Hudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management
Enforcement
- Recall Number
- Z-2311-2019
- Event ID
- 83461
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Teleflex Medical
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 18, 2019
- Initiation Date
- July 30, 2019
- Classification Date
- September 9, 2019
- Termination Date
- February 15, 2022
- Address
- 3015 Carrington Mill Blvd, Morrisville, NC, 27560-5437, United States
Description
Hudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
a) REF 5-11108 Batch numbers: 73L1600010 73A1700644 73C1700236 73D1700438 73E1700473 73F1700385 73K1700066 73L1700501 73B1800123 73D1800137 73F1800172 73G1800124 73G1800925 73H1800731 b) REF 5-11110 Batch numbers: 73K1600368 73K1600727 73A1700645 73B1700410 73C1700237 73D1700624 73E1700474 73F1700386 73H1700040 73H1700482 73H1700599 73J1700291 73K1700067 73L1700502 73M1700452 73B1800124 73D1800313 73F1800173 73G1800125 73K1800170 73L1800224 c) REF 5-11112 Batch numbers: 73K1600369 73L1600011 73A1700646 73C1700238 73D1700625 73E1700475 73F1700387 73H1700041 73H1700483 73J1700292 73L1700503 73M1700453 73A1800671 73B1800350 73D1800138 73F1800174 73G1800126 73G1800806 73J1800308 73K1800173 73L1800225 73K1800172
US Nationwide Distribution
7963 units