FDA Enforcement
Class II
Terminated
RayStation 6 Service Pack 3, UDI # 07350002010242 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Recall: Z-2343-2020
·
Reported June 17, 2020
Enforcement
- Recall Number
- Z-2343-2020
- Event ID
- 85723
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- RAYSEARCH LABORATORIES AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- June 17, 2020
- Initiation Date
- May 18, 2020
- Classification Date
- June 10, 2020
- Termination Date
- August 6, 2021
- Address
- Sveavagen 9, N/A, Stockholm, N/A, N/A, Sweden
Description
RayStation 6 Service Pack 3, UDI # 07350002010242 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason
Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Code Info
UDI # 07350002010242 Build number: 6.3.0.6
Distribution
US Nationwide distribution including in the states of Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.
Quantity
N/A