lnvacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16. Packed one per carton. The lnvacare Precise Rx Pediatric Flowmeter adapts to lnvacare Concentrators for low-flow usage.The gauge has a scale printed on the front side and the back side. The front scale is the primary source for monitoring oxygen flow. The rear scale is an aide for monitoring oxygen flow.
Enforcement
- Recall Number
- Z-2364-2016
- Event ID
- 74581
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Invacare Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 17, 2016
- Initiation Date
- June 21, 2016
- Classification Date
- August 8, 2016
- Termination Date
- February 7, 2017
- Address
- 1 Invacare Way, Elyria, OH, 44035-4190, United States
Description
lnvacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16. Packed one per carton. The lnvacare Precise Rx Pediatric Flowmeter adapts to lnvacare Concentrators for low-flow usage.The gauge has a scale printed on the front side and the back side. The front scale is the primary source for monitoring oxygen flow. The rear scale is an aide for monitoring oxygen flow.
The rear gradation scale was screen printed upside down on the flowmeter gauge.
Serial numbers for Model IRCPF16: 16BF023724, 16BF023729, 16BF023734, 16BF023739, 16BF023744,16BF023720, 16BF023725, 16BF023730, 16BF023735, 16BF023740, 16BF023745,16BF023721, 16BF023726, 16BF023731, 16BF023736, 16BF023741, 16BF023746,16BF023722, 16BF023727, 16BF023732, 16BF023737, 16BF023742, 16BF023747,16BF023723, 16BF023728, 16BF023733, 16BF023738, 16BF023743, 16BF023748,16BF023749, 16BF030593, 16CF015205, 16CF019539, 16CF026064, 16CF030100,16BF023750, 16BF030594, 16CF015206, 16CF019540, 16CF026065, 16CF030101,16BF023751, 16CF016169, 16CF015207, 16CF019541, 16CF026066, 16CF030102,16BF023752, 16CF016170, 16CF015208, 16CF019542, 16CF026067, 16CF030103,16BF023753, 16CF016171, 16CF015209, 16CF019543, 16CF026068,16CF030104,16BF023754, 16CF016172, 16CF015210, 16CF019544, 16CF026069, 16CF030105,16BF023755, 16CF016173, 16CFOI5211, 16CF019545, 16CF026070, 16CF030106,16BF025606, 16CF016174, 16CF015212, 16CF019546, 16CF026071, 16CF030107,16BF025607, 16CF016175, 16CFOI5213, 16CF019547, 16CF026072, 16CF030108,16BF025608, 16CF016176, 16CF015214, 16CF019548, I6CF026073, 16CF030109,16BF025609, 16CF016177, 16CF015215, 16CF019549, l6CF026074, 16CF030110,16BF025610, 16CF016178, 16CF015216, l6CF019550, l6CF026075, 16CF0301ll,16BF025611, 16CF016179, 16CF018612, 16CF019551, 16CF026076, 16CF030112,16BF025612, 16CF016180, 16CF018613, 16CF019552, l6CF026077, 16CF030113,16BF025613, 16CF016181, 16CF018614, 16CF019553, 16CF026078, 16CF030114,16BF025614, 16CF016182, 16CF018615, 16CF019554, 16CF026079, 16CF030115,16BF025615, 16CF016183, 16CF018616, 16CF019555, 16CF026080, l6CF030116,16BF025616, 16CF016184, 16CF018617, 16CF019556, 16CF026081, 16CF030117,16BF025617, 16CF016185, 16CF018618, 16CF019557, 16CF026082, 16CF030118,16BF025618, 16CF016186, 16CF018619, 16CF019558, 16CF026083, 16CF030119,16BF025619, 16CF016187, 16CFOI8620, 16CF019559, 16CF026084, 16CF030120,16BF025620, 16CF016188, 16CF018621, 16CF019560, 16CF026085, 16CF030121,16BF025621, 16CF016189, 16CF018622, 16CF019561, 16CF026086, 16CF030122,16BF025622, 16CF016190, 16CF018623, 16CF019562, 16CF026087, 16CF030123,16BF025623, 16CF016191, 16CF018624, 16CF019563, 16CF026088, 16CF030124,16BF025624, 16CF016192, 16CF018625, 16CF019564, 16CF026089, 16CF032012,16BF025625, 16CF016193, 16CF018626, 16CF019565, 16CF026090, 16CF032013,16BF025626, 16CF016194, 16CF018627, 16CF019566, 16CF026091, 16CF032014,16BF025627, 16CF016195, 16CF018628, 16CF023329, 16CF026092, 16CF032015,16BF025628, 16CF016196, 16CF018629, 16CF023330, 16CF027673, 16CF032016,16BF025629, 16CF016197, 16CF018630, 16CF023331, 16CF027674, 16CF032017,16BF025630, 16CF016198, 16CF018631, 16CF023332, 16CF027675, 16CF032018,16BF025631, 16CF015181, 16CF018632, 16CF023333, 16CF027676, 16CF032019,16BF025632, 16CF015182, 16CF018633, 16CF023334, 16CF027677, 16CF032020,16BF025633, 16CF015183, 16CF018634, 16CF023335, 16CF027678, 16CF032021,16BF025634, 16CF015184, 16CF018635, 16CF023336, 16CF027679, 16CF032022,16BF025635, 16CF015185, 16CF018636, 16CF023337, 16CF027680, 16CF032023,16BF025636, 16CF015186, 16CF018637, 16CF023338, 16CF027681, 16CF032024,16BF025637, 16CF015187, 16CF018638, 16CF023339, 16CF027682, 16CF032025,16BF025638, 16CF015188, 16CF018639, 16CF023340, 16CF027683, 16CF032026,16BF025639, 16CF015189, 16CFOI8640, 16CF023341, 16CF027684, 16CF032027,16BF025640, 16CF015190, 16CF018641, 16CF023342, 16CF027685, 16CF032028,16BF025641, 16CF015191, 16CF018642, 16CF023343, 16CF027686, 16CF032029,16BF027950, 16CF015192, 16CF018643, 16CF023344, 16CF027687, 16CF032030,16BF027951, 16CF015193, 16CF018644, 16CF023345, 16CF027688, 16CF040048,16BF027952, 16CF015194, 16CF018645, 16CF023346, 16CF030089, 16CF040049,16BF030583, 16CF015195, 16CF018646, 16CF023347, 16CF030090, 16CF040050,16BF030584, 16CF015196, 16CF018647, 16CF023348, 16CF030091, 16CF040051,16BF030585, 16CF015197, 16CF019531, 16CF023349, 16CF030092, 16CF040052,16BF030586, 16CF015198, 16CF019532, 16CF026057, 16CF030093, 16CF040053,16BF030587, 16CF015199, 16CF019533, 16CF026058, 16CF030094, 16CF040054,16BF030588, 16CF015200, 16CF019534, 16CF026059, 16CF030095, 16CF040055,16BF030589, 16CF015201, 16CF019535, 16CF026060, 16CF030096, 16CF040056,16BF030590, 16CF015202, 16CF019536, 16CF026061, 16CF030097, 16CF040057,16BF030591, 16CF015203, 16CF019537, 16CF026062, 16CF030098, 16CF040058,16BF030592, 16CF015204, 16CF019538, 16CF026063, 16CF030099, 16CF040059,16CF040060,16CF040061,16DF017166, 16DF017167, 16DF017910,16DF017911, 16DF040084 16DF040085, 16DF040632,16DF040633, 16DF045327,16DF045328,16CF040062, 16DF017168, 16DF017912, 16DF040086, 16DF040634, 16DF045329,16CF040063, 16DF017169, 16DF017913, 16DF040087, 16DF040635, 16DF045330,16CF040064, 16DF017170, 16DF017914, 16DF040088, 16DF040636, 16DF045331,16CF040065, 16DF017171, 16DF017915, 16DF040089, 16DF040637, 16DF045332,16CF040066, 16DF017172, 16DF017916, 16DF040090, 16DF040638, 16DF045333,16CF040067, 16DF017173, 16DF017917, 16DF040091, 16DF040639, 16DF045334,16CF040068, 16DF017174, 16DF017918, 16DF040092, 16DF040640, 16DF045335,16CF040069, 16DF017175, 16DF017919, 16DF040093, 16DF040641, 16DF045336,16CF040070, 16DF017176, 16DF017920, 16DF040094, 16DF040642, 16DF045337,16CF040071, 16DF017177, 16DF017921, 16DF040095, 16DF040643, 16DF045338,16CF040072, 16DF017178, 16DF017922, 16DF040096, 16DF040644, 16DF045339,16CF040073, 16DF017179, 16DF017923, 16DF040097, 16DF040645, 16DF045340,16CF040074, 16DF017180, 16DF017924, 16DF040098, 16DF040646, 16DF045341,16CF040075, 16DF017181, 16DF017925, 16DF040099, 16DF040647, 16DF045342,16CF040076, 16DF017182, 16DF017926, 16DF040100, 16DF040648, 16DF045343,16CF040077, 16DF017183, 16DF017927, 16DF040101, 16DF040649, 16EF000999,16CF040078, 16DF017184, 16DF017928, 16DF040102, 16DF040650, 16EF001000,16CF040079, 16DF017185, 16DF017929, 16DF040103, 16DF041832, 16EF001001,16CF040080, 16DF017186, 16DF019605, 16DF040104, 16DF041833, 16EF001002,16CF040081, 16DF017187, 16DF019606, 16DF040615, 16DF041834, 16EF001003,16CF040082, 16DF017894, 16DF019607, 16DF040616, 16DF045311, 16EF001004,16CF040083, 16DF017895, 16DF040069, 16DF040617, 16DF045312, 16EF001005,16DF017152, 16DF017896, 16DF040070, 16DF040618, 16DF045313, 16EF001006,16DF017153, 16DF017897, 16DF040071, 16DF040619, 16DF045314, 16EF001007,16DF017154, 16DF017898, 16DF040072, 16DF040620, 16DF045315, 16EF001008,16DF017155, 16DF017899, 16DF040073, 16DF040621, 16DF045316, 16EF001009,16DF017156, 16DF017900, 16DF040074, 16DF040622, 16DF045317, 16EF001010,16DF017157, 16DF017901, 16DF040075, 16DF040623, 16DF045318, 16EF001011,16DF017158, 16DF017902, 16DF040076, 16DF040624, 16DF045319, 16EF001012,16DF017159, 16DF017903, 16DF040077, 16DF040625, 16DF045320, 16EF001013,16DF017160, 16DF017904, 16DF040078, 16DF040626, 16DF045321, 16EF001014,16DF017161, 16DF017905, 16DF040079, 16DF040627, 16DF045322, 16EF001015,16DF017162, 16DF017906, 16DF040080, 16DF040628, 16DF045323, 16EF001016,16DF017163, 16DF017907, 16DF040081, 16DF040629, 16DF045324, 16EF001017,16DF017164, 16DF017908, 16DF040082, 16DF040630, 16DF045325, 16EF001020,16DF017165, 16DF017909, 16DF040083, 16DF040631, 16DF045326, 16EF001021.
Worldwide Distribution - US to CA, CO, CT, FL, GA, IA, ID, IL,IN,KS,KY,LA,MA,MD,MI,MS,MT,NC,NE,NJ,NM,NY,OH,PA,SC, TN,TX,UT,VA,WA and Internationally to Canada, Germany and New Zealand.