FDA Enforcement Class II Terminated

Aequalis Humeral Nail Targeting Jig, Product Code: 9020060. Orthopedic manual surgical instrument.

Recall: Z-2315-2015 · Reported August 5, 2015

Enforcement

Recall Number
Z-2315-2015
Event ID
71605
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Tornier, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 5, 2015
Initiation Date
June 25, 2015
Classification Date
July 29, 2015
Termination Date
August 31, 2017
Address
10801 Nesbitt Ave S, N/A, Bloomington, MN, 55437-3109, United States

Description

Aequalis Humeral Nail Targeting Jig, Product Code: 9020060. Orthopedic manual surgical instrument.

Reason

Recall for the Aequalis IM Nail instrumentation set (Tray Number 9020000) due to several reports that the mast of the targeting jig (Part Number 9020060) is separating from the jig boom. Separation will impact alignment and can affect the ability to fixate Aequalis IM Nail screws.

Code Info

Lots: DB0114069, DB0114349, CA46096, CA216052, DB0111264001, DB0112016, DB0113102.

Distribution

US: AL, AK, AZ, CA, CO, CT, FL, IL, IN, ME, MD, MI, MO, NC, OH, PA, TX. OUS: CANADA AND FRANCE.

Quantity

99