Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to support the implanted device, and for this recall the software that resides on the programmer (software model 3330) is the method of delivery of the new firmware.
Enforcement
- Recall Number
- Z-2335-2018
- Event ID
- 80244
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- St Jude Medical Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- July 11, 2018
- Initiation Date
- April 16, 2018
- Classification Date
- June 29, 2018
- Termination Date
- February 4, 2021
- Address
- 15900 Valley View Ct, Sylmar, CA, 91342-3577, United States
Description
Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to support the implanted device, and for this recall the software that resides on the programmer (software model 3330) is the method of delivery of the new firmware.
The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal.
Software Versions: 24.0.1 rev 1
U.S. Nationwide Distribution: PR, MI, NJ, IL, FL, AR, PA, KY, CA, TN, IN, AL, TX, AZ, SC, OK, VA, NH, NM, WA, CO, IA, WI, NV, UT, OH, NC, MO, DE, WV, MD, OR, NE, VT, KS, MA, IA, WY, CT, GA, MS, LA, DC, LA,
0