ARTIS icono biplane -Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327600

Recall: Z-2359-2023 · Reported August 16, 2023

Enforcement

Recall Number
Z-2359-2023
Event ID
92653
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 16, 2023
Initiation Date
June 19, 2023
Classification Date
August 8, 2023
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

ARTIS icono biplane -Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327600

Reason

Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system

Code Info

UDI-DI: 4056869063317 Serial Numbers: 180550 180517 180518 180546 180547 180560 180534 180556 180537 180545 180533 180566 180536 180563

Distribution

Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Italy, Japan, Latvia, Luxembourg, Macedonia, Mexico, Monaco, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.

Quantity

14 units