FDA Enforcement Class II Ongoing

Accu-Chek Inform II Test Strip, Whole Blood Glucose Test System, Model number 05942891001

Recall: Z-2354-2021 · Reported September 1, 2021

Enforcement

Recall Number
Z-2354-2021
Event ID
88362
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Roche Diabetes Care, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 1, 2021
Initiation Date
July 28, 2021
Classification Date
August 26, 2021
Address
9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States

Description

Accu-Chek Inform II Test Strip, Whole Blood Glucose Test System, Model number 05942891001

Reason

Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.

Code Info

Since this is a labeling correction, this recall is not lot specific and is applicable to all non-expired lots.

Distribution

Worldwide distribution - US Nationwide distribution.

Quantity

9,555,555 (US); 5,128,758 (OUS)