Kosmos on iOS impacting iOS, Model Numbers: P007639, P007944, software is part of a general-purpose diagnostic ultrasound system.
Enforcement
- Recall Number
- Z-2325-2024
- Event ID
- 94709
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Echonous Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 17, 2024
- Initiation Date
- May 9, 2024
- Classification Date
- July 10, 2024
- Address
- 8310 154th Ave Ne Ste 200 Bldg B, N/A, Redmond, WA, 98052-6180, United States
Description
Kosmos on iOS impacting iOS, Model Numbers: P007639, P007944, software is part of a general-purpose diagnostic ultrasound system.
Ultrasound system has a bug in affected iOS software, which after DICOM export, can cause inaccuracy in postprocessed measurements done in 3rd party software, such as PACS or other post-processing software, which may lead to inaccurate diagnoses or treatment.
Model/UDI-DI/Software versions: P007639/09353278000717/1.0, P007944/09353278000762/ 2.1, 2.2.
US: ID, TX, WI, KS, IL, GA, MI, CA, OH, FL, AK, UT, MD, NY, NV, CO, ND, WA, DC, LA, KY, MN, HI, NM, TN, NJ, ID, MI, VT, AZ, MO, AL, MA, NC OUS: Denmark, Switzerland, Austria, Italy, France, United Kingdom, Germany, Spain, Romania, Finland, Romania, Australia, India, Hong Kong, Thailand