FDA Enforcement Class II Terminated

EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together with Autostainer Link Instruments.

Recall: Z-2330-2016 · Reported August 10, 2016

Enforcement

Recall Number
Z-2330-2016
Event ID
74608
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Dako Denmark A/S
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 10, 2016
Initiation Date
July 6, 2016
Classification Date
August 3, 2016
Termination Date
October 13, 2016
Address
Produktionsvej 42, Glostrup, N/A, Denmark

Description

EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together with Autostainer Link Instruments.

Reason

The pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1.

Code Info

K8002 Lot no: 10112879 K8004 Lot no: 10111463

Distribution

Worldwide Distribution -- US, Canada, and Colombia.

Quantity