FDA Enforcement
Class II
Terminated
Volcano brand Intravascular Ultrasound, Volcano s5/s5i Family Software Version 3.2.1 or 3.2.2, Models/Part Numbers: 435-0607.01, 435-0607.04, 435-0602.01, 435-0602.02; Product is manufactured and distributed by Volcano Corporation, Rancho Cordova, CA. Medical device for use in imaging of histology.
Recall: Z-2345-2012
·
Reported September 19, 2012
Enforcement
- Recall Number
- Z-2345-2012
- Event ID
- 63102
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Volcano Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 19, 2012
- Initiation Date
- August 24, 2012
- Classification Date
- September 10, 2012
- Termination Date
- November 25, 2014
- Address
- 2870 Kilgore Rd, N/A, Rancho Cordova, CA, 95670-6133, United States
Description
Volcano brand Intravascular Ultrasound, Volcano s5/s5i Family Software Version 3.2.1 or 3.2.2, Models/Part Numbers: 435-0607.01, 435-0607.04, 435-0602.01, 435-0602.02; Product is manufactured and distributed by Volcano Corporation, Rancho Cordova, CA. Medical device for use in imaging of histology.
Reason
The display on the system has the potential to change. In house testing eventually identified the unique sequence of events that must occur in order for this issue to manifest itself.
Code Info
All units of these models. Models/Part Numbers: 435-0607.01, 435-0607.04, 435-0602.01, 435-0602.02
Distribution
Worldwide Distribution.
Quantity
1020 units