MX 16-slice SKD whole body computed tomography X-ray system. Imaging diagnostic tool.

Recall: Z-2347-2016 · Reported August 17, 2016

Enforcement

Recall Number
Z-2347-2016
Event ID
74402
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 17, 2016
Initiation Date
April 20, 2016
Classification Date
August 5, 2016
Termination Date
July 2, 2018
Address
595 Miner Rd, Cleveland, OH, 44143-2131, United States

Description

MX 16-slice SKD whole body computed tomography X-ray system. Imaging diagnostic tool.

Reason

The following issues have been found in MX 16-slice systems with software version 1.1.4.21426: 1. During the filming operation on MX16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patients images before pasting, a previous patients images may be present in the clipboard and be copied onto the film of the current patient. 2. During the Bolus Tracking scan, if the Auto Voice in Tracker scan is enabled, the Tracker scan will be aborted unexpectedly and the diagnostic scan after the tracker scan needs to be manually started. 3. If the scan protocol with SAS (Spiral Auto Start) function is selected to plan the scan, the SAS option may not be displayed on the Contrast tab of the scan protocol parameter area or the SAS option is displayed but not enabled as pre-configured in the protocol. This issue only occurs on the first helical scan after system startup that applies SAS function

Code Info

Model No. 728132; S/N: EP16ER100003, EP16ER100004, EP16ER100005, EP16ER100007, EP16ER100008, EP16ER110003, EP16ER110004, EP16ER110005, EP16ER110006, EP16ER110007, EP16ER110008, EP16ER110009, EP16ER110010, EP16ER110011, EP16ER110012, EP16ER110013, EP16ER110014, EP16ER110015, EP16ER110016, EP16ER110017, EP16ER110018, EP16ER110019, EP16ER110020, EP16ER110021, EP16ER110022, EP16ER110023, EP16ER110024, EP16ER120001, EP16ER120002, EP16ER120003, EP16ER120004, EP16ER120005, EP16ER120006, EP16ER120007, EP16ER120008, EP16ER120010, EP16ER120012, EP16ER120015, EP16ER120016, EP16ER120017, EP16ER120018, EP16ER130001, EP16ER130002, EP16ER130005, EP16ER130006, EP16ER130007, EP16ER130008, EP16ER130009, EP16ER130010, EP16ER130011, EP16ER130012, EP16ER130013, EP16ER130014, EP16ER130015, EP16ER130016, EP16ER130018, EP16ER130025, EP16ER130026, EP16ER130029, EP16ER130033, EP16ER130034, EP16EB150009 & EP16EB150014. Model No.728135; S/N: EP16E120009, EP16E120011, EP16E120013, EP16ER130003, EP16ER130004, EP16ER130017, EP16ER130019, EP16ER130020, EP16ER130021, EP16ER130022, EP16ER130023, EP16ER130024, EP16ER130027, EP16ER130028, EP16ER130030, EP16ER130031 & EP16ER130032.

Distribution

Products were sold in the following states: AL, AR, CA, FL, GA, IL, IN, MS, NC, ND, NJ, NV, NY, OH, PA, SC, TN & VA. Products were also shipped to the following countries: Albania, ALGERIA, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Burundi, Cape Verde, Chile, China, Colombia, Congo, Democratic, Costa Rica, Cyprus, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Hungary, India, INDONESIA, Ireland, Italy, Japan, Jordan, Latvia, Lebanon, Libyan Arab Jamahiri, Lithuania, Malawi, Mali, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, Nigeria, Oman, Palestine, Paraguay, Philippines, Poland, Romania, Russia, Senegal, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, United Kingdom, Venezuela, Viet Nam & Zambia.

Quantity

80 units