FDA Enforcement
Class II
Ongoing
Microdot Xtra blood glucose test strips
Recall: Z-2357-2019
·
Reported August 28, 2019
Enforcement
- Recall Number
- Z-2357-2019
- Event ID
- 82867
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Cambridge Sensors Limited
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 28, 2019
- Initiation Date
- July 26, 2018
- Classification Date
- August 22, 2019
- Address
- Units 9 & 10, Cardinal Park, Huntingdon, N/A, N/A, United Kingdom
Description
Microdot Xtra blood glucose test strips
Reason
The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.
Code Info
LOT NUMBERS 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702.
Distribution
Illinois
Quantity
17,784 (USA)