FDA Enforcement Class II Terminated

MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.

Recall: Z-2359-2019 · Reported September 4, 2019

Enforcement

Recall Number
Z-2359-2019
Event ID
82794
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Agilent Technologies, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
September 4, 2019
Initiation Date
August 15, 2018
Classification Date
August 23, 2019
Termination Date
September 28, 2021
Address
5301 Stevens Creek Blvd, N/A, Santa Clara, CA, 95051-7201, United States

Description

MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.

Reason

When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.

Code Info

Software versions B.07 or below

Distribution

US Nationwide. distribution.

Quantity

235