FDA Enforcement Class II Terminated

Artis zee/ zeego, Artis Q/ Q.zen, stand alone system, software controlled Model numbers: 10094135, 10094137, 10094139, 10094141, 10280959, 10848281, 10848282, 10848283, 10848353, 10848255 Artis zee / zeego and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Recall: Z-2344-2016 · Reported August 17, 2016

Enforcement

Recall Number
Z-2344-2016
Event ID
74655
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 17, 2016
Initiation Date
July 7, 2016
Classification Date
August 5, 2016
Termination Date
December 13, 2016
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

Artis zee/ zeego, Artis Q/ Q.zen, stand alone system, software controlled Model numbers: 10094135, 10094137, 10094139, 10094141, 10280959, 10848281, 10848282, 10848283, 10848353, 10848255 Artis zee / zeego and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Reason

Due to an error in the 19 Live Display, image reproduction may fail in the examination room and the potential exists for the loss of images immediately after system startup.

Code Info

Serial Numbers: 121185,147996,121169,105032,109190,109191,109197,109198,125100,158225,161021,154847,124083,123050,109187,124102,121167,125021,121163,117125,148001,158234,148010,158229,121171, 109183,109181,131101,158226,109194,117124,158244,109179,158238,109188,121179,137714,121168,158246,161022,121175,121173,154859,109186,158230,123046,158227,109189,123052,105029,105030

Distribution

Nationwide Distribution to AL,AZ,CA,CO,IA,IL,MA,MD,MI,MN,MO,NE,NJ,NY,OH,OR,PA,RI,TX,VA,WA,WI,WV

Quantity

51 units