BD FACS Sample Prep Assistant (SPA) III - Product Usage: Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II and BD FACSCalibur flow cytometry systems. Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto II flow cytometry systems

Recall: Z-2358-2020 · Reported June 17, 2020

Enforcement

Recall Number
Z-2358-2020
Event ID
85129
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Becton, Dickinson and Company, BD Biosciences
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 17, 2020
Initiation Date
July 8, 2019
Classification Date
June 11, 2020
Termination Date
January 31, 2023
Address
2350 Qume Dr, N/A, San Jose, CA, 95131-1812, United States

Description

BD FACS Sample Prep Assistant (SPA) III - Product Usage: Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II and BD FACSCalibur flow cytometry systems. Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto II flow cytometry systems

Reason

The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.

Code Info

Catalog (Ref) No. 647205

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AL, CO,CT, GA, IL, MA, MI, MD, NC, NH,NJ, NY, OH, PA, TN, TX, VA, and VT. The countries of Australia, Botswana, Brazil, Canada, Chile, China, Colombia, Costa Rica, Europe, Hong Kong, India, Italy, Jamaica, Japan, Kazakhstan, Malaysia, Martinique, Mexico, New Zealand, Nicaragua, Peru, Puerto Rico (US), Romania, Russia, Singapore, South Africa, South Korea, Swaziland, Switzerland, Turkey, US, Vietnam.

Quantity

467 units