BK Medical ApS Ultrasound System Scanner bk2300 - Model #: bk3500 with software releases 1.2.0, 1.2.1 and 1.2.2. The system is intended for diagnostic ultrasound imaging or fluid flow analysis of the human body, data processing and guidance of puncture and biopsy. The system performs simple geometric measurements and calculations in the following areas: Emergency Medicine, Anesthesia, MSK, Vascular, Cardiology, OB/GYN
Enforcement
- Recall Number
- Z-2325-2016
- Event ID
- 74691
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- B-K Medical A/S
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 10, 2016
- Initiation Date
- June 28, 2016
- Classification Date
- August 3, 2016
- Termination Date
- August 29, 2016
- Address
- Mileparken 34, Herlev, N/A, Denmark
Description
BK Medical ApS Ultrasound System Scanner bk2300 - Model #: bk3500 with software releases 1.2.0, 1.2.1 and 1.2.2. The system is intended for diagnostic ultrasound imaging or fluid flow analysis of the human body, data processing and guidance of puncture and biopsy. The system performs simple geometric measurements and calculations in the following areas: Emergency Medicine, Anesthesia, MSK, Vascular, Cardiology, OB/GYN
In the system (bk3500) software generic volume measurement (HxWxLxl) function the user may be unaware of the calculation factor (1) and assume the generic volume calculation factor is similar to that of the other (prostate, adenoma, testis, kidney, and bladder) volume calculation factors provided in the software.
Serial Numbers: 5007750, 5008004, 5007734,5007742, 5007596, and50074277
Worldwide Distribution - US to TX, FL, and MA. Internationally to Australia