FDA Enforcement Class I Ongoing

Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.

Recall: Z-2328-2025 · Reported August 27, 2025

Enforcement

Recall Number
Z-2328-2025
Event ID
97079
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Hamilton Medical AG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 27, 2025
Initiation Date
June 3, 2025
Classification Date
August 19, 2025
Address
Parc Industrial Vial 10, N/A, Domat/Ems, N/A, N/A, Switzerland

Description

Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.

Reason

A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.

Code Info

Model/Catalog Number: 160021 ESM board Part Number: 10144626 GTIN: 07630002808590 UDI-DI: 76300028PN160021ZW Affected SN range: 13636 till 13950 and 13966 till 14753 13797 13801 13806 13855 13865 13981 13988 13989 14012 14015 14019 14034 14035 14042 14052 14054 14102 14109 14112 14113 14114 14115 14116 14119 14120 14121 14122 14123 14124 14126 14127 14128 14163 14168 14169 14170 14172 14173 14175 14177 14187 14188 14189 14195 14197

Distribution

US Nationwide distribution.

Quantity

1040 systems