FDA Enforcement Class II Ongoing

OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004

Recall: Z-2310-2021 · Reported August 25, 2021

Enforcement

Recall Number
Z-2310-2021
Event ID
88344
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
OPTI Medical Systems, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 25, 2021
Initiation Date
June 25, 2021
Classification Date
August 17, 2021
Address
235 Hembree Park Dr, N/A, Roswell, GA, 30076-5738, United States

Description

OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004

Reason

The firm identified a precipitation issue which may cause late Ct reactors (a false positive result) with a rate of 5 to 10% for affected vials of RNA Master Mix.

Code Info

a) 99-57003, Lot Numbers: 20808, 20812, 20813 b) 99-57004, Lot Numbers: 20881, 20901, 211013

Distribution

Maine, New York, Alabama, Pennsylvania, Florida, Illinois, New Jersey, Connecticut, Georgia, Puerto Rico, Costa Rica, Hong Kong, Philippines, Barbados

Quantity

3552 kits