FDA Enforcement
Class II
Ongoing
LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 19-Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-119/00
Recall: Z-2350-2023
·
Reported August 16, 2023
Enforcement
- Recall Number
- Z-2350-2023
- Event ID
- 92758
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 16, 2023
- Initiation Date
- July 3, 2023
- Classification Date
- August 4, 2023
- Address
- Oststr. 4-10, N/A, Norderstedt, N/A, N/A, Germany
Description
LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 19-Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-119/00
Reason
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
Code Info
UDI-DI: 04026575182138 Lot Numbers: B922122 B932054 C005071 C211013
Distribution
US Nationwide distribution.
Quantity
24 units