FDA Enforcement Class II Completed

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Recall: Z-2332-2024 · Reported July 17, 2024

Enforcement

Recall Number
Z-2332-2024
Event ID
94731
Classification
Class II
Status
Completed
Product Type
Devices
Firm
Barco N.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 17, 2024
Initiation Date
April 24, 2024
Classification Date
July 11, 2024
Address
President Kennedypark 35, N/A, Kortrijk, N/A, N/A, Belgium

Description

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Reason

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

Code Info

serial numbers: 2530986299, 2530986300, 2530986301, 2530986302, 2530986303, 2530986304, 2530986305, 2530986306, 2530986307, 2530986308, 2530986309, 2530986310, 2530986311, 2530986312, 2530986313, 2530986314, 2530986315, 2530986316, 2530986317, 2530986318, 2530986319, 2530986320, 2530986321, 2530986322, 2530986323, 2530986324, 2530986325, 2530986326, 2530986327, 2531041594, 2531041591, 2531041578, 2531041585, 2531041595, 2531041596, 2531041586, 2531041588, 2531041587, 2531041577, 2531041584, 2531041581, 2531041612, 2531041608, 2531041600, 2531041601, 2531041604, 2531041599, 2531041602, 2531041605, 2531041606, 2531041607, 2531041609, 2531041610, 2531041580, 2531041583, 2531041598, 2531041593, 2531041590, 2531041614, 2531041582, 2531041613, 2531041589, 2531041616, 2531041579, 2531041615, 2531041603, 2531041597, 2531041592, 2530985205, 2530986280, 2530986281, 2530986282, 2530986283, 2530986284, 2530986285, 2530986286, 2530986287, 2530986288, 2530986289, 2530986290, 2530986291, 2530986292, 2530986293, 2530986294, 2530986295, 2530986296, 2530986297, 2530986298

Distribution

US: FL

Quantity

88 units