Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to support the implanted device, and for this recall the software that resides on the programmer (software model 3330) is the method of delivery of the new firmware.

Recall: Z-2335-2018 · Reported July 11, 2018

Enforcement

Recall Number
Z-2335-2018
Event ID
80244
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
St Jude Medical Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 11, 2018
Initiation Date
April 16, 2018
Classification Date
June 29, 2018
Termination Date
February 4, 2021
Address
15900 Valley View Ct, Sylmar, CA, 91342-3577, United States

Description

Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to support the implanted device, and for this recall the software that resides on the programmer (software model 3330) is the method of delivery of the new firmware.

Reason

The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal.

Code Info

Software Versions: 24.0.1 rev 1

Distribution

U.S. Nationwide Distribution: PR, MI, NJ, IL, FL, AR, PA, KY, CA, TN, IN, AL, TX, AZ, SC, OK, VA, NH, NM, WA, CO, IA, WI, NV, UT, OH, NC, MO, DE, WV, MD, OR, NE, VT, KS, MA, IA, WY, CT, GA, MS, LA, DC, LA,

Quantity

0