FDA Enforcement Class I Ongoing

Description/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS; STIMUCATH SFTY KIT: 20G CTH, 18Gx8CM NDL/AB-20608-KS; STIMUCATH KIT: 19G CATH, 17G x 9CM NDL/AB-05060-PK; FLEXBLOCK ULTRASOUND KIT: 17Gx9CM NS NDL/ASK-19609-VCU; CONTINUOUS NERVE BLOCK KIT/ASK-19600-CC2; CONTINUOUS NERVE BLOCK KIT 19 GA X 60/ASK-19608-CC; CONTINUOUS NERVE BLOCK KIT 19GA X 60CM/ASK-19608-KP; CONTINUOUS NERVE BLOCK KIT 19GA X 60CM/ASK-19608-SHC; FLEXBLOCK KIT:ECHO CATH, 17Gx8CM STM NDL/FB-19608-K; FLEXBLOCK ULTRASOUND KIT: 17Gx8CM STM ND/FB-19608-KU; FLEXBLOCK KIT:ECHO CATH, 17Gx9CM NS NDL/FB-19609-K; FLEXBLOCK ULTRASOUND KIT: 17Gx9CM NS NDL/FB-19609-KU; CONTINUOUS PERIPHERAL NERVE BLOCK KIT/ASK-19608-US1; EPIDURAL CATHETERIZATION KIT/ASK-05001-SLR1; EPIDURAL CATHETERIZATION KIT/MH-05001; EPIDURAL CATHETERIZATION KIT/WJ-05401; SPINAL ANES/EPIDURAL CATH 19GA X 90CM/ASK-05501-USC1; EPIDURAL NEEDLE KIT/LG-03000; EPIDURAL ANESTHESIA KIT/MTO-03000-VM; EPIDURAL ANESTHESIA KIT/SM-03001; Anesthesia conduction kit/FB-19611-K

Recall: Z-2364-2026 · Reported June 24, 2026

Enforcement

Recall Number
Z-2364-2026
Event ID
98793
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
ARROW INTERNATIONAL, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 24, 2026
Initiation Date
April 21, 2026
Classification Date
June 12, 2026
Address
3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437, United States

Description

Description/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS; STIMUCATH SFTY KIT: 20G CTH, 18Gx8CM NDL/AB-20608-KS; STIMUCATH KIT: 19G CATH, 17G x 9CM NDL/AB-05060-PK; FLEXBLOCK ULTRASOUND KIT: 17Gx9CM NS NDL/ASK-19609-VCU; CONTINUOUS NERVE BLOCK KIT/ASK-19600-CC2; CONTINUOUS NERVE BLOCK KIT 19 GA X 60/ASK-19608-CC; CONTINUOUS NERVE BLOCK KIT 19GA X 60CM/ASK-19608-KP; CONTINUOUS NERVE BLOCK KIT 19GA X 60CM/ASK-19608-SHC; FLEXBLOCK KIT:ECHO CATH, 17Gx8CM STM NDL/FB-19608-K; FLEXBLOCK ULTRASOUND KIT: 17Gx8CM STM ND/FB-19608-KU; FLEXBLOCK KIT:ECHO CATH, 17Gx9CM NS NDL/FB-19609-K; FLEXBLOCK ULTRASOUND KIT: 17Gx9CM NS NDL/FB-19609-KU; CONTINUOUS PERIPHERAL NERVE BLOCK KIT/ASK-19608-US1; EPIDURAL CATHETERIZATION KIT/ASK-05001-SLR1; EPIDURAL CATHETERIZATION KIT/MH-05001; EPIDURAL CATHETERIZATION KIT/WJ-05401; SPINAL ANES/EPIDURAL CATH 19GA X 90CM/ASK-05501-USC1; EPIDURAL NEEDLE KIT/LG-03000; EPIDURAL ANESTHESIA KIT/MTO-03000-VM; EPIDURAL ANESTHESIA KIT/SM-03001; Anesthesia conduction kit/FB-19611-K

Reason

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Code Info

REF(Material)/UDI-DI/Lot(Batch): ASK-04001-SUH-S/10801902210217, 10801902193275/33F25C0766, 33F25E0112, 33F25G0101, 33F25H0058, 33F25H0725, 33F25J0862, 33F25M0836, 33F25F0160; ASK-04001-VM1/10801902210217, 10801902202106, 10801902193275/33F24L0599, 33F26A0962, 33F24G0280; AB-19604-KS/10801902210217, 10801902209471/33F24L0608, 33F25G0474, 33F25J0814, 33F25M0887, 33F24D0064; AB-19608-KS/10801902210217, 10801902209471/33F24E0312, 33F24F0295, 33F24G0316, 33F24K0082, 33F24L0714, 33F25E0027, 33F25E0904, 33F25H0003, 33F25J0815, 33F25L0337, 33F26B0498, 33F24E0314; AB-20608-KS/)10801902210217/33F25H0004, 33F25J0816, 33F24D0437; AB-05060-PK/10801902209433, 10801902210217/33F24D0002, 33F24E0029, 33F24G0315, 33F24J0484, 33F24K0081, 33F25A0451, 33F25C0747, 33F25E0026, 33F25J0646, 33F24E0028; ASK-19609-VCU/10801902210217, 10801902206876, 10801902214291/33F24E0536, 33F24J0630, 33F24L0605, 33F25C0179, 33F25E0220, 33F25H0140, 33F26B1003, 33F24D0628; ASK-19600-CC2/10801902210057, 10801902210217, 10801902210064/33F24D0393, 33F24E0533, 33F24G0494, 33F24K0251, 33F24L0665, 33F25C0177, 33F25E0217, 33F25H0138, 33F24C0230; ASK-19608-CC/10801902210217, 10801902210064, 10801902202236/33F24G0309, 33F24K0252, 33F24L0639, 33F25B0691, 33F25C0178, 33F25C0761, 33F25E0218, 33F25E1022, 33F25H0750, 33F25J0918, 33F26B0230, 33F24D0394; ASK-19608-KP/10801902210217/33F24G0495, 33F24G0086; ASK-19608-SHC/10801902210217, 40801902202220, 10801902202243/33F24G0496, 33F24J0629, 33F24K0253, 33F24L0573, 33F25A0571, 33F25C0733, 33F25E0219, 33F25G0757, 33F25H0139, 33F25J0919, 33F25K0400, 33F25L0480, 33F26B0606, 33F24C0231; FB-19608-K/10801902210217, 10801902210125/33F24G0533, 33F24J0650, 33F24K0279, 33F24L0568, 33F25C0186, 33F25C0748, 33F25E0257, 33F25G0147, 33F25H0164, 33F25J0623, 33F25L0340, 33F25M0847, 33F24C0258; FB-19608-KU/10801902210217, 10801902210132/33F24E0026, 33F24E0574, 33F24G0312, 33F24J0328, 33F24L0557, 33F24M0268, 33F25C0187, 33F25C0940, 33F25E0258, 33F25H0165, 33F25J0624, 33F25L0344, 33F25M0851, 33F23G0580; FB-19609-K/10801902210132, 10801902210217, 10801902210132, 10801902210149/33F24C0259, 33F24D0407, 33F24E0575, 33F24G0534, 33F24K0281, 33F24L0805, 33F25C0721, 33F25E1042, 33F25J0625, 33F25L0351, 33F25M0856, 33F26B0504, 33F24B0386; FB-19609-KU/10801902210149, 10801902210217, 10801902210149, 10801902193275/33F24C0260, 33F24E0576, 33F24E0577, 33F24G0535, 33F24L0763, 33F25A0592, 33F25E0259, 33F25E1043, 33F25H0758, 33F25K0402, 33F25L0471, 33F26B0594, 33F26C0480, 33S24L0763, 33F24L0738; ASK-19608-US1/10801902210217/33F24E0535, 33F24K0254, 33F24L0770, 33F24E0014; ASK-05001-SLR1/10801902210217, 10801902209617/33F24J0596, 33F24L0560, 33F25A0546, 33F25C0939, 33F25E0177, 33F25E0998, 33F25J0530, 33F24C0194; MH-05001/10801902210217/33F24G0540, 33F24J0331, 33F24J0654, 33F25E0261, 33F25H0168, 33F25J0937, 33F25L0478, 33F24K0286; WJ-05401/10801902210385, 10801902210217, 10801902210385/33F24B0166, 33F24C0054, 33F24J0672, 33F24K0307, 33F24L0548, 33F25A0609, 33F25A0625, 33F25C0723, 33F25E0285, 33F25E1062, 33F25G0156, 33F25H0187, 33F25J0640, 33F25J0952, 33F25L0414, 33F26B0559, 33F26C0858; ASK-05501-USC1/10801902210217, 10801902214857, 10801902209853/33F24E0506, 33F24E0507, 33F24G0478, 33F24L0561, 33F25A0561, 33F25C0167, 33F25C0769, 33F25E0200, 33F25G0132, 33F25H0745, 33F25J0908, 33F25L0412, 33F26B0557, 33F24C0211; LG-03000/10801902210217/33F24L0668, 33F24E0581; MTO-03000-VM/10801902210217, 10801902202076, 10801902119916/33F24E0586, 33F24G0547, 33F24J0658, 33F24K0291, 33F24L0632, 33F25A0596, 33F25C0869, 33F25E0265, 33F25H0173, 33F25J0939, 33F25L0380, 33F26C0306, 33F24A1057; SM-03001/10801902210217, 10801902202496, 10801902210354/33F24E0779, 33F24J0243, 33F24J0434, 33F24M0012, 33F25A0099, 33F25B0035, 33F25B0128, 33F25C0410, 33F25C0521, 33F25F0421, 33F25G0531, 33F25J0053, 33F25L0171, 33F26B0643, 33F24C0276; FB-19611-K/10801902210156/33F24F0014

Distribution

US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.

Quantity

205,685