FDA Enforcement Class II Terminated

ANSPACH - 4MM Diamond Ball, Extends 27MM; REF S-4D-L-G1 all lot numbers Cutting shaping bone including spine and cranium.

Recall: Z-2308-2015 · Reported August 5, 2015

Enforcement

Recall Number
Z-2308-2015
Event ID
71212
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
The Anspach Effort, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 5, 2015
Initiation Date
April 13, 2015
Classification Date
July 29, 2015
Termination Date
August 25, 2016
Address
4500 Riverside Dr, N/A, Palm Beach Gardens, FL, 33410-4235, United States

Description

ANSPACH - 4MM Diamond Ball, Extends 27MM; REF S-4D-L-G1 all lot numbers Cutting shaping bone including spine and cranium.

Reason

G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.

Code Info

REF S-4D-L-G1 ; ALL lot numbers.

Distribution

Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.

Quantity

246