FDA Enforcement Class II Terminated

microtargeting" Depth Stop Adapter, Product Model Number, Product Number 66-AC-DS(1.8),

Recall: Z-2359-2018 · Reported July 11, 2018

Enforcement

Recall Number
Z-2359-2018
Event ID
80148
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
FHC, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 11, 2018
Initiation Date
May 10, 2018
Classification Date
July 5, 2018
Termination Date
April 30, 2019
Address
1201 Main St, Bowdoin, ME, 04287-7302, United States

Description

microtargeting" Depth Stop Adapter, Product Model Number, Product Number 66-AC-DS(1.8),

Reason

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

Code Info

UDI: 00873263006375, 00873263001493

Distribution

The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.

Quantity

2952 total