FDA Enforcement Class II Ongoing

Microdot Xtra blood glucose test strips

Recall: Z-2357-2019 · Reported August 28, 2019

Enforcement

Recall Number
Z-2357-2019
Event ID
82867
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Cambridge Sensors Limited
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 28, 2019
Initiation Date
July 26, 2018
Classification Date
August 22, 2019
Address
Units 9 & 10, Cardinal Park, Huntingdon, N/A, N/A, United Kingdom

Description

Microdot Xtra blood glucose test strips

Reason

The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.

Code Info

LOT NUMBERS 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702.

Distribution

Illinois

Quantity

17,784 (USA)