BD FACS Sample Prep Assistant (SPA) II Upgrade - Product Usage: K050191: The BD FACSCanto system with BD FACSCanto clinical software when used with the BD FACS Sample Prep Assistant II /III is intended for use as an in vitro diagnostic device for identification and enumeration of lymphocyte subsets in human cells in suspension using a lyse / no-wash sample preparation method for flow cytometry.
Enforcement
- Recall Number
- Z-2357-2020
- Event ID
- 85129
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton, Dickinson and Company, BD Biosciences
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 17, 2020
- Initiation Date
- July 8, 2019
- Classification Date
- June 11, 2020
- Termination Date
- January 31, 2023
- Address
- 2350 Qume Dr, N/A, San Jose, CA, 95131-1812, United States
Description
BD FACS Sample Prep Assistant (SPA) II Upgrade - Product Usage: K050191: The BD FACSCanto system with BD FACSCanto clinical software when used with the BD FACS Sample Prep Assistant II /III is intended for use as an in vitro diagnostic device for identification and enumeration of lymphocyte subsets in human cells in suspension using a lyse / no-wash sample preparation method for flow cytometry.
The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.
Catalog (Ref) No. 650686
Worldwide distribution - US Nationwide distribution including in the states of AL, CO,CT, GA, IL, MA, MI, MD, NC, NH,NJ, NY, OH, PA, TN, TX, VA, and VT. The countries of Australia, Botswana, Brazil, Canada, Chile, China, Colombia, Costa Rica, Europe, Hong Kong, India, Italy, Jamaica, Japan, Kazakhstan, Malaysia, Martinique, Mexico, New Zealand, Nicaragua, Peru, Puerto Rico (US), Romania, Russia, Singapore, South Africa, South Korea, Swaziland, Switzerland, Turkey, US, Vietnam.
51 units