Invacare¿ Pronto¿ Air Personal Transporter (power wheelchair), Invacare Corporation, One Invacare Way, Elyria, Ohio, 44035-4190, (800) 333 -6900.
Enforcement
- Recall Number
- Z-2358-2015
- Event ID
- 71613
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Invacare Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 19, 2015
- Initiation Date
- February 2, 2015
- Classification Date
- August 10, 2015
- Termination Date
- September 14, 2015
- Address
- 1 Invacare Way, N/A, Elyria, OH, 44035-4190, United States
Description
Invacare¿ Pronto¿ Air Personal Transporter (power wheelchair), Invacare Corporation, One Invacare Way, Elyria, Ohio, 44035-4190, (800) 333 -6900.
The batteries used in the Invacare¿ Pronto¿ Air Personal Transporter models power wheelchairs have the possibility to overheat during charging which could potentially result in leaking battery acid and/or swelling battery, loss of use, chemical burns and/or property damage.
Model #AIRPTMBDY18, Serial #s: 12JSZD60025, 12JSZD60026, 12JSZD60027, 12KSZD60001, 12KSZD60002, 12KSZD60004, 12KSZD60005, 12KSZD60006, 12KSZD60007, 12KSZD60008, 12KSZD60010, 12KSZD60013, 12KSZD60014, 12KSZD60015, 12KSZD60016, 12KSZD60017, 12KSZD60018, 12KSZD60019, 12KSZD60020, 12KSZD60021, 12KSZD60022, 12KSZD60023, 12KSZD60024, 12KSZD60026, 12KSZD60028, 12KSZD60029, 12KSZD60030, 12KSZD60031, 12KSZD60032, 12KSZD60033, 12KSZD60034, 12KSZD60035, 12KSZD60036, 12KSZD60037, 12KSZD60038, 12KSZD60039, 12KSZD60040, 12KSZD60041, 12KSZD60042, 12KSZD60043, 12KSZD60045, 12KSZD60046, 12KSZD60047, 12KSZD60048, 12KSZD60049, 12KSZD60050, 12KSZD60051, 12KSZD60052, 12KSZD60053, 12KSZD60054, 12KSZD60055, 12KSZD60056, 12KSZD60057, 12KSZD60058, 12KSZD60059, 12KSZD60060, 12KSZD60061, 12KSZD60063, 12KSZD60064, 12KSZD60065, 12KSZD60066, 12KSZD60067, 12KSZD60069, 12KSZD60070, 12KSZD60071, 12KSZD60072, 12KSZD60073, 12KSZD60075, 12KSZD60076, 12KSZD60077, 12KSZD60078, 12KSZD60079, 12KSZD60080, 12KSZD60081, 12KSZD60082, 12KSZD60083, 12KSZD60085, 12KSZD60086, 12KSZD60087, 12KSZD60088, 12KSZD60091, 12KSZD60092, 12KSZD60093, 12KSZD60094, 12KSZD60095, 12KSZD60096, 12KSZD60098, 12KSZD60099, 12KSZD60100, 12KSZD60101, 12KSZD60103, 12KSZD60105, 12KSZD60107, 12KSZD60108, 12KSZD60109, 12KSZD60110, 12KSZD60111, 12KSZD60112, 12KSZD60114, 12KSZD60115, 12KSZD60116, 12KSZD60117, 12KSZD60118, 12KSZD60120, 12KSZD60121, 12KSZD60122, 12KSZD60123, 12KSZD60124, 12KSZD60125, 12KSZD60126, 12KSZD60127, 12KSZD60128, 12KSZD60129, 12KSZD60130, 12KSZD60131, 12KSZD60134, 12KSZD60136, 12KSZD60137, 12KSZD60138, 12KSZD60139, 12KSZD60140, 12KSZD60141, 12KSZD60142, 12KSZD60143, 12KSZD60144, 12KSZD60145, 12KSZD60146, 12KSZD60147, 12KSZD60148, 12KSZD60149, 12KSZD60151, 12KSZD60152, 12KSZD60153, 12KSZD60154, 12KSZD60155, 12KSZD60156, 12KSZD60158, 12KSZD60159, 12KSZD60160, 12KSZD60162, 12KSZD60163, 12KSZD60164, 12KSZD60165, 12KSZD60166, 12KSZD60168, 12KSZD60169, 12KSZD60170, 12KSZD60171, 12KSZD60172, 12KSZD60173, 12KSZD60174, 12KSZD60176, 12KSZD60177, 12KSZD60178, 12KSZD60179, 12KSZD60180, 12KSZD60181, 12KSZD60182, 12KSZD60183, 12KSZD60186, 12KSZD60187, 12KSZD60188, 12KSZD60189, 12KSZD60190, 12KSZD60191, 12KSZD60192, 12KSZD60193, 12KSZD60194, 12KSZD60195, 12KSZD60197, 12KSZD60198, 12KSZD60199, 12KSZD60200, 12KSZD60201, 12KSZD60203, 12KSZD60204, 12KSZD60205, 12KSZD60206, 12KSZD60207, 12KSZD60209, 12KSZD60211, 12KSZD60212, 12KSZD60213, 12KSZD60214, 12KSZD60215, 12KSZD60216, 12KSZD60217, 12KSZD60219, 12KSZD60220, 12LSZD60001, 12LSZD60002, 12LSZD60003, 12LSZD60004, 12LSZD60006, 12LSZD60007, 12LSZD60008, 12LSZD60009, 12LSZD60011, 12LSZD60012, 12LSZD60014, 12LSZD60015, 12LSZD60017, 12LSZD60018, 12LSZD60020, 12LSZD60021, 12LSZD60024, 12LSZD60025, 12LSZD60026, 12LSZD60027, 12LSZD60028, 12LSZD60029, 12LSZD60030, 12LSZD60031, 12LSZD60032, 12LSZD60033, 12LSZD60034, 12LSZD60035, 12LSZD60036, 12LSZD60037, 12LSZD60040, 12LSZD60042, 12LSZD60043,12LSZD60044, 12LSZD60046, 12LSZD60047, 12LSZD60048, 12LSZD60049, 12LSZD60050, 12LSZD60051, 12LSZD60052, 12LSZD60053, 12LSZD60054,12LSZD60055, 12LSZD60056, 12LSZD60058, 12LSZD60059, 12LSZD60063, 12LSZD60064, 12LSZD60067, 12LSZD60068, 12LSZD60069, 12LSZD60070, 12LSZD60071, 12LSZD60072, 12LSZD60073, 12LSZD60074, 12LSZD60075, 12LSZD60076, 12LSZD60077, 12LSZD60078, 12LSZD60079, 12LSZD60080, 12LSZD60081, 12LSZD60083, 12LSZD60084, 12LSZD60085, 12LSZD60086, 12LSZD60088, 12LSZD60089, 12LSZD60091, 12LSZD60093, 12LSZD60094, 12LSZD60095, 12LSZD60096, 12LSZD60097, 12LSZD60098, 12LSZD60099, 12LSZD60100, 12LSZD60101, 12LSZD60102, 12LSZD60103, 12LSZD60104, 12LSZD60105, 12LSZD60107, 12LSZD60109, 12LSZD60110, 12LSZD60112, 12LSZD60113, 12LSZD60114, 12LSZD60115, 12LSZD60116, 12LSZD60117, 12LSZD60121, 12LSZD60122, 12LSZD60123, 12LSZD60124, 12LSZD60126, 12LSZD60127, 12LSZD60128, 12LSZD60129, 12LSZD60130, 12LSZD60131, 12LSZD60132, 12LSZD60134, 12LSZD60135, 12LSZD60136, 12LSZD60137, 12LSZD60139, 12LSZD60140, 12LSZD60141, 12LSZD60142, 12LSZD60144, 12LSZD60145, 12LSZD60146, 12LSZD60147, 12LSZD60149, 12LSZD60150, 12LSZD60151, 12LSZD60152, 12LSZD60155, 12LSZD60156, 12LSZD60158, 12LSZD60159, 12LSZD60160, 12LSZD60161, 12LSZD60162, 12LSZD60163, 12LSZD60164, 12LSZD60167, 12LSZD60168, 12LSZD60169, 12LSZD60170, 12LSZD60171, 12LSZD60172, 12LSZD60173, 12LSZD60174, 12LSZD60175, 12LSZD60176, 12LSZD60177, 12LSZD60178, 12LSZD60181, 12LSZD60182, 12LSZD60185, 12LSZD60187, 12LSZD60190, 12LSZD60192, 12LSZD60193, 12LSZD60194, 12LSZD60195, 12LSZD60197, 12LSZD60198, 12LSZD60199, 12LSZD60200, 12LSZD60201, 12LSZD60202, 12LSZD60203, 12LSZD60205, 12LSZD60206, 12LSZD60207, 12LSZD60208, 12LSZD60209, 12LSZD60210, 12LSZD60211, 12LSZD60212, 12LSZD60213, 12LSZD60214, 12LSZD60215, 12LSZD60216, 13CSZD60001, 13CSZD60002, 13CSZD60003, 13CSZD60004, 13CSZD60005, 13CSZD60006, 13CSZD60008, 13CSZD60009, 13CSZD60011, 13CSZD60013, 13CSZD60015, 13CSZD60016, 13CSZD60018, 13CSZD60019, 13CSZD60022, 13CSZD60023, 13CSZD60024, 13CSZD60025, 13CSZD60026, 13CSZD60028, 13CSZD60029, 13CSZD60031, 13CSZD60032, 13CSZD60034, 13CSZD60035, 13CSZD60036, 13CSZD60037, 13CSZD60039, 13CSZD60040, 13CSZD60041, 13CSZD60042, 13CSZD60044, 13CSZD60045, 13CSZD60046, 13CSZD60047, 13CSZD60048, 13CSZD60050, 13CSZD60051, 13CSZD60052, 13CSZD60054, 13CSZD60055, 13CSZD60056, 13CSZD60057, 13CSZD60058, 13CSZD60059, 13CSZD60066, 13CSZD60071, 13CSZD60072, 13CSZD60074, 13CSZD60075, 13CSZD60076, 13CSZD60077, 13CSZD60078, 13CSZD60079, 13CSZD60080, 13CSZD60082, 13CSZD60083, 13CSZD60084, 13CSZD60085, 13CSZD60087, 13CSZD60088, 13CSZD60091, 13CSZD60093, 13CSZD60095, 13CSZD60097, 13CSZD60098, 13CSZD60099, 13CSZD60100, 13CSZD60101, 13CSZD60102, 13CSZD60103, 13CSZD60104, 13CSZD60105, 13CSZD60107, 13CSZD60108, 13CSZD60110, 13CSZD60112, 13CSZD60113, 13CSZD60114, 13CSZD60115, 13CSZD60116, 13CSZD60117, 13CSZD60118, 13CSZD60119, 13CSZD60128, 13CSZD60130, 13CSZD60131, 13CSZD60133, 13CSZD60134, 13CSZD60135, 13CSZD60136, 13CSZD60138, 13CSZD60140, 13CSZD60141, 13CSZD60144, 13CSZD60146, 13CSZD60147, 13CSZD60148, 13CSZD60149, 13CSZD60150, 13CSZD60151, 13CSZD60153, 13CSZD60154, 13CSZD60155, 13CSZD60156, 13CSZD60157, 13CSZD60159, 13CSZD60163, 13CSZD60164, 13CSZD60166, 13CSZD60167, 13CSZD60168, 13CSZD60170, 13CSZD60171, 13CSZD60173, 13CSZD60175, 13CSZD60178, 13CSZD60180, 13CSZD60182, 13CSZD60183, 13CSZD60188, 13CSZD60189, 13CSZD60191, 13CSZD60193, 13CSZD60194, 13CSZD60198. 13CSZD60199, 13CSZD60200, 13CSZD60201, 13CSZD60203, 13CSZD60204, 13CSZD60205, 13CSZD60207, 13CSZD60208, 13CSZD60211, 13CSZD60213, 13CSZD60214, 13CSZD60215, 13CSZD60216, 13CSZD60219, 13CSZD60224, 13CSZD60225, 13CSZD60227, 13CSZD60229, 13DSZD60001, 13DSZD60002, 13DSZD60003, 13DSZD60004, 13DSZD60005, 13DSZD60007, 13DSZD60008, 13DSZD60010, 13DSZD60012, 13DSZD60013, 13DSZD60015, 13DSZD60016, 13DSZD60017, 13DSZD60019, 13DSZD60020, 13DSZD60021, 13DSZD60023, 13DSZD60024, 13DSZD60026, 13DSZD60027, 13DSZD60029, 13DSZD60030, 13DSZD60032, 13DSZD60033, 13DSZD60034, 13DSZD60035, 13DSZD60036, 13DSZD60037, 13DSZD60039, 13DSZD60040, 13DSZD60041, 13DSZD60042, 13DSZD60043, 13DSZD60044, 13DSZD60045, 13DSZD60046, 13DSZD60047, 13DSZD60055, 13DSZD60056, 13DSZD60057, 13DSZD60058, 13DSZD60059, 13DSZD60060, 13DSZD60061, 13DSZD60062, 13DSZD60065, 13DSZD60066, 13DSZD60067, 13DSZD60068, 13DSZD60069, 13DSZD60074, 13DSZD60075, 13DSZD60076, 13DSZD60077, 13DSZD60078, 13DSZD60079, 13DSZD60080, 13DSZD60081, 13DSZD60082, 13DSZD60084, 13DSZD60085, 13DSZD60086, 13DSZD60087, 13DSZD60088, 13DSZD60089, 13DSZD60091, 13DSZD60092, 13DSZD60095, 13DSZD60096, 13DSZD60097, 13DSZD60098, 13DSZD60099, 13DSZD60100, 13DSZD60101, 13DSZD60102, 13DSZD60103, 13DSZD60104, 13DSZD60105, 13DSZD60107, 13DSZD60108, 13DSZD60109, 13DSZD60110, 13ESZD60002, 13ESZD60003, 13ESZD60004, 13ESZD60005, 13ESZD60007, 13ESZD60008, 13ESZD60009, 13ESZD60010, 13ESZD60011l, 13ESZD60012, 13ESZD60014, 13ESZD60015, 13ESZD60016, 13ESZD60017, 13ESZD60018, 13ESZD60019, 13ESZD60020, 13ESZD60021, 13ESZD60022, 13ESZD60023, 13ESZD60024, 13ESZD60025, 13ESZD60027, 13ESZD60030, 13ESZD60031, 13ESZD60032, 13ESZD60033, 13ESZD60034, 13ESZD60035, 13ESZD60036, 13ESZD60037, 13ESZD60038, 13ESZD60039, 13ESZD60040, 13ESZD60041, 13FSZD60001, 13FSZD60002, 13FSZD60003, 13FSZD60004, 13FSZD60005, 13FSZD60006, 13FSZD60009, 13FSZD60011, 13FSZD60012, 13FSZD60013, 13FSZD60014, 13FSZD60015, 13FSZD60016, 13FSZD60017, 13FSZD60018, 13FSZD60019, 13FSZD60020, 13FSZD60021, 13FSZD60022, 13FSZD60023, 13FSZD60024, 13FSZD60025, 13FSZD60027, 13FSZD60030, 13FSZD60031, 13FSZD60032, 13FSZD60033, 13FSZD60034, 13FSZD60035, 13FSZD60036, 13FSZD60037, 13FSZD60040, 13FSZD60041, 13FSZD60042, 13FSZD60043, 13FSZD60044, 13FSZD60045, 13FSZD60048, 13FSZD60049, 13FSZD60050, 13FSZD60051, 13FSZD60052, 13FSZD60053,13FSZD60054, 13FSZD60055, 13FSZD60056, 13FSZD60057, 13FSZD60058, 13FSZD60060, 13FSZD60061, 13FSZD60062, 13FSZD60063, 13FSZD60064, 13FSZD60067, 13FSZD60068, 13FSZD60069, 13FSZD60070, 13FSZD60071, 13FSZD60072, 13FSZD60075, 13FSZD60076, 13FSZD60078, 13FSZD60079, 13FSZD60080, 13FSZD60081, 13FSZD60082, 13FSZD60084, 13FSZD60085, 13FSZD60086, 13FSZD60088, 13FSZD60089, 13FSZD60094, 13FSZD60096, 13FSZD60097, 13FSZD60098, 13FSZD60100, 13FSZD60101, 13FSZD60102,13FSZD60103, 13FSZD60104, 13FSZD60106, 13FSZD60107, 13FSZD60111, 13FSZD60112, 13FSZD60114, 13FSZD60115, 13FSZD60118, 13FSZD60119, 13FSZD60120, 13FSZD60121, 13GSZD60001, 13GSZD60003, 13GSZD60004, 13GSZD60006, 13GSZD60007, 13GSZD60010, 13GSZD60011, 13GSZD60013, 13GSZD60014, 13GSZD60015, 13GSZD60016, 13GSZD60017, 13GSZD60018, 13GSZD60019, 13GSZD60022, 13GSZD60023, 13GSZD60024, 13GSZD60025, 13GSZD60026, 13GSZD60027, 13GSZD60030, 13GSZD60031, 13GSZD60032, 13GSZD60033, 13GSZD60036, 13GSZD60037, 13GSZD60038, 13GSZD60040, 13GSZD60042, 13GSZD60043, 13GSZD60045, 13GSZD60046, 13GSZD60048, 13HSZD60004, 13HSZD60005, 13HSZD60006, 13HSZD60007, 13HSZD60009, 13HSZD60010, 13HSZD60011, 13HSZD60013, 13HSZD60014, 13HSZD60015, 13HSZD60017, 13JSZD60002, 13JSZD60004, 13JSZD60010, 13JSZD60011, 13JSZD60013, 13JSZD60014, 13JSZD60015, 13JSZD60017, 13JSZD60018, 13JSZD60019, 13JSZD60020, 13JSZD60021, 13JSZD60023, 13JSZD60026, 13JSZD60034, 13JSZD60035, 13JSZD60037, 13JSZD60038, 13JSZD60039, 13JSZD60040, 13JSZD60041, 13JSZD60043, 13JSZD60045, 13JSZD60046, 13JSZD60047, 13JSZD60048, 13JSZD60049, 13JSZD60050, 13JSZD60051, 13JSZD60052, 13JSZD60053, 13JSZD60054, 13JSZD60055, 13JSZD60058, 13JSZD60060, 13JSZD60061, 13JSZD60063, 13JSZD60064, 13JSZD60065, 13JSZD60067, 13JSZD60069, 13JSZD60071, 13JSZD60074, 13JSZD60076, 13JSZD60077, 13JSZD60078, 13JSZD60079, 13JSZD60080, 14ESZD60005, 14ESZD60007, 14ESZD60008, 14ESZD60015, 14ESZD60016, 14ESZD60020, 14ESZD60021, 14ESZD60023, 14ESZD60024, 14ESZD60025, 14ESZD60038, 14ESZD60039, 14ESZD60044,14ESZD60057, 14ESZD60058, 14ESZD60061, 14ESZD60062, 14ESZD60063, 14ESZD60067, 14ESZD60069, 14ESZD60070, 14ESZD60071, 14ESZD60075, 14ESZD60076, 14ESZD60078, 14ESZD60080, 14ESZD60082, 14ESZD60087, 14ESZD60089, 14ESZD60090, 14ESZD60092, 14ESZD60093, 14ESZD60094, 14ESZD60096, 14ESZD60101, 14ESZD60103, 14ESZD60104, 14ESZD60105, 14ESZD60106, 14ESZD60108, 14ESZD60109, 14ESZD60111, 14ESZD60112, 14ESZD60113 & 14ESZD60114. Model #AIRPTMBDY20, Serial #s: 12KSZD60009, 12KSZD60025, 12KSZD60084, 12KSZD60097, 12KSZD60104, 12KSZD60132, 12KSZD60135, 12KSZD60150, 12LSZD60016, 12LSZD60045, 12LSZD60061, 12LSZD60065, 12LSZD60082, 12LSZD60087, 12LSZD60111, 12LSZD60119, 12LSZD60143, 12LSZD60148, 12LSZD60153, 13CSZD60021, 13CSZD60049, 13CSZD60060, 13CSZD60106, 13CSZD60111, 13CSZD60123, 13CSZD60129, 13CSZD60142, 13CSZD60158, 13CSZD60192, 13CSZD60202, 13CSZD60210, 13DSZD60004, 13DSZD60006, 13DSZD60011, 13DSZD60022, 13DSZD60038, 13DSZD60050, 13DSZD60051, 13DSZD60052, 13DSZD60054, 13ESZD60001, 13ESZD60013, 13ESZD60026, 13FSZD60007, 13FSZD60026, 13FSZD60029, 13FSZD60077, 13FSZD60091, 13FSZD60092, 13FSZD60095, 13FSZD60099, 13GSZD60029, 13GSZD60044, 13HSZD60012, 13HSZD60016, 13JSZD60024, 13JSZD60033, 13JSZD60044, 13JSZD60056, 13JSZD60066, 14ESZD60017, 14ESZD60018, 14ESZD60027, 14ESZD60037, 14ESZD60040, 14ESZD60077, 14ESZD60081, 14ESZD60083 & 14ESZD60084. Model #AIRPTBASE, Serial #s: 12KSZD60068, 12KSZD60119, 12LSZD60106, 12LSZD60166, 12LSZD60180, 13CSZD60092, 13DSZD60063, 13DSZD60064, 13DSZD60090, 13FSZD60008, 13FSZD60028, 13FSZD60046, 13FSZD60047, 13HSZD60001, 13JSZD60001, 13JSZD60068, 14ESZD60014, 14ESZD60022, 14ESZD60065, 14ESZD60068, 14ESZD60072, 14ESZD60074, 14ESZD60086, 14ESZD60088, 14ESZD60091, 14ESZD60097, 14ESZD60098, 14ESZD60102 & 14ESZD60107.
Worldwide Distribution - US (nationwide) Distribution including Puerto Rico, and to the states of : AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA and WI., and to the countries of : AUSTRALIA, CANDA, FRANCE, NEW ZEALAND, SOUTH AFRICA, TRINIDAD AND TOBAGO
1,057 finished devices