microTargeting" Drive System Distributed by Medtronic, Product Number 9033G0601
Enforcement
- Recall Number
- Z-2356-2018
- Event ID
- 80148
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- FHC, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 11, 2018
- Initiation Date
- May 10, 2018
- Classification Date
- July 5, 2018
- Termination Date
- April 30, 2019
- Address
- 1201 Main St, Bowdoin, ME, 04287-7302, United States
Description
microTargeting" Drive System Distributed by Medtronic, Product Number 9033G0601
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
REF 9033G0601
The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.
2952 total