FDA Enforcement
Class II
Terminated
PathoDx Strep B Grouping Latex, REF 62031, 3 ml packaged in plastic dropper bottle with a pink cap. Latex reagent for streptococcal group B; Identify beta-hemolytic streptococci of Lancefield Groups A, B, C, F and G from primary culture plates.
Recall: Z-2351-2015
·
Reported August 19, 2015
Enforcement
- Recall Number
- Z-2351-2015
- Event ID
- 71614
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Remel Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 19, 2015
- Initiation Date
- June 30, 2015
- Classification Date
- August 7, 2015
- Termination Date
- February 13, 2017
- Address
- 12076 Santa Fe Trail Dr, N/A, Lenexa, KS, 66215-3519, United States
Description
PathoDx Strep B Grouping Latex, REF 62031, 3 ml packaged in plastic dropper bottle with a pink cap. Latex reagent for streptococcal group B; Identify beta-hemolytic streptococci of Lancefield Groups A, B, C, F and G from primary culture plates.
Reason
A reagent present may produce weak or slow reactions; continued use may result in a failure to identify target species.
Code Info
lot 1463129, exp. date 31 May 2016
Distribution
Nationwide Distribution, including Puerto Rico and the country of Canada.
Quantity
4420 bottles