FDA Enforcement
Class II
Ongoing
Brand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 Product Description: Surgical Awl Component: No
Recall: Z-2357-2024
·
Reported July 24, 2024
Enforcement
- Recall Number
- Z-2357-2024
- Event ID
- 94872
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- SEASPINE ORTHOPEDICS CORPORATION
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 24, 2024
- Initiation Date
- June 4, 2024
- Classification Date
- July 15, 2024
- Address
- 5770 Armada Dr, Carlsbad, CA, 92008-4608, United States
Description
Brand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 Product Description: Surgical Awl Component: No
Reason
Insufficient weld around the cap component of the instrument is insufficient to withstand impaction forces.
Code Info
Lot Code: UDI 10889981216368 Lot Numbers TT0190H, TT0191H
Distribution
The product was distributed to the following US states: Nevada, Colorado, Texas, Washington, Indiana, Georgia, Michigan, Florida, Pennsylvania, Illinois, Tennessee, Virginia, Missouri, South Dakota, California, Montana, Minnesota, Arizona
Quantity
100 units