FDA Enforcement Class II Terminated

Merlin PCS programmer (Model # 3650), Software (Model # 3330)

Recall: Z-2329-2018 · Reported July 4, 2018

Enforcement

Recall Number
Z-2329-2018
Event ID
80212
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
St Jude Medical Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 4, 2018
Initiation Date
April 16, 2018
Classification Date
June 28, 2018
Termination Date
March 8, 2024
Address
15900 Valley View Ct, Sylmar, CA, 91342-3577, United States

Description

Merlin PCS programmer (Model # 3650), Software (Model # 3330)

Reason

The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.

Code Info

Software Versions: 24.1.1 rev1 (US), 23.1.2 rev 1 (OUS), 23.1.3 (Japan), and 23.1.4 rev 1 (Canada)

Distribution

U.S. Nationwide distribution

Quantity

71651