FDA Enforcement Class II Ongoing

UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a cystoscope shaft into the bladder. Model: UL-VO

Recall: Z-2364-2023 · Reported August 16, 2023

Enforcement

Recall Number
Z-2364-2023
Event ID
92700
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Scholly Fiberoptic Gmbh
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 16, 2023
Initiation Date
June 23, 2023
Classification Date
August 9, 2023
Address
Robert-Bosch-Str. 1-3, N/A, Denzlingen, N/A, N/A, Germany

Description

UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a cystoscope shaft into the bladder. Model: UL-VO

Reason

There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.

Code Info

UDI-DI: 04250480106617 Lot Number: 141980

Distribution

Nationwide Foreign: Japan, South Korea

Quantity

98 units