25J0593, 33F26B0039, 33F24A1251; AK-04020/10801902190939, 10801902193275, 60801902121549, 10801902121551/33F24C0727, 33F24C1254, 33F24D0439, 33F24E0697, 33F24F0010, 33F24G0049, 33F24G0116, 33F24G0220, 33F24G0645, 33F24G0646, 33F24G0680, 33F24H0064, 33F24J0013, 33F24J0457, 33F24K0033, 33F24M0004,33F24M0052, 33F24M0070, 33F24M0103,33F24M0146,33F25A0084, 33F25A0335, 33F25A0346, 33F25A0362, 33F25A0397, 33F25B0080, 33F25B0307, 33F25B0414, 33F25C0058, 33F25C0097, 33F25C0494, 33F25C0575, 33F25D0027, 33F25D0083, 33F25D0102, 33F25D0296, 33F25D0703, 33F25F0293, 33F25F0398, 33F25G0493, 33F25H0322, 33F25H0842, 33F25J0037, 33F25J0767, 33F25K0276, 33F25L0154, 33F25L0169, 33F25L0218, 33F26A1264, 33F26A1290, 33F26A1313, 33F26B0291, 33F24C0728, 33F24C0729, 33F24E0699, 33F24G0596, 33F24H0057, 33F24J0350, 33F24J0458, 33F24K0615, 33F24M0126, 33F25A0338, 33F25B0039, 33F25B0131, 33F25D0192, 33F25D0281, 33F25E1109, 33F25F0496, 33F25J0764, 33F26B0054, 33F23M0173; AK-04020-C/60801902121549, 10801902193275, 10801902121551/33F24G0050, 33F24G0595, 33F24G0619, 33F24G0681, 33F24H0060, 33F24H0085, 33F24H0244, 33F24J0421, 33F24J0471, 33F24L0074, 33F24L0517, 33F24M0051, 33F24M0102, 33F24M0145,33F25A0354,33F25B0013, 33F25B0079, 33F25B0130, 33F25B0279, 33F25C0096, 33F25C0411, 33F25C0523, 33F25C0595, 33F25D0014, 33F25D0082, 33F25D0169, 33F25E0404, 33F25E1088, 33F25E1164, 33F25F0430, 33F25J1033, 33F25K0343, 33F25L0175, 33F25M0472, 33F26A0212, 33F26A1275, 33F26B0053, 33F26B0290, 33F23M0171; AK-04120/10801902190922, 10801902193275, 20801902121572/33F24A1254, 33F24B0432, 33F24C0731, 33F24D0440, 33F24E0701, 33F24H0444, 33F24J0006, 33F24K0623, 33F24L0035, 33F24M0127, 33F25A0129, 33F25B0057, 33F25B0328, 33F25E1096, 33F25F0405, 33F25F0522, 33F25G0511, 33F25J1067, 33F26C0419, 33F24C0789; AK-04122/20801902121572, 10801902193275, 10801902121582/33F24G0051, 33F24G0148, 33F24G0682, 33F24K0535, 33F24M0071, 33F25A0092, 33F25A0380, 33F25B0005, 33F25D0108, 33F25J0762, 33F25K0003, 33F24A1009; AK-04220/10801902190946, 10801902193275, 10801902121636/33F24C0733, 33F24E0703, 33F24G0117, 33F24G0223, 33F24G0683, 33F24H0246, 33F24H0414, 33F24J0009, 33F24J0447, 33F24M0003, 33F24M0086, 33F25C0085, 33F25D0171, 33F25G0065, 33F25J0059, 33F26C0392, 33F24A1257; ASK-04020-AH/10801902193275, 10801902114898/33F24L0219, 33F24L0398, 33F24L0950, 33F25B0174, 33F25C0436, 33F25D0332, 33F25D0648, 33F25E0848, 33F25H0647, 33F25J0350, 33F26C0572, 33F24G0403; ASK-04020-ATH/10801902193275, 10801902193237, 10801902205916/33F24J0553, 33F24K0167, 33F24L0777, 33F25C0133, 33F25C0982, 33F25E0964, 33F25G0106, 33F25G0670, 33F25H0063, 33F25H0727, 33F25L0242, 33F25L0452, 33F25M0947, 33F26B0583, 33F24D0856; ASK-04020-AU2/10801902205916, 10801902193275, 10801902117660/33F24E0645, 33F24F1044, 33F24K0365, 33F24L1020, 33F25A0230, 33F25C0225, 33F25D0333, 33F25E0523, 33F25E0849, 33F25H0474, 33F25J0199, 33F25L0120, 33F25M0208, 33F26A1419, 33F26B0449, 33F26C0030, 33F24A0814; ASK-04020-BHS/10801902193275, 30801902144578, 10801902211429/33F25A0231, 33F25D0589, 33F25E0850, 33F25F0095, 33F25G0320, 33F25J0200, 33F26A1514, 33R24D0228, 33S24C0678, 33S24C1077, 33F24D0277; ASK-04020-BIDB2/10801902193275/33F24K0168, 33F25A0508, 33F25C0134, 33F25C0835, 33F25E0121, 33F25G0107, 33F25H0065, 33F25H0728, 33F25J0669, 33F25K0399, 33F25L0429, 33F25M0929, 33F24J0365; ASK-04020-CCD1/10801902193275/33F24J0289, 33F24J0556, 33F24L0641, 33F25A0510, 33F25C0772, 33F25E0123, 33F25G0108, 33F25H0730, 33F25M0874, 33F24G0282; ASK-04020-CR/10801902193275, 10801902206814/33F24K0169, 33F24L0679, 33F25A0511, 33F25C0785, 33F25E0124, 33F25E0969, 33F25G0109, 33F25G0669, 33F25H0067, 33F25H0731, 33F25J0868, 33F25L0241, 33F25M0881, 33F26B0523, 33F24G0407; ASK-04020-EUH2/10801902193275, 10801902213379, 10801902193244/33F24J0290, 33F24J0557, 33F24L0753, 33F25A0512, 33F25C0930, 33F25E0125, 33F25H0068, 33F25J0869, 33F25M0894, 33F26B0236, 33F23H0888; ASK-04020-FAM1/10801902193275/ 33F24H0374, 33F24J0181, 33F24L0734, 33F25C0135, 33F25C0781, 33F25E0126, 33F25E0970, 33F25J0502, 33F24G0408; ASK-04020-FMC/10801902193275, 10801902121384/33F24J0558, 33F24K0171, 33F25A0513, 33F25C0773, 33F25E0127, 33F25E0971, 33F25H0069, 33F25J0870, 33F25L0388, 33F25M0892, 33F24A0828; ASK-04020-KSF2/10801902193275/33F25A0516, 33F25C0851, 33F25E0132, 33F25H0075, 33F25J0876, 33F25M0906, 33F24K0178; ASK-04020-LHC2/10801902193275, 10801902225525/33F24L0716, 33F25A0517, 33F25C0734, 33F25E0133, 33F25J0503, 33F25L0491, 33F25M0922, 33F26C0672; ASK-04020-LHC3/10801902193275, 10801902225525/33F24L0716, 33F25A0517, 33F25C0734, 33F25E0133, 33F25J0503, 33F25L0491, 33F25M0922, 33F26C0672, 33F24L0653; ASK-04020-LM/10801902193275/33F25B0690, 33F25E0977, 33F25H0076, 33F25J0877, 33F25L0422, 33F25M0925, 33F24G0417; ASK-04020-MCH1/10801902193275, 20801902117711/33F24L0920, 33F24A0380; ASK-04020-MIH1/10801902193275/33F24L0744, 33F25A0522, 33F25C0978, 33F25E0138, 33F25H0080, 33F25J0879, 33F26A0996, 33F25A0442; ASK-04020-NWMX/10801902193275/33F24K0186, 33F25J0882, 33F25L0363, 33F25M0872, 33F24G0423; ASK-04020-PHP/10801902193275, 10801902138498/33F24K0187, 33F25A0148, 33F25C0401, 33F25D0283, 33F25E1145, 33F25J1027, 33F24C0164; ASK-04020-RIHM/10801902193275/33F24L0243, 33F24L1034, 33F25B0177, 33F25C0256, 33F25D0462, 33F25E0525, 33F25E0773, 33F25G0317, 33F25H0645, 33F25J0351, 33F26A1408, 33F24M0095; ASK-04020-RMC1/10801902193275/33F25J0509, 33F25L0464, 33F24L0747; ASK-04020-RU/10801902193275/33F25A0524, 33F25C0807, 33F25E0143, 33F25G0112, 33F25J0510, 33F25M0863, 33F24G0427; ASK-04020-SE/10801902193275, 10801902213263, 10801902138757/33F24L0611, 33F25A0525, 33F25C0849, 33F25E0144, 33F25E0984, 33F25J0511, 33F26B0999, 33F24E0433; ASK-04020-SHM/10801902193275/33F24G0428, 33F24J0576, 33F24K0190, 33F24L0684, 33F25A0526, 33F25C0971, 33F25E0145, 33F25G0113, 33F25J0512, 33F25L0378, 33F25M0883, 33F25C0143; ASK-04020-SRMC1/10801902193275, 10801902117837/33F24J0296, 33F24K0192, 33F24L0710, 33F25C0144, 33F25C0820, 33F25E0146, 33F25H0086, 33F25J0884, 33F25M0842, 33F24A0868; ASK-04020-UCL1/10801902193275, 10801902144550/33F24K0412, 33F24L1081, 33F25A0233, 33F25C0257, 33F25D0393, 33F25E0774, 33F25G0360, 33F25J0201, 33F26B0160, 33F24G0178; ASK-04020-UMI/10801902193275, 10801902145571/33F24L1123, 33F24E0726; ASK-04020-UN/10801902145571, 10801902193275, 60801902117849/33F24K0021, 33F24L0533, 33F24M0053, 33F24M0107, 33F25A0089, 33F25B0016, 33F25B0092, 33F25B0314, 33F25D0015, 33F25D0113, 33F25D0172, 33F25E1146, 33F25H0797, 33F25K0230, 33F26B0090, 33F26B0262, 33F24A0373; ASK-04020-UNC/60801902117849, 10801902193275, 10801902143782/33F24A1035, 33F24B0270, 33F24C0170, 33F24G0432, 33F24J0297, 33F24J0577, 33F24L0799, 33F25A0527, 33F25C0145, 33F25C0994, 33F25E0147, 33F25H0088, 33F25J0513, 33F25L0359, 33F25M0868, 33F26A0993, 33F26B0975, 33F24F0461; ASK-04020-UNC1/10801902193275, 10801902143782, 10801902117851/33F24G0286, 33F24J0298, 33F24K0194, 33F24L0806, 33F25C0146, 33F25E0148, 33F25H0089, 33F25J0514, 33F25L0469, 33F25M0959, 33F26C0599, 33F24A0255; ASK-04220-PSU/10801902193275/33F25A0152, 33F25H0737, 33F24K0495; ASK-04220-UCL1/10801902193275, 10801902159127, 10801902143805/33F24L0290, 33F25A0206, 33F25C0226, 33F25D0463, 33F25E0669, 33F25G0351, 33F25H0428, 33F25J0352, 33F26B0153, 33F24F0619; NA-04020-S1A/10801902193275, 10801902139396/33F25A0600, 33F25C0195, 33F25C0786, 33F25E0270, 33F25E1050, 33F25G0152, 33F25H0177, 33F25H0759, 33F25J0633, 33F25K0404, 33F25L0472, 33F25M0962, 33F26B0595, 33F24J0683; ASK-04550-JMC1/10801902193275, 10801902130836/33F25C0837, 33F25E0175, 33F25E0997, 33F25L0146, 33F24F0685; ASK-04001-BW2/10801902193275, 10801902221602/33F25A0500, 33F25E0110, 33F25H0057, 33F25J0496, 33F25L0360, 33F26A0994, 33S23M0157, 33F25A0226; ASK-04001-WUH1/10801902193275, 10801902220834, 10801902119268/33F24J0549, 33F24K0162, 33F24L0685, 33F25H0061, 33F25J0864, 33F26B0683, 33F26B1021, 33F24A0078; ASK-04510-HUM/10801902115048/33F24C0956, 33F24D0699, 33F24E0647, 33F24F0911, 33F24L0885
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
785,999