FDA Enforcement
Class II
Terminated
Medtronic Passive Biopsy Needle Kit, Part Number 9731754 - Passive Biopsy Fee Per Use (FPU) Kit. The image guided Passive Biopsy Needle is an accessory instrument with the StealthStation System. The Biopsy Needle is intended to be used for stereotaxy biopsy of brain tissue.
Recall: Z-2325-2015
·
Reported August 12, 2015
Enforcement
- Recall Number
- Z-2325-2015
- Event ID
- 71756
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medtronic Navigation, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 12, 2015
- Initiation Date
- July 17, 2015
- Classification Date
- August 3, 2015
- Termination Date
- April 7, 2016
- Address
- 826 Coal Creek Cir, N/A, Louisville, CO, 80027-9710, United States
Description
Medtronic Passive Biopsy Needle Kit, Part Number 9731754 - Passive Biopsy Fee Per Use (FPU) Kit. The image guided Passive Biopsy Needle is an accessory instrument with the StealthStation System. The Biopsy Needle is intended to be used for stereotaxy biopsy of brain tissue.
Reason
The adhesive between the sphere assembly and the biopsy needle shaft may be compromised, causing the sphere assembly to move on the needle shaft. This may result in inaccurate navigation and improper placement of the needles biopsy window in the brain.
Code Info
LOT No: 0007525461. 0007599312
Distribution
Worldwide Distribution - US (nationwide) and Canada.
Quantity
1415 total