FDA Enforcement Class II Terminated

RayStation 5 Service Pack 3, UDI # 07350002010020 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.

Recall: Z-2340-2020 · Reported June 17, 2020

Enforcement

Recall Number
Z-2340-2020
Event ID
85723
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
RAYSEARCH LABORATORIES AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 17, 2020
Initiation Date
May 18, 2020
Classification Date
June 10, 2020
Termination Date
August 6, 2021
Address
Sveavagen 9, N/A, Stockholm, N/A, N/A, Sweden

Description

RayStation 5 Service Pack 3, UDI # 07350002010020 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.

Reason

Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.

Code Info

UDI # 07350002010020 Build number: 5.0.3.17

Distribution

US Nationwide distribution including in the states of Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.

Quantity

N/A